The RoScan Thermal 3D Body Scanning System: Medical Applicability and Benefits for Unobtrusive Sensing and Objective Diagnosis

Chromý, Adam; Žalud, Luděk

Abstract

The RoScan is a novel, high-accuracy multispectral surface scanning system producing colored 3D models that include a thermal layer. (1) Background: at present, medicine still exhibits a lack of objective diagnostic methods. As many diseases involve thermal changes, thermography may appear to be a convenient technique for the given purpose; however, there are three limiting problems: exact localization, resolution vs. range, and impossibility of quantification. (2) Methods: the basic principles and benefits of the system are described. The procedures rely on a robotic manipulator with multiple sensors to create a multispectral 3D model. Importantly, the structure is robust, scene-independent, and features quantifiable measurement uncertainty; thus, all of the above problems of medical thermography are resolved. (3) Results: the benefits were demonstrated by several pilot case studies: medicament efficacy assessment in dermatology, objective recovery progress assessment in traumatology, applied force quantification in forensic sciences, exact localization of the cause of pain in physiotherapy, objective assessment of atopic dermatitis, and soft tissue volumetric measurements. (4) Conclusion: the RoScan addresses medical quantification, which embodies a frequent problem in several medical sectors, and can deliver new, objective information to improve the quality of healthcare and to eliminate false diagnoses.

Full text

sensors Article The RoScan Thermal 3D Body Scanning System: Medical Applicability and Benefits for Unobtrusive Sensing and Objective Diagnosis Adam Chromy 1,2,* and Ludek Zalud 1,2 1CEITEC—Central European Institute of Technology, Brno University of Technology, Purkynova 656/123, 612 00 Brno, Czech Republic; [email protected].cz 2Faculty of Electrical Engineering and Communications, Brno University of Technology, Technicka 3082/12, 616 00 Brno, Czech Republic *Correspondence: adam.chr[email protected].cz Received: 19 October 2020; Accepted: 19 November 2020; Published: 20 November 2020   Abstract: The RoScan is a novel, high-accuracy multispectral surface scanning system producing colored 3D models that include a thermal layer. (1) Background: at present, medicine still exhibits a lack of objective diagnostic methods. As many diseases involve thermal changes, thermography may appear to be a convenient technique for the given purpose; however, there are three limiting problems: exact localization, resolution vs. range, and impossibility of quantification. (2) Methods: the basic principles and benefits of the system are described. The procedures rely on a robotic manipulator with multiple sensors to create a multispectral 3D model. Importantly, the structure is robust, scene-independent, and features quantifiable measurement uncertainty; thus, all of the above problems of medical thermography are resolved. (3) Results: the benefits were demonstrated by several pilot case studies: medicament efficacy assessment in dermatology, objective recovery progress assessment in traumatology, applied force quantification in forensic sciences, exact localization of the cause of pain in physiotherapy, objective assessment of atopic dermatitis, and soft tissue volumetric measurements. (4) Conclusion: the RoScan addresses medical quantification, which embodies a frequent problem in several medical sectors, and can deliver new, objective information to improve the quality of healthcare and to eliminate false diagnoses. Keywords: multimodal imaging; multispectral imaging; 3D thermography; high-accuracy 3D scanning; robotic 3D scanning 1. Introduction In general terms, the effectiveness of treatment, and thus also the recovery time and the amount of resources spent, stands on two prominent pillars: correct diagnosis and appropriate therapy [ 1 ]. At present, the latter pillar is comparatively well-characterized thanks to the availability of a wide range of knowledge sources generated through simplified global communication. Recommended therapies are often standardized, with the “best practice” methods systematically presented in research articles. Moreover, and importantly in the given context, almost any diagnosis can be paired with a suitable treatment approach [2]. The pace of progress in the former pillar, however, is markedly different. Although the correct diagnosis is at least as important as proper therapy, many of the current, commonly used diagnostic procedures still rely on healthcare professionals’ experience, intuition, and memory. The correctness of the decision then depends on the specific abilities of the medical specialist [ 3 ]. The diagnosis can be very close to the patient’s real condition, but it can also be completely wrong. The variance of the results then embodies a major problem within intuitive medicine. Sensors 2020,20, 6656; doi:10.3390/s20226656 www.mdpi.com/journal/sensors Sensors 2020,20, 6656 2 of 22 To prevent mistakes in determining the correct diagnosis, to decrease the required time, and to reduce the dependence of successful disease classification on the doctor’s expertise and experience, supporting diagnostic medical devices are introduced increasingly more often. Most hospitals are presently equipped with general-purpose imaging methods, such as X-ray, MRI, and CT; these tools, however, may be very expensive and thus less accessible to smaller medical centres. In such cases, smaller, cheaper, and more specialized instruments are employed. Almost all types of injuries, multiple diseases, and pathological changes share a significant symptom: the local temperature change. This thermal abnormality is caused by increased blood flow and stronger cellular metabolism in the affected area; the two aspects then induce a local temperature difference generated by the given negative health-related effect [ 4 ]. Such local thermal deviations can be detected and visualized via thermal cameras operating within the long-wave infrared spectrum (LWIR) [ 5 ]. As the resolution of thermal imagers is about 0.04 ◦ C [ 6 ], even the smallest changes are detectable, and the disease may be terminated before developing significantly enough to produce observable symptoms like redness or swelling (Figure 1). Sensors 2020, 20, x FOR PEER REVIEW 2 of 24 To prevent mistakes in determining the correct diagnosis, to decrease the required time, and to reduce the dependence of successful disease classification on the doctor’s expertise and experience, supporting diagnostic medical devices are introduced increasingly more often. Most hospitals are presently equipped with general-purpose imaging methods, such as X-ray, MRI, and CT; these tools, however, may be very expensive and thus less accessible to smaller medical centres. In such cases, smaller, cheaper, and more specialized instruments are employed. Almost all types of injuries, multiple diseases, and pathological changes share a significant symptom: the local temperature change. This thermal abnormality is caused by increased blood flow and stronger cellular metabolism in the affected area; the two aspects then induce a local temperature difference generated by the given negative health-related effect [4]. Such local thermal deviations can be detected and visualized via thermal cameras operating within the long-wave infrared spectrum (LWIR) [5]. As the resolution of thermal imagers is about 0.04 °C [6], even the smallest changes are detectable, and the disease may be terminated before developing significantly enough to produce observable symptoms like redness or swelling (Figure 1). Figure 1. The symptoms have not progressed far enough to be observable (left), but the inflammation is already clearly visible in the thermal image (right). Medical thermography is generally referred to as Digital Medical Thermal Imaging (DMTI); the technique currently finds use in several relevant applications (e.g., inflamed tissue analysis [7,8], cancer detection [9], blood perfusion [10], and human stress research [11]), although it has not been developed sufficiently to date. The main factors limiting the usability and popularity of this powerful imaging approach consist of three central drawbacks [12–16]: 1. Exact localization problem—due to the lack of easily recognizable and clearly bounded landmarks in the thermal image (which contains mostly smooth gradients instead of sharp edges), it is not possible to determine the exact position of a thermal deviation on the skin. This issue is illustrated in Figure 2, namely, the indicated spots with a locally lower temperature; however, we are presently unable to accurately assign the individual points to exact places on the body, and it is then impossible to distinguish whether such thermal footprints relate to, for example, an area of eczema or a birthmark in the vicinity. 2. Resolution vs. range problem—the low resolution of thermal imagers forces the user to search for a compromise between two options: (a) a wide-range view, where the entire part of the body is represented, but the resolution is too low to show enough details (Figure 3a); and (b) a detailed view providing sufficient resolution but containing a very narrow segment of the body, from which the context and exact position are difficult to understand (Figure 3b). As the thermogram analysis consists predominantly of comparing the temperature in body regions, the clinician must see a relatively large area of the body; simultaneously, the high resolution has to be preserved to deliver the required details. Overall, the resolution vs. range problems is one of the most limiting drawbacks restricting the development of DMTI. Figure 1. The symptoms have not progressed far enough to be observable ( left ), but the inflammation is already clearly visible in the thermal image (right). Medical thermography is generally referred to as Digital Medical Thermal Imaging (DMTI); the technique currently finds use in several relevant applications (e.g., inflamed tissue analysis [ 7 , 8 ], cancer detection [ 9 ], blood perfusion [ 10 ], and human stress research [ 11 ]), although it has not been developed sufficiently to date. The main factors limiting the usability and popularity of this powerful imaging approach consist of three central drawbacks [12–16]: 1. Exact localization problem—due to the lack of easily recognizable and clearly bounded landmarks in the thermal image (which contains mostly smooth gradients instead of sharp edges), it is not possible to determine the exact position of a thermal deviation on the skin. This issue is illustrated in Figure 2, namely, the indicated spots with a locally lower temperature; however, we are presently unable to accurately assign the individual points to exact places on the body, and it is then impossible to distinguish whether such thermal footprints relate to, for example, an area of eczema or a birthmark in the vicinity. 2. Resolution vs. range problem—the low resolution of thermal imagers forces the user to search for a compromise between two options: (a) a wide-range view, where the entire part of the body is represented, but the resolution is too low to show enough details (Figure 3a); and (b) a detailed view providing sufficient resolution but containing a very narrow segment of the body, from which the context and exact position are difficult to understand (Figure 3b). As the thermogram analysis consists predominantly of comparing the temperature in body regions, the clinician must see a relatively large area of the body; simultaneously, the high resolution has to be preserved to Sensors 2020,20, 6656 3 of 22 deliver the required details. Overall, the resolution vs. range problems is one of the most limiting drawbacks restricting the development of DMTI. 3. Qualitative only, no quantification—although 2D thermography basically constitutes a quantification tool (since it accurately measures the temperature in each pixel), in terms of the medical quantification of disease or injury it is a qualitative method only [ 13 , 17 ]. By using it, we can recognize the physiological and non-physiological regions of the body but remain unable to quantify them. According to [ 13 ], this issue stems from the following aspects: (1) changes in the thermal image are gradual and continuous, and thus the image contains very few clear landmarks (sharp edges or clear points), which are required for specific definition of the region of interest (ROI). The exactly bounded ROI embodies a necessary precondition to enable quantification within the impacted region. (2) The DMTI output is created by the lossy transformation of the real 3D world into a 2D projection; thus, the image is distorted. This scenario does not allow us to implement certain steps, for example, measuring the surface of the affected area: we can see only the projection, which is significantly influenced by the actual setup of the camera in relation to the scanned body. (3) In order to obtain two comparable thermograms, we have to exactly preserve the scene, including the exact position of the subject (as even tiny changes of the position may exert a substantial impact on the results). Such a condition is hardly achievable in currently used setups. The RoScan multispectral 3D scanning project, which is described in the following chapters, addresses and eliminates all of the three DMTI problems outlined above. Thus, the novel system upgrades DMTI to wide-range high-accuracy 3D thermography, which can be readily used as an objective quantitative tool in various medical subdomains [10]. At present, the actual medical quantification embodies a topic broadly discussed throughout the healthcare sector. Insufficiently accurate, inadequately sensitive, or excessively subjective diagnostic methods are perceived as very restrictive in several medical domains, including dermatology, traumatology, physiotherapy, and forensic sciences. Sensors 2020, 20, x FOR PEER REVIEW 3 of 24 3. Qualitative only, no quantification—although 2D thermography basically constitutes a quantification tool (since it accurately measures the temperature in each pixel), in terms of the medical quantification of disease or injury it is a qualitative method only [13,17]. By using it, we can recognize the physiological and non-physiological regions of the body but remain unable to quantify them. According to [13], this issue stems from the following aspects: (1) changes in the thermal image are gradual and continuous, and thus the image contains very few clear landmarks (sharp edges or clear points), which are required for specific definition of the region of interest (ROI). The exactly bounded ROI embodies a necessary precondition to enable quantification within the impacted region. (2) The DMTI output is created by the lossy transformation of the real 3D world into a 2D projection; thus, the image is distorted. This scenario does not allow us to implement certain steps, for example, measuring the surface of the affected area: we can see only the projection, which is significantly influenced by the actual setup of the camera in relation to the scanned body. (3) In order to obtain two comparable thermograms, we have to exactly preserve the scene, including the exact position of the subject (as even tiny changes of the position may exert a substantial impact on the results). Such a condition is hardly achievable in currently used setups. Figure 2. The thermal footprints cannot be clearly assigned to particular locations on the body. The lack of sharp edges and other spot landmarks makes exact localization in a single 2D thermal image impossible unless registration is performed with such a supplementary layer that yields clear features. Figure 2. The thermal footprints cannot be clearly assigned to particular locations on the body. The lack of sharp edges and other spot landmarks makes exact localization in a single 2D thermal image impossible unless registration is performed with such a supplementary layer that yields clear features. Sensors 2020,20, 6656 4 of 22 Sensors 2020, 20, x FOR PEER REVIEW 4 of 24 (a) (b) Figure 3. The impossibility of simultaneously reaching both a wide range and a high resolution restrains the use of thermal imaging in medicine: (a) in a wide-range view, the whole body part is visible, but the details of the temperature distribution are not; (b) in a narrow-range view, the details are visible, but the region cannot be localized, and context characterizing the whole affected area is missing. The RoScan multispectral 3D scanning project, which is described in the following chapters, addresses and eliminates all of the three DMTI problems outlined above. Thus, the novel system upgrades DMTI to wide-range high-accuracy 3D thermography, which can be readily used as an objective quantitative tool in various medical subdomains [10]. At present, the actual medical quantification embodies a topic broadly discussed throughout the healthcare sector. Insufficiently accurate, inadequately sensitive, or excessively subjective diagnostic methods are perceived as very restrictive in several medical domains, including dermatology, traumatology, physiotherapy, and forensic sciences. In dermatology, coarse and subjective scoring systems remain in wide use as the primary means of evaluation because an objective and reliable in vivo quantification method applicable to a particular region of interest has not been developed to date. Due to this deficiency, a diagnosis based on such subjective data is then unable to reveal tiny abnormalities to detect the disease in its early stage. Traumatology and forensic sciences would greatly welcome an objective method for quantifying and monitoring diverse effects or symptoms, such as bruises and their severity; however, this type of method is still missing even in the two fields. In physiotherapy, one of the common symptoms is a change in the body volume. These manifestations have to be measured objectively and sensitively in order to reveal an emerging disease already at the initial stage. Although the changes are currently measurable in a comparatively accurate manner, we are unable to distinguish between a physiological (e.g., muscle growth) and a non-physiological (e.g., swelling) change. Moreover, it is also difficult to assess the impact of the treatment procedures because the existing treatment evaluation exploits subjective health surveys or low-resolution scoring systems with poor inter-observer correlation. These problems (and many others) are tackled by the RoScan multispectral 3D scanning system. In this study, the basic principles and benefits of the approach are characterized and demonstrated on several medical applications where pilot case studies were performed with positive results. Figure 3. The impossibility of simultaneously reaching both a wide range and a high resolution restrains the use of thermal imaging in medicine: ( a ) in a wide-range view, the whole body part is visible, but the details of the temperature distribution are not; ( b ) in a narrow-range view, the details are visible, but the region cannot be localized, and context characterizing the whole affected area is missing. In dermatology, coarse and subjective scoring systems remain in wide use as the primary means of evaluation because an objective and reliable in vivo quantification method applicable to a particular region of interest has not been developed to date. Due to this deficiency, a diagnosis based on such subjective data is then unable to reveal tiny abnormalities to detect the disease in its early stage. Traumatology and forensic sciences would greatly welcome an objective method for quantifying and monitoring diverse effects or symptoms, such as bruises and their severity; however, this type of method is still missing even in the two fields. In physiotherapy, one of the common symptoms is a change in the body volume. These manifestations have to be measured objectively and sensitively in order to reveal an emerging disease already at the initial stage. Although the changes are currently measurable in a comparatively accurate manner, we are unable to distinguish between a physiological (e.g., muscle growth) and a non-physiological (e.g., swelling) change. Moreover, it is also difficult to assess the impact of the treatment procedures because the existing treatment evaluation exploits subjective health surveys or low-resolution scoring systems with poor inter-observer correlation. These problems (and many others) are tackled by the RoScan multispectral 3D scanning system. In this study, the basic principles and benefits of the approach are characterized and demonstrated on several medical applications where pilot case studies were performed with positive results. 2. Materials and Methods This paper discusses novel approaches to and applications within 3D thermal scanning. The presented scanning system, developed completely at Brno University of Technology, combines a high operational flexibility with high absolute accuracy of the resulting 3D model. Such a fusion of capabilities is still uncommon in current commercial 3D scanners, as these usually embody a compromise between the scanning accuracy and flexibility [ 18 ]; in medicine, however, both of these parameters are needed. A flexible and accurate 3D surface scanning system is thus very likely to find wide use in various medical domains, especially if it yields extensive 3D models and maintains the high accuracy across all the spectral layers, thus providing options not regularly available within the state of the art. The RoScan multispectral 3D system consists of a robotic manipulator and several sensors, grouped together at the endpoint into a sensing head (Figure 4a). To allow medical use, we combine a laser scanner, a color camera (wavelength range between 390 and 750 nm), and a thermal imager Sensors 2020,20, 6656 5 of 22 (wavelength range 8 to 14 µ m); however, the system is modular, meaning that the cameras can be replaced by other sensors providing data in various spectral ranges. Sensors 2020, 20, x FOR PEER REVIEW 5 of 24 2. Materials and Methods This paper discusses novel approaches to and applications within 3D thermal scanning. The presented scanning system, developed completely at Brno University of Technology, combines a high operational flexibility with high absolute accuracy of the resulting 3D model. Such a fusion of capabilities is still uncommon in current commercial 3D scanners, as these usually embody a compromise between the scanning accuracy and flexibility [18]; in medicine, however, both of these parameters are needed. A flexible and accurate 3D surface scanning system is thus very likely to find wide use in various medical domains, especially if it yields extensive 3D models and maintains the high accuracy across all the spectral layers, thus providing options not regularly available within the state of the art. The RoScan multispectral 3D system consists of a robotic manipulator and several sensors, grouped together at the endpoint into a sensing head (Figure 4a). To allow medical use, we combine a laser scanner, a color camera (wavelength range between 390 and 750 nm), and a thermal imager (wavelength range 8 to 14 µm); however, the system is modular, meaning that the cameras can be replaced by other sensors providing data in various spectral ranges. The manipulator moves the sensors along the body being scanned (Figure 4d), and the software processes, in real time, the distance data measured by the laser scanner into the form of a 3D surface model (Figure 4b). In addition to the laser scanner necessary for acquiring the 3D model itself, the robot’s endpoint is equipped with a color camera (Figure 4c), which delivers color information relating to each point of the 3D mesh; a thermal imager is also included (but will be discussed only later in the text). To this point, the RoScan’s output may have seemed to be almost the same as that of commercially available 3D scanners; such an impression, however, will change quickly if we emphasize the fact that the output model can be large and complex without any influence on the accuracy. Compared to other current 3D scanning instruments, the robotic manipulator ensures very accurate and independent localization of the sensors; the accuracy of the resulting model then does not affect the object’s appearance or shape. The method is robust, and the accuracy constraints are identical regardless of the scanned scene. Smooth homogeneous structures or repeated patterns are processed flawlessly (this being a substantial difference from the problematic procedure that characterizes, for example, methods based on stereophotogrammetric principles [19]), and the 3D model does not exhibit a cumulative deviation (unlike the products of, for example, handheld scanners [20] or other methods based on Iterative closest point (ICP) principle [21–23]). Figure 4. The RoScan system: ( a ) the main hardware components, namely, the robotic manipulator and head with several sensors; ( b ) the software yields a 3D mesh in real time; ( c ) the sensing head comprises a 2D laser scanner, a thermal imager, and an RGB camera; ( d ) the sensing head moves along the body, capturing thousands of distance profiles needed to create a 3D spatial model and taking several images from the thermal and visible spectra; the spectra are then mapped onto the surface. The manipulator moves the sensors along the body being scanned (Figure 4d), and the software processes, in real time, the distance data measured by the laser scanner into the form of a 3D surface model (Figure 4b). In addition to the laser scanner necessary for acquiring the 3D model itself, the robot’s endpoint is equipped with a color camera (Figure 4c), which delivers color information relating to each point of the 3D mesh; a thermal imager is also included (but will be discussed only later in the text). To this point, the RoScan’s output may have seemed to be almost the same as that of commercially available 3D scanners; such an impression, however, will change quickly if we emphasize the fact that the output model can be large and complex without any influence on the accuracy. Compared to other current 3D scanning instruments, the robotic manipulator ensures very accurate and independent localization of the sensors; the accuracy of the resulting model then does not affect the object’s appearance or shape. The method is robust, and the accuracy constraints are identical regardless of the scanned scene. Smooth homogeneous structures or repeated patterns are processed flawlessly (this being a substantial difference from the problematic procedure that characterizes, for example, methods based on stereophotogrammetric principles [ 19 ]), and the 3D model does not exhibit a cumulative deviation (unlike the products of, for example, handheld scanners [ 20 ] or other methods based on Iterative closest point (ICP) principle [21–23]). An additional information layer, unavailable in current 3D scanners, is generated by the thermal imager. The advantageously composed structure utilizes the robotic manipulator to capture the temperature images and 3D spatial data at different positions: The images are taken at a significantly greater proximity to compensate for the low resolution of thermal cameras and to achieve the same resolution in the thermogram, the other layers, and the entire 3D model. Consequently, a true, Sensors 2020,20, 6656 6 of 22 non-interpolated temperature value is assignable to each model point, even in extensive and highly detailed models. This capability then resolves and eliminates one of the problems that accompany the use of DMTI (see Sections 1and 2.4, issue No. 2, resolution vs. range problem). In the RoScan, the output is a multispectral 3D computer model providing a color shape coupled with the thermal distribution. The combination of 3D data, thermal data, and a color image addresses also the DMTI drawbacks Nos. 1 and 3, as further explained in Section 2.4. This paper briefly exposes the abilities, functions, and benefits of the entire RoScan system, focusing on its applicability; multiple details on the implementation are then outlined in other papers written by the authors of this article. The studies in question are dedicated to the actual 3D scanning [24–26], the mapping of 2D images onto the surface of the 3D model [ 27 ], and the laser scanner and camera calibration [28,29]. 2.1. Capturing 3D Models with the RoScan A multispectral 3D model is captured via the robotic manipulator moving the sensing head around the scanned object and along the predefined trajectory. The standard approach consists in defining different trajectories for each body part to be scanned; such predefined trajectories are simply selected according to the body part where a disease is expected. During the execution of a trajectory, the laser scanner captures the distance profiles, while the cameras collect the thermal and color images. The color images and distance profiles are obtained at similar proximities because both sensors exhibit an almost identical resolution rate. The resolution of the thermal imager is nevertheless significantly lower, making us capture (compared to the corresponding process performed with the color camera) markedly more thermal images from a significantly shorter distance. Thus, we finally obtain the same resolution in the color and the thermal layers (Figure 5). Sensors 2020, 20, x FOR PEER REVIEW 7 of 24 (a) (b) Figure 5. The procedures of reaching the same resolution rates in all of the spectral layers despite the native resolution of the sensors differ significantly: (a) hundreds of thermal images are captured at a very short distance, yielding a high resolution in a single image; the range, however, is narrow. Multiple images are taken and then merged by mapping onto the surface to form a single, highresolution layer covering the whole model; (b) the native resolution of the color camera is markedly higher, meaning that only several images taken at a greater distance are needed to reach the final resolution identical with that achieved in scenario (a). All of the collected distance profiles are aligned to the same system of coordinates by using homogeneous transformations; subsequently, the 3D shaded mesh is generated [25]. In the next step, the 2D images from the cameras are mapped to the surface of the 3D model [27]. At this stage, the ray-tracing algorithm [30] is employed to examine the visibility of each point of the mesh from the camera. If a single point is visible in multiple images, the resulting temperature is defined as the average of the values, as follows: the color in such points is computed as linear interpolation between the colors from these images, weighted by the angle relative to the 3D surface normal because the lightness of the color is influenced by the light reflection angle. The output model is delivered in a particular file format, which can be opened in the RoScan Analyzer software tool, described in the next section. 2.2. Tool to Allow 3D Model View and Analysis Multispectral 3D surface models created with the RoScan can be displayed via the RoScan Analyzer. In addition to the common features, such as rotating, scaling, and adjusting the modes of view, the software instrument also facilitates measuring many spatial parameters of the selected region of interest (ROI) and allows exporting the 3D data in several standard file formats (PTS, XYZ, and PLY); the data are then processable with any third-party 3D modeling program. Each model can be viewed in the following modes (Figure 6): • Color—the basic view of the clinician. In the model, this layer is used mainly for localization, either by natural landmarks of the skin (such as pigmented spots), via pen-marked markers, or through the use of region boundaries; such landmarks can be highlighted to be visible also in other layers (Figure 7). Alternatively, when highlighting abnormal places in the visualized thermal layer, we can adopt the opposite approach, which relies on localizing the spots in a color representation. Figure 5. The procedures of reaching the same resolution rates in all of the spectral layers despite the native resolution of the sensors differ significantly: ( a ) hundreds of thermal images are captured at a very short distance, yielding a high resolution in a single image; the range, however, is narrow. Multiple images are taken and then merged by mapping onto the surface to form a single, high-resolution layer covering the whole model; ( b ) the native resolution of the color camera is markedly higher, meaning that only several images taken at a greater distance are needed to reach the final resolution identical with that achieved in scenario (a). Sensors 2020,20, 6656 7 of 22 All of the collected distance profiles are aligned to the same system of coordinates by using homogeneous transformations; subsequently, the 3D shaded mesh is generated [25]. In the next step, the 2D images from the cameras are mapped to the surface of the 3D model [ 27 ]. At this stage, the ray-tracing algorithm [ 30 ] is employed to examine the visibility of each point of the mesh from the camera. If a single point is visible in multiple images, the resulting temperature is defined as the average of the values, as follows: the color in such points is computed as linear interpolation between the colors from these images, weighted by the angle relative to the 3D surface normal because the lightness of the color is influenced by the light reflection angle. The output model is delivered in a particular file format, which can be opened in the RoScan Analyzer software tool, described in the next section. 2.2. Tool to Allow 3D Model View and Analysis Multispectral 3D surface models created with the RoScan can be displayed via the RoScan Analyzer. In addition to the common features, such as rotating, scaling, and adjusting the modes of view, the software instrument also facilitates measuring many spatial parameters of the selected region of interest (ROI) and allows exporting the 3D data in several standard file formats (PTS, XYZ, and PLY); the data are then processable with any third-party 3D modeling program. Each model can be viewed in the following modes (Figure 6): • Color—the basic view of the clinician. In the model, this layer is used mainly for localization, either by natural landmarks of the skin (such as pigmented spots), via pen-marked markers, or through the use of region boundaries; such landmarks can be highlighted to be visible also in other layers (Figure 7). Alternatively, when highlighting abnormal places in the visualized thermal layer, we can adopt the opposite approach, which relies on localizing the spots in a color representation. • Temperature—the layer with false colors that correspond to the temperature of the skin; here, the exact value of a particular point can be examined by clicking. To reach the appropriate contrast between the physiological and the non-physiological areas, we can adjust the coloring palette by moving a slider. • Roughness—this layer shows false colors that correspond to the roughness of the scanned surface. This output utilizes the measuring principle of the applied laser scanner, which exploits triangulation [ 31 ]. As a side result of this measuring method, the divergence of the reflected beam is acquired; such divergence correlates with the roughness of the scanned surface. • Surface only—in certain cases, the coloring and shading of the model may hide some details or produce an optical illusion. Representing the surface in a single color allows us to see even the tiny details that normally often remain uncovered. When spatial details are to be examined, it is often convenient to show the 3D surface only, with the coloring eliminated. On each layer, the user can select points of interest (POIs), facilitating spatial measurements with the model. In the selected points or regions of interest (ROIs) defined thereby, the following parameters can be assessed: • Distance [mm]—separates several selected points. The distances are measured either in the direct mode or via the shortest path along the surface (these options are applicable in determining, e.g., the circumference of the ROI and the distance between the focus and the landmark). • Angle [degree]—the angle between every three subsequent POIs, including, for example, those that allow an analysis of vertebrae positions. • Surface area [mm 2 ]—relating to the whole model or the ROI only, for example, the area of burn, or extent of lesion. • Volume [mm 3 ]—concerning the entire model or a part thereof. Such a portion can be defined by the cutting plane or deflected cutting surface. Useful for, e.g., measurements of the limb volume. Sensors 2020,20, 6656 8 of 22 • Color—of a selected POI or the average color of the ROI. It is a more objective alternative to the currently used scoring systems for evaluating lesion redness. • Roughness—defined by the beam divergence width (in the sensor pixels), which is the dimensionless index correlating with the surface roughness at a particular point. We can either examine such roughness or calculate the average roughness of the ROI. • Temperature [ ◦ C]—of a highlighted point or the average temperature of the ROI. This parameter has the highest information value, since it correlates with inflammation in the area. Sensors 2020, 20, x FOR PEER REVIEW 8 of 24 • Temperature—the layer with false colors that correspond to the temperature of the skin; here, the exact value of a particular point can be examined by clicking. To reach the appropriate contrast between the physiological and the non-physiological areas, we can adjust the coloring palette by moving a slider. • Roughness—this layer shows false colors that correspond to the roughness of the scanned surface. This output utilizes the measuring principle of the applied laser scanner, which exploits triangulation [31]. As a side result of this measuring method, the divergence of the reflected beam is acquired; such divergence correlates with the roughness of the scanned surface. • Surface only—in certain cases, the coloring and shading of the model may hide some details or produce an optical illusion. Representing the surface in a single color allows us to see even the tiny details that normally often remain uncovered. When spatial details are to be examined, it is often convenient to show the 3D surface only, with the coloring eliminated. Figure 6. The output model of the RoScan system consists of several layers: (a) a 3D spatial representation without coloring; (b) a surface with true colors; (c) false-color-mapped thermal data; (d) false-color-mapped roughness data. On each layer, the user can select points of interest (POIs), facilitating spatial measurements with the model. In the selected points or regions of interest (ROIs) defined thereby, the following parameters can be assessed: • Distance [mm]—separates several selected points. The distances are measured either in the direct mode or via the shortest path along the surface (these options are applicable in determining, e.g., the circumference of the ROI and the distance between the focus and the landmark). • Angle [degree]—the angle between every three subsequent POIs, including, for example, those that allow an analysis of vertebrae positions. • Surface area [mm2]—relating to the whole model or the ROI only, for example, the area of burn, or extent of lesion. • Volume [mm3]—concerning the entire model or a part thereof. Such a portion can be defined by the cutting plane or deflected cutting surface. Useful for, e.g., measurements of the limb volume. Figure 6. The output model of the RoScan system consists of several layers: ( a ) a 3D spatial representation without coloring; ( b ) a surface with true colors; ( c ) false-color-mapped thermal data; ( d ) false-color-mapped roughness data. Sensors 2020, 20, x FOR PEER REVIEW 9 of 24 • Color—of a selected POI or the average color of the ROI. It is a more objective alternative to the currently used scoring systems for evaluating lesion redness. • Roughness—defined by the beam divergence width (in the sensor pixels), which is the dimensionless index correlating with the surface roughness at a particular point. We can either examine such roughness or calculate the average roughness of the ROI. • Temperature [°C]—of a highlighted point or the average temperature of the ROI. This parameter has the highest information value, since it correlates with inflammation in the area. (a) (b) Figure 7. The navigation between several layers carried out by using points of interest (POIs): (a) selecting the area with suspicious features in the color layer; (b) examining the temperature at highlighted points in the thermal layer. In this case, we did not observe any thermal changes relating to the effect indicated in the image (a); the observation corresponds to reality, as the injury was created by a make-up artist. Such a multimodal 3D computer model finds use in, for example, inflammation monitoring. The process then comprises the following stages: first, the thermal layer is studied, and possible thermal abnormalities are highlighted in the 3D model. Second, we switch to the color mode, which facilitates exact localization of the inflammation foci and allows us to set up the region of interest (ROI). Afterwards, we can quantify the inflammation activity (usually in the form of the mean temperature within the ROI). At the last stage, the swelling-induced volume variation is measured accurately. The inflammation monitoring and quantification constitute only one of the RoScan’s potential target applications. To illustrate the claim, we can note in this context that while inflammation leads to a positive thermal data gradient, ischemia results in a local decrease; thus, the RoScan may be employed in the monitoring of diabetic necrotic tissues. 2.3. Key Parameters of the RoScan The central advantage of the discussed concept rests in the applied combination of spatial, color, and thermal data within a single 3D model; such an approach delivers new research opportunities, which would otherwise remain virtually impossible to implement. More concretely, the RoScan has a major potential in the objective evaluation of spatial changes of the body; this medical task includes, above all, the monitoring of oedemas, muscle growth and atrophy, inflammation, and necrosis. The color layer guarantees the precise selection of a ROI (via natural landmarks or markers drawn on the subject’s skin). The thermal layer helps to find inflamed areas or determine if a spatial change is caused by physiological or non-physiological processes. Finally, 3D scanning yields an undistorted view, thus offering an effective alternative to standard 2D imaging technologies, where the distortion prevents objective measurements from being performed. The essential capabilities and properties of the RoScan system are summarized in Table 1. Figure 7. The navigation between several layers carried out by using points of interest (POIs): ( a ) selecting the area with suspicious features in the color layer; ( b ) examining the temperature at highlighted points in the thermal layer. In this case, we did not observe any thermal changes relating to the effect indicated in the image ( a ); the observation corresponds to reality, as the injury was created by a make-up artist. Sensors 2020,20, 6656 9 of 22 Such a multimodal 3D computer model finds use in, for example, inflammation monitoring. The process then comprises the following stages: first, the thermal layer is studied, and possible thermal abnormalities are highlighted in the 3D model. Second, we switch to the color mode, which facilitates exact localization of the inflammation foci and allows us to set up the region of interest (ROI). Afterwards, we can quantify the inflammation activity (usually in the form of the mean temperature within the ROI). At the last stage, the swelling-induced volume variation is measured accurately. The inflammation monitoring and quantification constitute only one of the RoScan’s potential target applications. To illustrate the claim, we can note in this context that while inflammation leads to a positive thermal data gradient, ischemia results in a local decrease; thus, the RoScan may be employed in the monitoring of diabetic necrotic tissues. 2.3. Key Parameters of the RoScan The central advantage of the discussed concept rests in the applied combination of spatial, color, and thermal data within a single 3D model; such an approach delivers new research opportunities, which would otherwise remain virtually impossible to implement. More concretely, the RoScan has a major potential in the objective evaluation of spatial changes of the body; this medical task includes, above all, the monitoring of oedemas, muscle growth and atrophy, inflammation, and necrosis. The color layer guarantees the precise selection of a ROI (via natural landmarks or markers drawn on the subject’s skin). The thermal layer helps to find inflamed areas or determine if a spatial change is caused by physiological or non-physiological processes. Finally, 3D scanning yields an undistorted view, thus offering an effective alternative to standard 2D imaging technologies, where the distortion prevents objective measurements from being performed. The essential capabilities and properties of the RoScan system are summarized in Table 1. Table 1. The key parameters. Parameter Value Positional accuracy of model points (3 σ). 0.12 mm Scanning speed 10 mm/s Scanning trajectory flexible, programmable Measured values position, color, temperature, roughness Operational cost approximately 0.1 EUR per model Price of a device approximately 30,000 EUR (or less) Data storage format internal Export formats for other software PLY, PTS, XYZ The price of the device indicated in the table above relates to the research setup characterized herein; that is, a configuration which utilizes a robotic manipulator. Advantageously, this component delivers superior flexibility and versatility, allowing the entire device to find use in many different applications, even outside of medicine. If deployed in practice, however, the system is usually not required to perform an overly wide range of activities, and thus its six-degrees-of-freedom flexibility can become overkill. The manipulator can then be replaced by a cheaper (but still very accurate) positioning system with fewer degrees of freedom (Rails, Cartesian robots, etc.), maintaining the key parameters. As the price of the manipulator amounts to 19,000 EUR out of the total of 30,000 EUR, the acquisition costs could then drop markedly. Although the purchase costs are significantly lower than those in the current imaging methods, the most extensive savings are achieved in regular operation, where the RoScan generates almost zero additional costs, regardless of the frequency of use. This is a major benefit compared to, for example, MRI, in which each scan is worth 200–400 EUR [32,33]. Sensors 2020,20, 6656 16 of 22 days, when the condition was not yet evaluable by means of the common methods (which mostly exploit visual observation and manual examination). Sensors 2020, 20, x FOR PEER REVIEW 17 of 24 for example, we can refer to the evidence of inflammation in a subject suffering from entezopathy of the deltoid muscle, revealed during the first examination (Figure 14a). The inflammation had recessed markedly before the second examination (Figure 14b), and improved blood flow in the patella area, accompanying the post-surgical recovery between the first (Figure 14c) and the second examinations (Figure 14d), is also evident. Positive progress was determined from the 3D models already after two days, when the condition was not yet evaluable by means of the common methods (which mostly exploit visual observation and manual examination). The benefit of the RoScan in physiotherapy rests mainly in the exact localization capability; however, the device can also function as a quantification tool in cases where it is possible to suppress the external fluctuations that affect body temperature. Such a situation is presented in Figure 14a,b; the images show an area with a homogeneous temperature distribution where the external influences can be suppressed via comparison with a reference point nearby. In item 14ac, such normalization is difficult to achieve, and the user will thus benefit from the above-mentioned localization only. Figure 14. An upper limb affected by entezopathy of the deltoid muscle at (a) the first visit and (b) the second visit; the healing of the inflammation is clearly observable. The recovery following a surgery of ruptured knee ligaments at (c) the first visit and (d) the second visit; improved blood circulation is visible. Note: the regular hot spots in (a,b) are caused by cellulite. 3.5. Volumetric Measurements of Soft Tissues Within this experiment, the current volumetric techniques were compared with the RoScan via measuring the volumes of reference objects (phantoms). When performing against the current gold standard, i.e., water displacement volumetry, the RoScan delivered volumetric measurements at higher accuracy and repeatability, as shown in Figure 15. Figure 14. An upper limb affected by entezopathy of the deltoid muscle at ( a ) the first visit and ( b ) the second visit; the healing of the inflammation is clearly observable. The recovery following a surgery of ruptured knee ligaments at ( c ) the first visit and ( d ) the second visit; improved blood circulation is visible. Note: the regular hot spots in (a,b) are caused by cellulite. The benefit of the RoScan in physiotherapy rests mainly in the exact localization capability; however, the device can also function as a quantification tool in cases where it is possible to suppress the external fluctuations that affect body temperature. Such a situation is presented in Figure 14a,b; the images show an area with a homogeneous temperature distribution where the external influences can be suppressed via comparison with a reference point nearby. In item 14ac, such normalization is difficult to achieve, and the user will thus benefit from the above-mentioned localization only. 3.5. Volumetric Measurements of Soft Tissues Within this experiment, the current volumetric techniques were compared with the RoScan via measuring the volumes of reference objects (phantoms). When performing against the current gold standard, i.e., water displacement volumetry, the RoScan delivered volumetric measurements at higher accuracy and repeatability, as shown in Figure 15. In terms of its parameters and capabilities, the RoScan embodies a faster, safer, and more accurate approach with very low operational costs; the system is especially valuable in specific cases where advanced ROI examination and high accuracy are needed. Such special volumetric cycles normally require the standard MRI or CT; however, the RoScan, if employed, reaches more accurate results at significantly lower costs [38]. As the RoScan is generally the most accurate method, it will allow measuring of even very small changes in the volume, which can enable us to discover pathological changes in a timely manner, well before they have developed significantly. Additional details on the experiment can be found in source [ 12 ], and the current volumetric methods are compared between one another and with the RoScan in [39] and [38,40], respectively. Sensors 2020,20, 6656 17 of 22 Sensors 2020, 20, x FOR PEER REVIEW 18 of 24 Figure 15. The RoScan compared with the circumferential techniques and the water displacement method implemented in a reference object having a known volume: the RoScan displayed the most accurate and repeatable performance [12]. In terms of its parameters and capabilities, the RoScan embodies a faster, safer, and more accurate approach with very low operational costs; the system is especially valuable in specific cases where advanced ROI examination and high accuracy are needed. Such special volumetric cycles normally require the standard MRI or CT; however, the RoScan, if employed, reaches more accurate results at significantly lower costs [38]. As the RoScan is generally the most accurate method, it will allow measuring of even very small changes in the volume, which can enable us to discover pathological changes in a timely manner, well before they have developed significantly. Additional details on the experiment can be found in source [12], and the current volumetric methods are compared between one another and with the RoScan in [39] and [38,40], respectively. 3.6. Objective Assessment of Atopic Dermatitis There are many diagnostic methods for assessing atopic dermatitis (AD), varying from scoring systems (very subjective, insensitive, and providing low resolution) to sophisticated and expensive options [41]. In this context, DMTI is considered a relatively objective (and thus a potentially reliable) method, as it measures increased blood flow, which manifests itself almost identically in all patients [42]. The technique, however, is rarely used due to the low resolution and complicated standardization of the data that facilitate comparison between the results (the DMTI exact localization problem). As mentioned above, the RoScan’s structure eliminates this difficulty, and thus it finds use in AD lesion assessment. The capability was demonstrated in the experiment, whereby an AD patient was examined via DMTI; nevertheless, the thermal data could not be causally linked to the real lesions. The relationship was then established only by using the RoScan and its ability to select the nonphysiological locations and to navigate in the color layer (Figure 8). The thermal information was then properly related to the AD lesions, and the progress was quantified to meet the goal of the experiment. Figure 15. The RoScan compared with the circumferential techniques and the water displacement method implemented in a reference object having a known volume: the RoScan displayed the most accurate and repeatable performance [12]. 3.6. Objective Assessment of Atopic Dermatitis There are many diagnostic methods for assessing atopic dermatitis (AD), varying from scoring systems (very subjective, insensitive, and providing low resolution) to sophisticated and expensive options [41]. In this context, DMTI is considered a relatively objective (and thus a potentially reliable) method, as it measures increased blood flow, which manifests itself almost identically in all patients [ 42 ]. The technique, however, is rarely used due to the low resolution and complicated standardization of the data that facilitate comparison between the results (the DMTI exact localization problem). As mentioned above, the RoScan’s structure eliminates this difficulty, and thus it finds use in AD lesion assessment. The capability was demonstrated in the experiment, whereby an AD patient was examined via DMTI; nevertheless, the thermal data could not be causally linked to the real lesions. The relationship was then established only by using the RoScan and its ability to select the nonphysiological locations and to navigate in the color layer (Figure 8). The thermal information was then properly related to the AD lesions, and the progress was quantified to meet the goal of the experiment. 4. Discussion The RoScan project, completely developed at Brno University of Technology, embodies a novel concept of using a robotic manipulator combined with several sensors producing data in different spectra. For medical applications, it appears to be advantageous to combine a laser scanner, a color camera (wavelength range between 390 and 750 nm), and a thermal imager (wavelength range from 8 to 14 µ m); such a configuration is beneficial because the thermal changes commonly accompany the non-physiological states. This innovative method for multispectral 3D scanning brings several advantages compared to the state-of-the-art approaches. Thanks to the projection of the spectral data onto the surface of the 3D model via an independent localization system, the process of registering the 2D images is robust and independent of the scanned object. The procedure is applicable to any smooth and homogeneous surface or texture, thus being especially valuable in merging several thermal images into a large one if there is a lack of reliably detectable registration landmarks. Sensors 2020,20, 6656 18 of 22 Another major benefit of independent localization lies in cross-spectral image registration (color vs. thermal). With this method, one can reach the same resolution in both the thermal and the color layers, even if the native resolutions of the sensor differ significantly. Moreover, both layers are properly aligned to the 3D model and to each other; importantly, the uncertainty is exactly specified and equal in the entire dataset. Advancement beyond the possibilities of the state-of-the-art is recognizable also within 3D scanning; unlike many of the current, commercially available 3D methods, the RoScan allows each point of the captured model to exhibit an equal, absolute spatial accuracy independent of the scanned scene and lacking cumulative deviations. Additionally, the accuracy is precisely quantifiable from the parameters of the manipulator and sensor, bringing a major advantage in precise measurements, where the uncertainty must be specified in an exact manner. Most of the current 3D scanners are not capable of such operations [43,44]. The RoScan offers a considerable potential in those medical applications where objective quantification plays the key role. The system quantifies inflammatory processes inside the human body; several pilot case studies described in this paper showed that using the RoScan to quantify inflammation is feasible when the inflamed region remains close to the skin, has sufficient “strength” to influence the skin temperature, and ideally appears in an area where the temperature distribution is as homogeneous as possible, enabling us to find an index with the potential to eliminate the environmental factors. The RoScan-based assessment of medicament efficacy in dermatology showed the possibility of comparing the impacts of two different drugs on the same subject and even the same lesion, an objective virtually unachievable via the current methods. Standard clinical tests are usually carried out on two groups of subjects, of which one is treated with the tested medicament, while the other is not; such an approach then may lead to wrong conclusions if the selected group does not constitute a representative sample of the population, and its sensitivity is rather low (meaning that only major progress is distinguishable). A traumatology-related study discussing the assessment of objective progress in recovery pointed out that the RoScan enables early detection of inflamed regions before they start to give pain or develop significantly, stressing also the facts that the system is thus capable of reducing the treatment and recovery time and cost. Moreover, the technique embodies a valuable tool to yield an objective record of the current state of a disease; producing such records is a common problem of present medicine, which intensively relies on the personal memory of the clinician. Another research subdomain where the RoScan reaches beyond the state of the art lies in investigating the quantification of applied force: the system proved to be highly sensitive, with the traces of a physical assault remaining visible on the thermogram for more than 3 h after the violent act, even though the actual pain caused by the punching disappeared in only 10 min. The measured temperature correlated with the number of punches received. In physiotherapy, the RoScan delivered objective evaluation of the treatment effects, documenting objectively blood circulation improvement after an invasive surgery. Finally, the tool has been verified as suitable for soft tissue volumetry, bringing superior accuracy and repeatability. An additional benefit is the possibility to select the edema only, stemming from the ability of the thermal layer to distinguish the edema from a healthy tissue. Neither of the achievements is obtainable via the current volumetric methods. At this point, however, it is necessary to emphasize that 3D thermography remains a mere tool requiring expert assessment by a doctor. The RoScan does not aim to offer automatic detection and localization of inflammation; rather, the device is designed to embody a source of objective data that will allow clinicians to decide and quantify better and more correctly than if such data are lacking. Areas with higher temperatures are not always inflamed; often the effect expresses an anatomical constitution (e.g., when the arteries lie closer to the skin). A significant impact is generated also by Sensors 2020,20, 6656 19 of 22 external influences, including the ambient temperature, the patient’s previous activity, and current footwear or clothing. Although the effect of these elements can be limited in many cases by choosing a suitable relative index, this approach is not applicable generally, meaning that the 3D thermogram will still have to be interpreted and evaluated by a medical doctor, who nevertheless may not be as experienced as a physician diagnosing without the thermogram. 5. Conclusions Generally, precise 3D thermography tools, such as the RoScan, can yield major benefits in various healthcare domains. The relevant instruments facilitate early detection of inflamed areas, the reason being that their sensitivities surpass those of the commonly used methods, which are mostly based on visual observations [45–47]. The precise 3D thermography as performed by the RoScan addresses three main limitations of Digital Medical Thermal Imaging (DMTI), discussed in the present scientific literature: the exact localization problem, the resolution vs. range problem, and quantification impossibility. In this context, DMTI has been characterized as having a substantial potential if the drawbacks were resolved. The RoScan system eliminates all of these issues, thus opening new possibilities for thermal imaging in medicine. Several pilot case studies confirmed the benefits introduced by the RoScan into diverse subdomains of healthcare, where the advanced concept overcomes the limits inherent in the other state of-the-art methods. This scanning approach appears to be particularly beneficial within medical efficacy assessment, dermatology, traumatology, and physiotherapy, where it facilitates objective monitoring of the recovery progress. In all of these areas, the novel multispectral 3D method can be advantageously adopted as an objective quantification tool to fill the void in the set of options available for the given purpose. More concretely, the RoScan is especially useful when the common symptoms have not yet fully developed; however, multispectral 3D scanning in general finds application also when the symptoms are already visible. Importantly, the capability of preserving the exact state of the patient’s body (to allows comparison during the next visit at the clinician) leads to objective evaluation of the progress of the disease. The positive results of the initial case studies presented herein pointed out the areas where the opportunity to enhance the state of the art is potentially high. In future projects, we will focus predominantly on these domains, and we will attempt to transfer the proposed technology into everyday medical practice. Although it will probably be necessary to modify the system for the given applications, this task will not be problematic, especially due to the fact that the entire system was fully developed at our department and is thus completely controlled by the authors of the paper. Major portions of the research will be dedicated to the maximum simplification and automation of the scanning process because if the system is to be practically usable, it must offer easy operability without expert knowledge and significant time intensity. The system is convenient for not only medicine but also technology and science: thermal 3D scanning procedures are usable in, for example, electrical engineering to ensure thermal analyses centered on the efficiency of electrical components [ 48 ]; civil engineering to monitor thermal loss in materials [ 49 ]; mechanical engineering to investigate mechanical strain [ 50 ]; and rescue robotics or experimental biology to evaluate the indispensable life functions [51–53]. Moreover, if we also take into account the fact all algorithms are universal and thus make it possible to easily replace the color and thermal camera with another sensor of a different spectral range, the usability of the RoScan system will expand even more significantly. Author Contributions: Concept—L.Z. and A.C.; methodology—A.C.; software—A.C.; validation—A.C. and L.Z.; resources—L.Z.; data curation—A.C.; writing/original draft structuring—A.C.; writing—review and editing—A.C.; visualization—A.C.; supervision—L.Z.; project administration—A.C.; funding acquisition—L.Z. All authors have read and agreed to the published version of the manuscript. Sensors 2020,20, 6656 20 of 22 Funding: This research was funded by the grant “Advancing Smart Optical Imaging and Sensing for Health (ASTONISH)”, H2020-EU.2.1.1.7.—ECSEL programme, topic ECSEL-04-2015—Smart Health, grant agreement ID: 692470. The research has been financially supported by the Ministry of Education, Youth and Sports of the Czech Republic under the project CEITEC 2020 (LQ1601). The completion of this paper was made possible by the grant No. FEKT-S-20-6205—“Research in Automation, Cybernetics and Artificial Intelligence within Industry 4.0” financially supported by the Internal science fund of Brno University of Technology. Acknowledgments: The authors would like to thank all the patients who volunteered to participate in the experiments which involved 3D thermal scanning of their body parts. The experiments were performed in compliance with the Declaration of Helsinki, i.e., the participants agreed to the form and content of the experiment appointments. Conflicts of Interest: The authors declare no conflict of interest. The funding institutions had no role in designing the study; in collecting, analyzing, and interpreting the data; in writing the manuscript; or in the decision to publish the results. References 1. Papadakis, M.A.; McPhee, S.J.; Rabow, M.W. CURRENT Medical Diagnosis and Treatment 2019; McGraw Hill Professional: New York, NY, USA, 2019; ISBN 978-1-260-11744-8. 2. New, A. Evidence-based medicine. A new approach to teaching the practice of medicine. Evidence-Based Medicine Working Group. JAMA J. Am. Med. Assoc. 1992,268, 2420–2425. [CrossRef] 3. Sackett, D.L. Evidence-based medicine. Semin. Perinatol. 1997,21, 3–5. [CrossRef] 4. Chang, T.-C.; Hsiao, Y.-L.; Liao, S.-L. Application of digital infrared thermal imaging in determining inflammatory state and follow-up effect of methylprednisolone pulse therapy in patients with Graves’ ophthalmopathy. Graefes Arch. Clin. Exp. Ophthalmol. 2008,246, 45–49. [CrossRef] 5. Gabrlik, P.; la Cour-Harbo, A.; Kalvodova, P.; Zalud, L.; Janata, P. Calibration and accuracy assessment in a direct georeferencing system for UAS photogrammetry. Int. J. Remote Sens. 2018 ,39, 4931–4959. [CrossRef] 6. Coffey, V.C. Multispectral Imaging Moves into the Mainstream. Opt. Photonics News 2012 ,23, 18–24. [CrossRef] 7. Hilton-Jones, D. Diagnosis and Treatment of Inflammatory Muscle Diseases. J. Neurol. Neurosurg. Psychiatry 2003,74, ii25–ii31. [CrossRef] 8. Ring, E.F.J.; Ammer, K. Infrared thermal imaging in medicine. Physiol. Meas. 2012,33, R33. [CrossRef] 9. Lu, G.; Fei, B. Medical hyperspectral imaging: A review. J. Biomed. Opt. 2014,19, 010901. [CrossRef] 10. Aretusi, G.; Fontanella, L.; Ippoliti, L.; Merla, A. Supervised Classification of Thermal High-Resolution IR Images for the Diagnosis of Raynaud’s Phenomenon. In New Perspectives in Statistical Modeling and Data Analysis; Ingrassia, S., Rocci, R., Vichi, M., Eds.; Springer: Berlin/Heidelberg, Germany, 2011; pp. 419–427. 11. Engert, V.; Merla, A.; Grant, J.A.; Cardone, D.; Tusche, A.; Singer, T. Exploring the Use of Thermal Infrared Imaging in Human Stress Research. PLoS ONE 2014,9, e90782. [CrossRef] 12. Chromy, A. High Resolution Multispectral 3D Scanning and Its Medical Applications. Ph.D. Thesis, Department of Control and Instrumentation, Faculty of Electrical Engineering and Communication, Brno University of Technology, Brno, Czech Republic, 2017. 13. Vardasca, R.; Simoes, R. Current Issues in Medical Thermography. In Topics in Medical Image Processing and Computational Vision; Lecture Notes in Computational Vision and Biomechanics; Springer: Dordrecht, The Netherlands, 2013; pp. 223–237. ISBN 978-94-007-0725-2. 14. Gul, K.M.; Ahmadi, N.; Wang, Z.; Jamieson, C.; Nasir, K.; Metcalfe, R.; Hecht, H.S.; Hartley, C.J.; Naghavi, M. Digital thermal monitoring of vascular function: A novel tool to improve cardiovascular risk assessment. Vasc. Med. Lond. Engl. 2009,14, 143–148. [CrossRef] 15. Bruehl, S.; Lubenow, T.R.; Nath, H.; Ivankovich, O. Validation of thermography in the diagnosis of reflex sympathetic dystrophy. Clin. J. Pain 1996,12, 316–325. [CrossRef] [PubMed] 16. Di Carlo, A. Thermography and the possibilities for its applications in clinical and experimental dermatology. Clin. Dermatol. 1995,13, 329–336. [CrossRef] 17. Ju, X.; Nebel, J.-C.; Siebert, J.P. 3D thermography imaging standardization technique for inflammation diagnosis. In Infrared Components and Their Applications; Gong, H., Cai, Y., Chatard, J.-P., Eds.; International Society for Optics and Photonics: Beijing, China, 2005; Volume 5640, p. 266. 18. Curless, B. From Range Scans to 3D Models. SIGGRAPH Comput. Graph. 1999,33, 38–41. [CrossRef] 19. Brown, L.G. A Survey of Image Registration Techniques. ACM Comput. Surv. 1992 ,24, 325–376. [CrossRef] Sensors 2020,20, 6656 21 of 22 20. Strobl, K.H.; Mair, E.; Hirzinger, G. Image-based pose estimation for 3-D modeling in rapid, hand-held motion. In Proceedings of the 2011 IEEE International Conference on Robotics and Automation (ICRA), Shanghai, China, 9–13 May 2011; pp. 2593–2600. 21. Alshawa, M. ICL: Iterative closest line—A novel point cloud registration algorithm based on linear features. Ekscentar 2007,10. 22. Poppinga, J.; Vaskevicius, N.; Birk, A.; Pathak, K. Fast plane detection and polygonalization in noisy 3D range images. In Proceedings of the IEEE/RSJ International Conference on Intelligent Robots and Systems, IROS 2008, Nice, France, 22–26 September 2008; pp. 3378–3383. 23. Bellocchio, F.; Borghese, N.A.; Ferrari, S.; Piuri, V. 3D Surface Reconstruction; Springer: New York, NY, USA, 2013; ISBN 978-1-4614-5631-5. 24. Chromy, A.; Zalud, L. Novel 3D Modelling System Capturing Objects with Sub-Millimetre Resolution. Adv. Electr. Electron. Eng. 2014,12, 476–487. [CrossRef] 25. Chromy, A.; Zalud, L. Robotic 3D scanner as an alternative to standard modalities of medical imaging. SpringerPlus 2014,3, 13. [CrossRef] 26. Chromy, A.; Kocmanova, P.; Zalud, L. Creating Three-Dimensional Computer Models Using Robotic Manipulator and Laser Scanners. In Proceedings of the 12th IFAC Conference on Programmable Devices and Embedded Systems (2013), Velke Karlovice, Czech Republic, 25–27 September 2013; Elsevier B.V.: Velke Karlovice, Czech Republic, 2013; pp. 268–273. 27. Chromy, A.; Klima, O. A 3D Scan Model and Thermal Image Data Fusion Algorithms for 3D Thermography in Medicine. J. Healthc. Eng. 2017,2017, 5134021. [CrossRef] 28. Kocmanova, P.; Zalud, L.; Chromy, A. 3D proximity laser scanner calibration. In Proceedings of the 2013 18th International Conference on Methods and Models in Automation and Robotics (MMAR), Miedzyzdroje, Poland, 26–29 August 2013; pp. 742–747. 29. Chromy, A. Mutual Calibration of Sensors for Multispectral 3D Scanning of Surface. In Proceedings of the 9th International Congress on Ultra Modern Telecommunications and Control Systems, Munich, Germany, 6–8 November 2017; IEEE: Munich, Germany, 2017; pp. 199–204. 30. Suffern, K. Ray Tracing from the Ground Up; CRC Press: Boca Raton, FL, USA, 2016; ISBN 978-1-4987-7470-3. 31. Smith, K.B.; Zheng, Y.F. Accuracy Analysis of Point Laser Triangulation Probes Using Simulation. J. Manuf. Sci. Eng. 1998,120, 736–745. [CrossRef] 32. Magnetic Resonance Imaging. Available online: http://www.mri-portal.com/clanky/magneticka_rezonance. php (accessed on 22 April 2018). 33. Novelline, R.A.; Squire, L.F. Squire’s Fundamentals of Radiology, 6th ed.; Harvard University Press: Cambridge, MA, USA, 2004; ISBN 978-0-674-01279-0. 34. Chromy, A. Multispectral 3D Surface Scanning System RoScan and its Application in Inflammation Monitoring and Quantification. In Proceedings of the 11th International Joint Conference on Biomedical Engineering Systems and Technologies: BIOIMAGING, Madeira, Portugal, 19–21 January 2018; Volume 2, pp. 106–113. 35. Chromy, A.; Zalud, L. Three-dimensional Thermal Imaging in Medicine. In Proceedings of the 3DBODY.TECH 2017—8th International Conference and Exhibition on 3D Body Scanning and Processing Technologies, Montreal, QC, Canada, 11–12 October 2017; Hometrica Consulting—Dr. Nicola D’Apuzzo: Montreal, QC, Canada, 2017; pp. 232–238. 36. V é le, F. Funkˇcn í diagnostika—Pˇredpoklad ú spˇechu fyzioterapeuta. Rehabil. Fyzik á ln í L é kaˇrstv í2012 ,19, 155–158. 37. Vele, F. Vysetreni Hybnych Funkci z Pohledu Neurofyziologie: Prirucka pro Terapeuty Pracujici v Neurorehabilitaci; Triton: Praha, Czech Republic, 2012; ISBN 978-80-7387-608-1. 38. Chromy, A. Application of High-Resolution 3D Scanning in Medical Volumetry. Int. J. Electron. Telecommun. 2016,62, 23–31. [CrossRef] 39. Chromy, A.; Zalud, L.; Dobsak, P.; Suskevic, I.; Mrkvicova, V. Limb volume measurements: Comparison of accuracy and decisive parameters of the most used present methods. SpringerPlus 2015 ,4, 707. [CrossRef] [PubMed] 40. Chromy, A. High-Accuracy Volumetric Measurements of Soft Tissues using Robotic 3D Scanner. IFAC-Pap 2015,48, 318–323. [CrossRef] Sensors 2020,20, 6656 22 of 22 41. Sprikkelman, A.B.; Tupker, R.A.; Burgerhof, H.; Schouten, J.P.; Brand, P.L.; Heymans, H.S.; van Aalderen, W.M. Severity scoring of atopic dermatitis: A comparison of three scoring systems. Allergy 1997 ,52, 944–949. [CrossRef] [PubMed] 42. Johansen, J.D.; Frosch, P.J.; Lepoittevin, J.-P. Contact Dermatitis; Springer Science & Business Media: Berlin/Heidelberg, Germany, 2010; ISBN 978-3-642-03827-3. 43. Polo, M.-E.; Felic í simo, Á .M. Analysis of Uncertainty and Repeatability of a Low-Cost 3D Laser Scanner. Sensors 2012,12, 9046–9054. [CrossRef] [PubMed] 44. Beraldin, J.-A. Basic theory on surface measurement uncertainty of 3D imaging systems. Three-Dimens. Imaging Metrol. 2009,7239. [CrossRef] 45. Lipsky, B.A.; Berendt, A.R.; Deery, H.G.; Embil, J.M.; Joseph, W.S.; Karchmer, A.W.; LeFrock, J.L.; Lew, D.P.; Mader, J.T.; Norden, C.; et al. Diagnosis and Treatment of Diabetic Foot Infections. Clin. Infect. Dis. 2004 ,39, 885–910. [CrossRef] 46. Krysko, D.V.; Vanden Berghe, T.; D’Herde, K.; Vandenabeele, P. Apoptosis and necrosis: Detection, discrimination and phagocytosis. Methods 2008,44, 205–221. [CrossRef] 47. Merla, A.; Romani, G.L. Biomedical Applications of Functional Infrared Imaging. In Proceedings of the 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference, Shanghai, China, 17–18 January 2006; pp. 690–693. 48. Merryman, S.A.; Nelms, R.M. Diagnostic technique for power systems utilizing infrared thermal imaging. IEEE Trans. Ind. Electron. 1995,42, 615–628. [CrossRef] 49. Taileb, A.; Dekkiche, H. Infrared Imaging as a Means of Analyzing and Improving Energy Efficiency of Building Envelopes: The case of a LEED Gold Building. Procedia Eng. 2015,118, 639–646. [CrossRef] 50. Munz, D.; Yang, Y.Y. Stress Singularities at the Interface in Bonded Dissimilar Materials under Mechanical and Thermal Loading. J. Appl. Mech. 1992,59, 857–861. [CrossRef] 51. Nejdl, L.; Kudr, J.; Cihalova, K.; Chudobova, D.; Zurek, M.; Zalud, L.; Kopecny, L.; Burian, F.; Ruttkay-Nedecky, B.; Krizkova, S.; et al. Remote-controlled robotic platform ORPHEUS as a new tool for detection of bacteria in the environment. Electrophoresis 2014,35, 2333–2345. [CrossRef] [PubMed] 52. Nejdl, L.; Kudr, J.; Ruttkay-Nedecky, B.; Heger, Z.; Zima, L.; Zalud, L.; Krizkova, S.; Adam, V.; Vaculovicova, M.; Kizek, R. Remote-Controlled Robotic Platform for Electrochemical Determination of Water Contaminated by Heavy Metal Ions. Int. J. Electrochem. Sci. 2015,10, 3635–3643. 53. Zalud, L.; Kopecny, L.; Burian, F.; Florian, T. CASSANDRA—Heterogeneous reconnaissance robotic system for dangerous environments. In Proceedings of the 2011 IEEE/SICE International Symposium on System Integration, SII 2011, Kyoto, Japan, 20–22 December 2011; pp. 1275–1280. Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).