Coronary vasospasm during pulsed field focal ventricular ablation with solid tip catheter
Full text
Europace (2025) 27, euaf226 https://doi.org/10.1093/europace/euaf226 RAPID COMMUNICATION Coronary vasospasm during pulsed field focal ventricular ablation with solid tip catheter Kazuto Hayasaka 1 *, Petr Peichl 1 , Nicoletta Ventrella 1,2 , Josef Marek 1,2 , Peter Štiavnický 1 , Eva Borišicová 1 , Jana Hašková 1 , Predrag Stojadinovič 1 , Robert Čihák 1 , Dan Wichterle 1,2 , and Josef Kautzner 1 1 Department of Cardiology, Institute for Clinical and Experimental Medicine, Vídeňská 1958/9, Prague 140 00, Czechia; and 2 2nd Department of Medicine—Department of Cardiovascular Medicine, Charles University Medical School I, Prague, Czechia Received 30 July 2025; accepted after revision 15 September 2025; online publish-ahead-of-print 19 September 2025 --------------------------------------------------------------------------------------------------------------------------------------------------------------- Keywords Pulsed-field ablation • Ventricular arrhythmia • Coronary spasm • Irrigated-tip catheter • Coronary venous system * Corresponding author. Tel: +420 261 365 006; fax: +420 236 052 985. E-mail address: [email protected] © The Author(s) 2025. Published by Oxford University Press on behalf of the European Society of Cardiology. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited. Introduction Coronary spasm is a recognized complication of pulsed-field ablation (PFA). 1 Although most reports have focused on large-footprint systems, risks of PFA delivered via focal catheter are less known. The newly available CENTAURI generator (Cardiofocus) enables delivery of monopolar, biphasic PFA using compatible conventional ablation catheters and it has been successfully used for treatment of ventricular arrhythmias (VA). 2 Recent reports indicate that focal PFA can be performed even from within the coronary venous system (CVS) without inducing spasm. 3 However, we have encountered several cases of spasm following focal PFA. Our aim is to share insights from these experiences. Methods The study included patients who underwent treatment for VA using CENTAURI system from May 2023 to June 2025. For navigation, the CARTO 3 mapping system (Biosense Webster) and intracardiac echocardiography (ICE) were utilized. A 3.5-mm irrigated-tip catheter (ThermoCool SmartTouch™, Biosense Webster) was used for both mapping and ablation. For PFA, 25A applications with a pulse width of 3.4 ms were delivered up to three times at each target site. Coronary angiography was conducted before and after each application to check for the presence of spasm. The distance between the catheter tip and coronary arteries were evaluated using biplane fluoroscopy or ICE. Results Of 13 patients treated for VA using PFA near coronary arteries, vasospasm occurred in four patients (31%). The ablation sites were located in the left ventricular outflow tract involving the great cardiac vein (GCV) in 11 patients, the LV posterior wall in one patient, and the right ventricular outflow tract (RVOT) in one patient. Regarding the distance between the catheter and the coronary artery, in the four cases with spasm the mean distance was 0.75 ± 0.65 mm, whereas in the nine cases without spasm it was 3.8 ± 2.5 mm. Case 1 A 51-year-old woman with prior failed radiofrequency ablation (RFA) for summit-origin premature ventricular contractions (PVC) underwent a redo procedure. Pulsed-field ablation was performed in the distal GCV, where the earliest activation was recorded 32 ms pre-QRS. Post-ablation angiography showed 90% spasm of the left circumflex artery (LCX) without ST-segment changes (Figure 1A). During PFA, the catheter was positioned directly over the coronary artery. The vessel diameter returned to baseline one hour after intracoronary nitrate (1 mg) administration. Additional RFA and PFA from the left coronary cusp (LCC) successfully suppressed the PVC without further spasm, and no recurrence has been observed during the 3-month follow-up. Case 2 A 57-year-old woman with a 21% PVC burden from the left ventricular summit underwent ablation. Early activation was detected at the LCC, but RFA at this site was ineffective. At the GCV, 15 ms pre-QRS prompted the delivery of PFA. Despite intranasal nitrate administration (0.8 mg), the patient developed 90% spasm of LCX after PFA (Figure 1B) without ST-segment changes. The coronary spasm resolved within 10 min following intracoronary nitrate (0.5 mg) administration. The distance between the catheter and the artery was 1 mm. Downloaded from https://academic.oup.com/europace/article/27/10/euaf226/8258532 by Institut klinicke a experimentalni mediciny, Vedecka lekarska knihovna user on 30 October 2025
Subsequent RF and PF above the pulmonary valve eliminated the PVC, no recurrence was observed during the 3-month follow-up. Case 3 An 80-year-old woman with dilated cardiomyopathy experienced recurrent ventricular tachycardia (VT) storm despite seven prior ablation procedures. During substrate mapping, low-voltage areas were identified in the lateral wall and summit. Coronary angiography demonstrated 25–50% pre-existing stenosis on LCX. Pulsed-field and RF were applied endocardially and epicardially to the basal superior left ventricle via the GCV, rendering VT non-inducible. However, a 75% spasm was observed in LCX (Figure 1C) with pre-existing stenosis. The distance between the catheter and the artery was 0.5 mm. Intracoronary administration of a total of 5 mg nitrate resolved the spasm. Despite acute VT non-inducibility, VT has recurred, and adjunctive pharmacological therapy was administered. Case 4 A 15-year-old boy who had previously undergone outflow tract PVC ablations experienced recurrence with a burden of 46%, prompting a redo procedure. The earliest activation was observed in the anteroseptal region of the RVOT. Initial angiography suggested a safe distance from the coronary arteries; however, additional PFA at a more inferior position induced 99% spasm in the left anterior descending artery (LAD) without ST-segment changes (Figure 1D). Intracardiac echocardiography revealed the catheter was 1.5 mm from the artery (Figure 1E). The spasm resolved 30 min after intracoronary nitrate administration with a total dose of 12.5 mg. Premature ventricular contractions were successfully eliminated, with no further recurrence at the 1-year follow-up. Discussion In PFA, lesion formation is determined by catheter electrode configuration. With a bipolar electrode configuration, current loss occurs into the blood pool as the energy is transmitted between the two electrodes, resulting in relatively shallow lesions. 4 In contrast, the CENTAURI system employs a unipolar PF system, in which one electrode is positioned at a site distant from the other, allowing the energy to traverse the myocardium towards the return electrode and thereby creating deeper lesions. Animal studies reported depths of 8.2 ± 1.6 mm 5 after unipolar PFA, indicating potential benefit in cases resistant to bipolar PF or RF ablation. Its ability to deliver energy in high-impedance regions, such as a distal GCV, also broadens applicability. Given these features, wider use may be expected; however, coronary spasm remains a Figure 1 Spasm cases and the spatial relationship and distance between the catheter tip and coronary arteries. (A) The angiography showed 90% spasm in left circumflex artery (LCX) #12. The catheter was positioned directly over the coronary artery. (B) Pulsed-field ablation (PFA) resulted in 90% LCX #11 spasm. The small window showed that the catheter crossed the artery at a distance of approximately 1 mm. (C) The catheter was parallel to the artery over a relatively long distance. (D) Left descending artery #6 developed 99% spasm by PFA from right ventricular outflow tract. (E) Intracardiac echography showed that the catheter was close to the artery during PFA. LAD, left anterior descending artery. 2 K. Hayasaka et al. Downloaded from https://academic.oup.com/europace/article/27/10/euaf226/8258532 by Institut klinicke a experimentalni mediciny, Vedecka lekarska knihovna user on 30 October 2025
concern, and the risk may be greater than with bipolar PF due to deeper lesion formation. Reports of spasm during VA ablation using focal PFA have been limited and included a case of mild spasm on angiography. 6 However, as demonstrated in the current series, clinically significant spasm may occur easily with focal PFA, especially when the catheter tip is positioned directly on the coronary artery, and such may not be accompanied by prominent ST-segment changes. This is noteworthy, as prior reports primarily focused on PFA for atrial arrhythmias. 7,8 These findings underscore the importance of coronary angiography before and after ablation, even with focal PFA delivered by solid tip catheter. While prophylactic nitrate administration is often considered, its effectiveness remains uncertain for a focal PF system. A previous study using high-dose intravenous nitroglycerin successfully prevented spasm during cavotricuspid isthmus ablation (CTI) with a multispline catheter. 9 However, the catheter characteristics in that study differ from those used in a focal PF system, which may limit generalizability. In our series, nitrates were administered to only one patient, limiting our ability to conclude their preventive efficacy. Another study reported spasm despite 400 μg of intravenous nitroglycerin prior to focal PFA for CTI. 10 Moreover, spasm in the proximal LAD or LCX during VA ablation could compromise a significantly larger portion of myocardium compared to spasm in the distal right coronary artery during flutter ablation, emphasizing the clinical relevance of this complication. Further research is warranted to determine the optimal nitrate strategy for use with focal PFA. Conclusion In the present study, we observed coronary spasm during PFA from the CVS and RVOT. These findings indicate that such complications can occur with focal PFA, warranting further attention. Funding This study was supported by the project National Institute for Research of Metabolic and Cardiovascular Diseases (Programme EXCELES, Project No. LX22NPO5104)—Funded by the European Union—Next Generation EU. This work was also funded by the project (Ministry of Health, Czech Republic) for development of research organization 00023001 (IKEM, Prague, Czech Republic)—Institutional support. This work was supported by MH CZ - DRO (‘Institute for Clinical and Experimental Medicine - IKEM, IN 00023001’). Conflict of interest: J.K. reports personal fees from Biosense Webster, Boston Scientific, GE Healthcare, Medtronic, and St. Jude Medical (Abbott) for participation in scientific advisory boards, and has received speaker honoraria from Biosense Webster, Biotronik, Boston Scientific, Medtronic, ProMed CS, St. Jude Medical (Abbott) and Viatris. P.P. has received speaker honoraria from St Jude Medical (Abbott), Medtronic, Boston Scientific and Biosense Webster. K.H. has received a scholarship from the Fukuda Foundation for Medical Technology and consultation fees from Biotronik Japan, DVX, Japan Life Line and Win International. All remaining authors have declared no conflicts of interest. Data availability The data underlying this article will be shared on reasonable request to the corresponding author. References 1. Della Rocca DG, Del Monte A, Bala G, Pannone L, Ströker E, Monaco C et al. Transient inferior st-segment elevation and ventricular fibrillation after cavotricuspid isthmus pulsed-field ablation. JACC Clin Electrophysiol 2023;9:704–6. 2. Della Rocca DG, Cespón-Fernández M, Keelani A, Raffa S, Pannone L, Almorad A et al. Focal pulsed field ablation for premature ventricular contractions: a multicenter experience. Circ Arrhythm Electrophysiol 2024;17:e012826. 3. Peichl P, Bulava A, Wichterle D, Schlosser F, Stojadinović P, Borišincová E et al. Efficacy and safety of focal pulsed-field ablation for ventricular arrhythmias: two-centre experience. Europace 2024;26:euae192. 4. Verma A, Asivatham SJ, Deneke T, Castellvi Q, Neal RE II. Primer on pulsed electrical field ablation: understanding the benefits and limitations. Circ Arrhythm Electrophysiol 2021;14:e010086. 5. Verma A, Neal R, Evans J, Castellvi Q, Vachani A, Deneke T et al. Characteristics of pulsed electric field cardiac ablation porcine treatment zones with a focal catheter. J Cardiovasc Electrophysiol 2023;34:99–107. 6. Chaldoupi SM, Nissen SD, Vriesendorp PA, Farnir FIP, Jerltorp K, Linz B et al. Focal monopolar pulsed field ablation from within the great cardiac vein for idiopathic premature ventricular contractions after failed radiofrequency ablation. Heart Rhythm 2024; 21:1877–87. 7. Chun KJ, Miklavčič D, Vlachos K, Bordignon S, Scherr D, Jais P et al. State-of-the-art pulsed field ablation for cardiac arrhythmias: ongoing evolution and future perspective. Europace 2024;26:euae134. 8. Reddy VY, Petru J, Funasako M, Kopriva K, Hala P, Chovanec M et al. Coronary arterial spasm during pulsed field ablation to treat atrial fibrillation. Circulation 2022;146:1808–19. 9. Malyshev Y, Neuzil P, Petru J, Funasako M, Hala P, Kopriva K et al. Nitroglycerin to ameliorate coronary artery spasm during focal pulsed-field ablation for atrial fibrillation. JACC Clin Electrophysiol 2024;10:885–96. 10. Farnir FIP, Chaldoupi SM, Hermans B, Johannessen A, Haugdal MA, Ruwald MH et al. Ablation of cavotricuspid isthmus-dependent atrial flutter using a focal monopolar pulsed-field ablation catheter: feasibility, periprocedural coronary spasms and conduction disorders. Heart Rhythm 2025:S1547-5271(25)02630-X. Coronary vasospasm during pulsed field focal ventricular ablation 3 Downloaded from https://academic.oup.com/europace/article/27/10/euaf226/8258532 by Institut klinicke a experimentalni mediciny, Vedecka lekarska knihovna user on 30 October 2025