On-chip electrochemical detection of dissolved oxygen: Eliminating the requirement for a permeable selective membrane
Full text
Title: On-chip Electrochemical Detection of Dissolved Oxygen: Eliminating the Requirement for a Permeable Selective Membrane. Authors: Ian Seymour, Fiona Barry, Alan O’Riordan, James F. Rohan* * Corresponding author. Affiliation: Tyndall National Institute, Lee Maltings, University College Cork, T12 R5CP, Cork, Ireland Abstract: Membrane-free electrochemical dissolved oxygen sensors were developed using micro interdigitated electrode arrays. The optimised sensor for detection of dissolved oxygen should contain a gold generator electrode to produce hydrogen peroxide, and a platinum collector electrode for the subsequently detection. The microfabrication process required platinum electrodeposition onto one comb of a gold interdigitated array. The deposition also decreased the inter-electrode gap and increased the collection efficiency. The dual metal IDE was then used to sense dissolved oxygen in water. A pre-treatment step was also required to improve the sensitivity of platinum to hydrogen peroxide ensuring a clean, reproducible surface for each test. The developed sensor accurately detected oxygen concentrations between 0 and 9 ppm, with a sensitivity of 1.59 nA/ppm and an LOD of 0.36 ppm. The detection of oxygen was also tested in the presence of unknown interferents, in estuarine river water, and with known interferents. It was found that the magnitude of hydrogen peroxide oxidation was generally unaffected by the presence of interferents, indicating a potential for interferent free oxygen detection without the requirement of oxygen permeable membranes. Keywords: Microfabricated, dissolved oxygen, sensor, hydrogen peroxide, interdigitated array, dual-metal, generator-collector 1. Introduction: Dissolved Oxygen (DO) refers to the amount of gaseous oxygen that a water system contains. DO arises primarily from two main processes: photosynthesis and diffusion from the atmosphere. Agitated water systems will dissolve more oxygen through diffusion, therefore flowing and tidal water systems are expected to have higher DO concentrations than stagnant or quiescent waters.[1] Temperature, salinity, and altitude are just some of the natural environmental factors that can influence the DO concentration. Dramatic and unexpected decreases in DO concentration typically result from contamination of the water source. The most common contaminants are bacteria and organic matter. Nutrient run-off from fertilizers is also a major source of contamination that can dramatically reduce DO in water.[2] From this, it is evident that DO can be an indicator of the overall This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
health of the water system. A decrease in DO or the presence of pollutants becomes apparent quickly in river and lake waters, due to eutrophication. For potable water, however, this is not as readily apparent. Therefore, it is necessary to actively monitor drinking water systems to ensure that the water is safe to consume. Regular monitoring of DO in drinking water can establish a baseline of normal or healthy conditions.[3] Significant deviations from this normal condition can therefore indicate an issue, most likely due to contamination. Current approaches to DO sensing are based on optical or electrochemical detection mechanisms, both of which can be undertaken at the point-of-interest. Optical based DO sensors typically operate by measurement of the luminescence response of a molecule in the presence of oxygen, where higher concentrations of oxygen mean less luminescence, leading to a quantitative measurement.[4] The luminescent species, typically an organic molecule, is provided as part of the overall device, simplifying the analysis. A drawback, however, is that luminescent species need to be replenished over time and can be influenced by environmental factors, such as temperature, pH and self-quenching. Electrochemical approaches to DO sensing, typically involve measurement by direct reduction of oxygen at an electrode surface. The reaction that governs this is shown in Eq. 1.[5][6] 𝑂 2 + 4𝐻 + + 4𝑒 ― → 2𝐻 2 𝑂 Eq. 1 However, the oxygen reduction reaction is subject to interference from other species, particularly in a complex matrix like water. It can also complicate the electrochemical detection of other species in water due to the large reaction potential window.[7] To this end, electrochemical approaches therefore require an oxygen permeable membrane behind which, the electrodes and electrolyte are placed. The membrane acts to ensure that only oxygen reaches the electrode surface by blocking other ions.[8] While this approach is less expensive than the optical method, issues arise when trying to miniaturise the sensor device due to the requirement for a membrane. Miniaturisation is important to facilitate portability, but also to minimise consumption of oxygen. Reduction of oxygen is a destructive process, as the oxygen is irreversibly reduced to water, therefore, larger electrodes may inadvertently deplete the DO concentration with consequent false negatives. However, complex fabrication methods are required for membrane integration at smaller length scales, and as a result much research focuses on improving these methods.[9][10][11] While elimination of the membrane would facilitate the miniaturisation of devices, it could also lead to interferents, such as chlorine, adversely affecting the DO detection. In this work, we propose a method to quantify DO using interdigitated electrode arrays (IDE) without the need for a membrane while also eliminating common interferent effects. In this approach, hydrogen peroxide (H2O2), produced as an intermediate species in oxygen reduction, is used as an indicator of overall DO concentration. This pathway is shown by Eq. 2 and Eq. 3 [12,13] 𝑂 2 + 2𝐻 + + 2𝑒 ― ⇌ 𝐻 2 𝑂 2 Eq. 2 𝐻 2 𝑂 2 + 2𝐻 + + 2𝑒 ― ⇌ 2𝐻 2 𝑂 Eq. 3 Eq. 1 – 3 effectively show a simplified mechanism for the oxygen reduction reaction. A more detailed discussion of this reaction is included in the supplementary information (SEq. 1 – 13). This method permits facile, interference-free DO quantification, without the need for an oxygen permeable membrane, consequently enabling miniaturisation. The detection of hydrogen peroxide during oxygen reduction has classically been achieved using a rotating ring disc electrode (RRDE) configuration.[14– 16] in that method, hydrogen peroxide produced at the disc moves to the outer ring electrode by convection forces, for detection. Typically, the collection efficiencies are quite low by comparison with generator-collector experiments which utilise a redox couple.[17] Collection efficiencies of 15-25 % are considered high for the RRDE with quite complex instrumentation which is unsuitable for miniaturised or portable sensing applications. An alternative to the RRDE for the detection of produced hydrogen peroxide is the use of IDEs. In this approach, detection is governed by diffusion of This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
hydrogen peroxide from the generator to the collector rather than convection. This has been shown to be a viable approach by Postlethwaite et al, using IDEs with 10 µm width and 10 µm gaps.[18] A drawback of this approach at the dimensions investigated was a lower collection efficiency of 5 % by comparison to the RRDE approaches. IDEs with smaller interelectrode gaps, such as 1 µm, would be expected to increase the efficiency significantly in line with ring-disc approaches. In the RRDE set-up, at the ring collector electrode the product of oxidised hydrogen peroxide (i.e., oxygen) is removed to the bulk electrolyte by the prevailing convection forces. In the IDE set-up, the product can diffuse back to the generator electrode to be reduced again, thus improving sensitivity. This IDE approach permits redox cycling, i.e., where the same molecule can be reduced and oxidised repeatedly or the control of the local environment, such as local pH control, by imposition of an appropriate potential at the collector electrode to enable novel analysis that is not otherwise possible. [19,20] In this work, we extended the previous research with IDE and local environment control to investigate the detection of oxygen through its electrochemical reduction to hydrogen peroxide at the generator electrode subsequently detected by oxidation of the peroxide at the collector electrode. Binary Gold and platinum IDEs were investigated to determine which material performs better at this scale. Fabrication of single metal IDEs requires fewer processing steps than a mixed metal binary IDE, therefore sensor devices comprising two IDEs were fabricated as either gold-gold or platinumplatinum. While gold was anticipated to be the superior metal in reducing DO via the two-electron reduction to hydrogen peroxide, platinum was also explored as it may have the appropriate behaviour at the ultra-micro scale to suit the desired detection method. The ability of both gold and platinum to oxidise hydrogen peroxide was also investigated to establish the preferred collector material. Parameters were established for the appropriate sensing technique, including the relevant pretreatment steps. Finally, the method was applied to detection of DO in estuarine river water samples and those containing common reductive interferents to establish the viability of the sensor in real world applications. 2. Materials and Methods: 2.1. Device Fabrication and Characterization Silicon chip-based devices were fabricated using methods similar to those described by Creedon et al and Barry et al.[21][22] Each chip consisted of two combs of gold working IDEs. A platinum pseudo reference and gold counter electrode were also employed on-chip. In brief, chips were designed to interface with external electronics via a microSD port to facilitate facile electrical connection. Devices were batch fabricated on 4-inch silicon wafers with a 300 nm thermally grown silicon dioxide insulation layer. Blanket deposits of titanium (10 nm) and gold (100 nm) was undertaken using a Temescal FC-2000 E-beam evaporator and lift-off technique to yield interdigitated microband (Length 55 µm x Width 1 µm x Height 110 nm) structures with separations between the tines of each comb maintained at 1.2 µm. A second metal evaporation and lift-off process was undertaken to define the interconnection tracks, contact pads and the gold counter electrode (90 µm x 7 mm). To prevent unwanted interactions along the connection tracks an insulating layer of silicon nitride was deposited by plasma enhanced chemical vapor deposition. Photolithography and dry etching were utilized to selectively open windows (45 μm x 100 μm) in the insulating nitride layer over the microband electrodes for electrolyte access. Openings were also created over the counter and reference electrodes and the contact pads. Each chip contains six IDE (sensors) which are separated by 0.94 mm to prevent cross talk between the sensors. Once sensor fabrication was completed, each wafer was diced to yield 28 separate chips. This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
A typical micrograph of one such chip is shown in Figure 1 (a), with the sensor array shown in Figure 1 (b). A custom-made cell holder was fabricated to allow measurement with small electrolyte volumes (~50 µL to 5 mL.). The cell was constructed from an aluminium base and a Teflon lid. Spring loaded probes (Coda Systems Ltd. PM4J Plain Radius Microprobes) were inserted into the lid in position above the peripheral contact pads, to permit electrical connection to external potentiostats. The cell was assembled with a Viton O-ring embedded in the lid to form a seal around the on-chip electrodes. Viton O-rings were chosen for their chemical resistance. The inner diameter of the O-ring was 7 mm with a cross section of 1.6 mm to allow an opening large enough for an electrolyte volume of 5 mls to expose all six sensors, counter, and reference electrodes on the device. Each chip was inspected using optical microscopy to identify any obvious defects or faults. Prior to any electrochemical characterisation chips were cleaned by rinsing for 20 seconds with acetone, iso-propyl alcohol and finally de-ionized water. The chips were dried in a flow of nitrogen and placed in the chip holder. Electrochemical analysis was performed using an Autolab Bipotentiostat (MAC80150 with BA Module, Metrohm). Cyclic voltammograms (CV) were performed from 0 V to 0.6 V at 50 mV/s in 1 mM ferrocene carboxylic acid (FCA, Sigma Aldrich, 97%) dissolved in 10 mM phosphate buffered saline (PBS, Sigma Aldrich). During these scans, the second interdigitated comb of electrodes was held at 0 V. All electrochemical measurements were recorded versus a commercial Ag/AgCl. Optical micrographs were acquired using a calibrated microscope (Axioskop II, Carl Zeiss Ltd.) with a chargecoupled detector camera (CCD; DEI-750, Optronics). SEM images were obtained using a calibrated field emission FEI Quanta FEG 650 at a beam voltage of 10 kV. Figure 1: (a) Fully fabricated silicon chip showing six sensor arrays and on chip counter and reference. (b) Microscope image of one interdigitated sensor array showing two combs of electrodes with each finger 0.8 µm width separated by a 1.2 µm gap. 2.2. Electrochemical Analysis of O2 and H2O2 on Gold and Platinum Electrochemical oxygen reduction was carried out in artificial drinking water (ADW) prepared by dissolving 1 g of sodium bicarbonate, 0.0654 g of magnesium sulphate (Sigma Aldrich, 99.5% anhydrous), 0.3414 g calcium sulphate dehydrate (Honeywell, 99%), 0.007 g potassium phosphate dibasic (Fluka, 98%), potassium phosphate monobasic (Sigma Aldrich, 99%) and 0.01 g sodium nitrate (Sigma Aldrich, 99%) in 10 L of deionised water. ADW was allowed to saturate with oxygen by exposure to air at room temperature (~17 oC). All oxygen concentrations were measured using a commercial optical DO probe (Hach, LDO101). Both gold-gold (Au-Au) and platinum-platinum (Pt-Pt) IDE arrays were used for this work. CVs were performed in each concentration of oxygen sweeping from 1.2 V to – 0.8 V vs. Ag/AgCl at 50 mV/s for 3 cycles. Varying concentrations of hydrogen peroxide solutions This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
(between 20 and 1000 µM) were prepared by diluting a hydrogen peroxide stock (Sigma Aldrich, 30% w/w) in ADW. LSVs were performed in each sample in the voltage range of 0.8 V to -0.2 V vs. Ag/AgCl at 50 mV/s. Scans were performed in triplicate for each concentration. 2.3. Optimization of Parameters for Detection of Intermediate H2O2 from O2 Reduction ADW samples saturated with oxygen were used to determine which metal, Au or Pt, produced higher quantities of hydrogen peroxide. Both metals can sufficiently oxidise hydrogen peroxide, therefore Au-Au and Pt-Pt generator - collector arrays were used for this analysis. CV analysis using the same parameters described in the previous section was undertaken. The only adjustment made being the biasing of the collector electrode comb at 0.4 V vs. Ag/AgCl. 2.4. Platinum Plating on Collector Electrode Comb to form a binary IDE array Platinum was plated on one comb of a gold IDE to produce Au-Pt arrays with a view to improve the sensitivity of the hydrogen peroxide detection. Platinum was plated using a commercial Platinum DNS bath (Johnson Matthey). The collector comb was held at a cathodic current of 0.108 µA for one minute to achieve a platinum plate approximately 83 nm thick. 2.5. Calibration of Dissolved O2 Using Gold-Platinum Generator-Collector Approach Samples of ADW with varying concentrations of oxygen were prepared by purging ADW solutions with nitrogen for 30 minutes. After purging, the ADW samples were covered with parafilm which contained a small perforation. This allowed oxygen to partition back into solution, albeit at a slow rate. It was found that oxygen typically re-dissolved at a rate of approximately 1 ppm per 3 minutes when the concentration was between 0 and 4 ppm. After 4 ppm was reached, the rate slowed to 1 ppm per 7 minutes until a 6 ppm concentration was reached. To bring the solution back to saturated concentrations, oxygen had to be bubbled into the solution, as the time to reach saturation by diffusion was considerably longer. This resulted in ADW solutions with oxygen concentration varied between fully nitrogen purged (~0.5 ppm), and fully saturated (~8.8 ppm). All oxygen concentrations were first measured using a commercial optical DO probe (Hach, LDO101). CV was performed for each concentration, using the Au-Pt mixed metal IDE, in the voltage window of 1.2 V to -0.8 V vs. Ag/AgCl. The Au generator was scanned while the Pt collector was biased at 0.4 V to oxidise the generated hydrogen peroxide. Prior to any scan, the Pt collector was biased at -0.8 V for 20 seconds as a pretreatment step to remove any oxide that may have formed on the surface of the Pt electrode 2.6. Detection of Dissolved O2 in the Presence of Known and Unknown Interferents Samples of ADW were spiked with 1 mg/L iron (FeCl3, Sigma Aldrich) and 1 mg/L chlorine (NaOCl, Milton Sterilizing Fluid), both of which are known to be active interferents in the oxygen reduction window. The method described above was used to test these samples to determine how both the generator and collector electrodes would be affected by the presence of additional reacting species. Furthermore, samples of river water (taken from River Lee, Cork city) were tested using the same approach. The spiked samples were measured versus a blank ADW sample with an equivalent concentration of DO, as measured by the optical DO probe. Similarly, the river water was tested versus a sample of ADW with equivalent concentrations of DO at the same temperature. This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
3. Results and Discussion: 3.1. Characterization of O2 Reduction and H2O2 Oxidation on Gold and Platinum Figure 2(a) shows a comparison of oxygen reduction at gold and platinum electrodes. Both electrodes were immersed in ADW with equivalent concentrations of DO. Oxygen reduction occurred at less negative potentials on platinum, however this reduction proceeds via the one step, four electron pathway to water. The gold electrode reduced oxygen at more cathodic potentials but does so via the two-step pathway. A full explanation of the oxygen reduction reaction is given in the supplementary information (SI Eq. 1 – SI Eq. 13). CVs and Calibration plots of DO for both gold and platinum are also presented in supplementary information (SI Fig. 1 (a) – (d)) and exhibit sensitivities of 3.2 nA/ppm and 3.8 nA/ppm, respectively. The current measured on the reverse sweep, 30 nA, at the platinum electrode, after the platinum oxide has been reduced, is equivalent to the current measured on gold after the second reduction step. By not allowing the platinum oxide to form, this peak behavior is removed, highlighting that it is solely attributed to the oxide reduction (SI Fig. 2). Both IDEs have the same dimensions, so it was expected that the measured currents should be similar. Diffusional overlap was also observed to be occurring such that the IDE electrodes were acting as one large electrode as opposed to an array of independent microelectrodes. As the detection method of this sensor involved the oxidation of hydrogen peroxide generated by the reduction of DO, both gold and platinum were evaluated as potential detection materials by comparing their LSV responses in 200 µM hydrogen peroxide. Figure 2(b) shows the oxidation of hydrogen peroxide on one comb of both a gold IDE and a platinum IDE, respectively. The platinum electrode showed a superior oxidation, with a current taken at 0.7 V that was 1.5 times higher than the gold electrode. An oxidation event was seen at 0.2 V on platinum that was absent for gold. This has been attributed to dissociative binding of hydrogen peroxide to the platinum surface, before its subsequent oxidation and released as oxygen. A full calibration of hydrogen peroxide on platinum is shown in the supporting information (SI Fig. 3 (a) and (b). The appropriate potential for collector electrode biasing was determined from these LSVs. The main peak was observed at 0.65 V for both materials. However, the platinum electrode could be biased at any potential from 0.2 V to 0.7 V and activity would be seen for hydrogen peroxide, which is another advantage over gold. The oxidation event on both materials displayed a peak shaped behaviour. Again, this can be explained by diffusional overlap. This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
Figure 2: (a) Comparison of the oxygen reduction reaction on gold (red) and platinum (black) in ADW samples. CVs were performed from 1.2 V to -1 V for gold and -0.8 V for platinum. 1st and 2nd on the gold scan refers to each step in the two step oxygen reduction reaction. (b) Comparison of LSVs in 200 µmol hydrogen peroxide for gold (red) and platinum (black). LSVs were performed from -0.2 V to 0.8 V at 50 mV/s. INSET: Full calibration of Hydrogen Peroxide from 20 µmol to 1000 µmol (Platinum collector). 3.2. Generator-Collector Detection of H2O2 Intermediate While gold had been established as the better material to differentiate the two-step reduction of oxygen, it was necessary to confirm that detectable amounts of hydrogen peroxide were being produced. This was undertaken using a gold IDE where one comb underwent CV from 1.2 V to -1 V (generator) while the other was biased at a constant potential at 0.6 V (collector), suitable for oxidising hydrogen peroxide. Experiments were performed at 50 mV/s to ensure that the produced hydrogen peroxide could diffuse to the collector before being further reduced at the generator. Figure 3(a) shows the result of this test performed in oxygen saturated ADW. The generator electrode delivered a current of 31.95 nA, while the collector detected a current of 11.82 nA. From this, a collection efficiency of 44 % was established, which is significantly higher than previous reports for IDEs and ringdisc systems. While this collection efficiency may appear low by comparison with ferrocene-based couples (~95 %), given the complex reaction mechanism associated with oxygen reduction, this is a very high for hydrogen peroxide analysis. Full conversion from oxygen to hydrogen peroxide may not occur, but it is also possible that the subsequent reaction consuming the hydrogen peroxide may result in lower amounts of the species reacting at the detector. This is also why the collector scan shows a peak rather than a steady-state response, as the concentration of hydrogen peroxide is subsequently decreasing. Similarly, the reactions of the two individual species must be considered. In this case, the oxygen is being chemically altered, therefore a simple calculation for collection efficiency is not readily applied. Significant redox cycling was observed, as evident by the change in overall shape of the CV. The first step of the oxygen reduction current is amplified, by redox cycling between the generator and collector of the oxygen/hydrogen peroxide redox couple. The total current remains the same however, as the second step in the oxygen reduction reaction cannot be redox cycled. A comparison of single mode and GC mode for gold is shown in the supporting Information (SI Fig. 4). The same test was performed on a platinum IDE to further confirm that hydrogen peroxide is not produced in greater quantities. Figure 3(b) shows that the collector response for platinum under the same conditions is extremely small. The reverse sweep of the generator electrode measured 29 nA, while the collector sweep measured approximately 1.2 nA. Again, the collector efficiency cannot be directly applied but This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
for platinum it is only ~4 %, which is significantly less than the gold electrode. Interestingly, the small amount of hydrogen peroxide produced is only measured on the forward sweep. On gold, hydrogen peroxide was produced on the reverse sweep, albeit lower than the forward sweep. Platinum, however, showed no hydrogen peroxide formation at all on the reverse sweep. This indicates that the presence of a surface platinum oxide in some way facilitates the production of hydrogen peroxide. The amount produced, however, was still too low for platinum to be considered a viable generator electrode for peroxide generation. Furthermore, there is no change to the platinum generator signal when the collector is appropriately biased, as no redox cycling occurs, highlighted in the supporting information (SI Fig. 5). Figure 3: (a) Au-Au array in ADW showing oxygen reduction at the generator (black) and hydrogen peroxide oxidation at the collector (red). CV was performed from 1.2 V to -1 V at 50 mV/s with the collector biased at 0.4 V. INSET: The effect of varying the collector potential on the resultant scan, between 0.4 V and 0.7 V. (b) Equivalent scan to (a) performed on a Pt-Pt array. (Inset) shows a close-up of the collector scan highlighting the relative lack of hydrogen peroxide. 3.3. Platinum Plating on Collector Comb of an Interdigitated Array Based on these observations, it was postulated that a gold generator and a platinum collector would provide the optimum sensor performance. Therefore, platinum was electroplated onto one comb of a gold IDE to create a mixed Au-Pt IDE based Sensor. This was undertaken using a commercial platinum plating bath as described in the experimental section. The datasheet defined optimum parameters for plating were a current bias of 1 A/dm2 of electrode surface to give a deposition rate of 5 µm/hr. As the surface area of one comb of electrodes was 1.08 x 10-7 dm2, the applied current was set at 0.108 µA. It was found that the optimum deposition thickness was 100 nm or 60 seconds plating, at which complete coverage of the electrode was observed. Following the plating procedure, the electrodes were characterised to determine the platinum coverage using SEM and energy dispersive x-ray (EDX) spectroscopy. Both techniques were performed on the gold-gold and the gold-platinum IDEs for comparison. Figure 4(a) shows the SEM and EDX, respectively, of a gold-gold IDE. From SEM analysis it was determined that both the generator and collector combs were approximately the same width (0.75 µm), and the gap was slightly over 1.24 µm. The electrodes shown in the SEM image were electrically independent of each other, such that plating on one of these electrodes should not impact the other electrode. EDX was performed on both electrode combs and spectra were obtained for each. The EDX spectra shown in Figure 4(b) shows that gold was the dominant species present, indicated by the arrows. The SEM image in Figure 4(c) for the platinum plated electrode indicates that approximately 100 nm of platinum was deposited on the collector comb of electrodes. The platinum comb width increased to 874 nm, and subsequently the gap decreased to 1.17 µm. The EDX spectrum This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
for a plated electrode shown in Figure 4(d) highlights that while gold was still the dominant species, platinum was also present on the surface, indicated by the silver-coloured arrows. The underlying gold was found to be dominant in the EDX spectrum since only a very thin layer of platinum was deposited. Thus, the X-rays can penetrate through the platinum layer and detect gold. The titanium adhesion layer beneath the gold, was also detected in trace amounts. An EDX was also performed on the unplated gold comb of electrodes. This was done as the width of the gold electrodes seemed larger than the non-plated IDE (777 nm vs. 758 nm). The EDX showed that no platinum was present on the gold electrode, and the larger width may have been due to variability in the fabrication process as it utilises optical lithography materials optimised for a 1.5 µm minimum feature fabrication process. Figure 4: (a) SEM of the Au-Au array prior to Pt plating. (b) EDX analysis of a Au-Au array, gold arrows indicate gold peak. (c) SEM image of an Au-Pt array. (d) EDX analysis of an Au-Pt array, gold arrows indicate gold peaks, grey arrows indicate platinum peaks. CV was then performed in single mode at the plated comb. Figure 5(a) shows a comparison of CVs of 1 mM FCA at a gold comb before and after the comb had been plated with platinum. The observed peak shaped behaviour is again a result of diffusional overlap, which resulted in time dependent diffusion-limited mass transfer behaviour and a reduction peak for the oxidised FCA on the reverse sweep at approximately 0.22 V. The oxidation peak current at 0.35 V increased by approximately 1 nA as a result of the platinum plating, consistent with an increased surface area. A significant behaviour change was seen on the platinum coated electrodes, with a reduction current apparent between 0 V and 0.1 V. This was determined to be the onset of oxygen reduction. Again, this further indicated that platinum had deposited onto the electrode comb, as it reduces oxygen at more anodic potentials than gold. Typical deposition data is shown in the inset of Figure 5(a). This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
[10] L.S.S. Santos, R. Landers, Y. Gushikem, Application of manganese (II) phthalocyanine synthesized in situ in the SiO2/SnO2 mixed oxide matrix for determination of dissolved oxygen by electrochemical techniques, Talanta 85 (2011) 1213–1216. https://doi.org/10.1016/j.talanta.2011.06.003. [11] S.M. Silva, L.F. Aguiar, R.M.S. Carvalho, A.A. Tanaka, F.S. Damos, R.C.S. Luz, A glassy carbon electrode modified with an iron N4-macrocycle and reduced graphene oxide for voltammetric sensing of dissolved oxygen, Microchimica Acta 183 (2016) 1251–1259. https://doi.org/10.1007/s00604-016-1750-6. [12] Y.L. Zheng, D. Mei, Y.-X. Chen, S. Ye, The redox reaction of hydrogen peroxide at an Au(100) electrode: Implications for oxygen reduction kinetics, Electrochemistry Communications 39 (2014) 19–21. https://doi.org/https://doi.org/10.1016/j.elecom.2013.12.005. [13] Y. Zheng, W. Chen, X.-Q. Zuo, J. Cai, Y.-X. Chen, The kinetics of the oxidation and reduction of H2O2 at a Pt electrode: A differential electrochemical mass spectrometric study, Electrochemistry Communications 73 (2016) 38–41. https://doi.org/https://doi.org/10.1016/j.elecom.2016.10.012. [14] U.A. Paulus, T.J. Schmidt, H.A. Gasteiger, R.J. Behm, Oxygen reduction on a high-surface area Pt/Vulcan carbon catalyst: a thin-film rotating ring-disk electrode study, Journal of Electroanalytical Chemistry 495 (2001) 134–145. https://doi.org/10.1016/s00220728(00)00407-1. [15] A. Bonakdarpour, M. Lefevre, R. Yang, F. Jaouen, T. Dahn, J.-P. Dodelet, J.R. Dahn, Impact of Loading in RRDE Experiments on Fe–N–C Catalysts: Twoor Four-Electron Oxygen Reduction?, Electrochemical and Solid-State Letters 11 (2008) B105. https://doi.org/10.1149/1.2904768. [16] V. Stamenković, T.J. Schmidt, P.N. Ross, N.M. Marković, Surface Composition Effects in Electrocatalysis: Kinetics of Oxygen Reduction on Well-Defined Pt3Ni and Pt3Co Alloy Surfaces, The Journal of Physical Chemistry B 106 (2002) 11970–11979. https://doi.org/10.1021/jp021182h. [17] A.J.C. Wahl, I.P. Seymour, M. Moore, P. Lovera, A. O’Riordan, J.F. Rohan, Diffusion profile simulations and enhanced iron sensing in generator-collector mode at interdigitated nanowire electrode arrays, Electrochimica Acta 277 (2018) 235–243. https://doi.org/https://doi.org/10.1016/j.electacta.2018.04.181. [18] T.A. Postlethwaite, J.E. Hutchison, R. Murray, B. Fosset, C. Amatore, Interdigitated Array Electrode as an Alternative to the Rotated Ring−Disk Electrode for Determination of the Reaction Products of Dioxygen Reduction, Analytical Chemistry 68 (1996) 2951–2958. https://doi.org/10.1021/ac960327b. [19] I. Seymour, B. O’Sullivan, P. Lovera, J. F. Rohan, A. O’Riordan, Elimination of Oxygen Interference in the Electrochemical Detection of Monochloramine, Using In Situ pH Control at Interdigitated Electrodes, ACS Sens 6 (2021) 1030–1038. https://doi.org/10.1021/acssensors.0c02264. [20] I. Seymour, B. O’Sullivan, P. Lovera, J.F. Rohan, A. O’Riordan, Electrochemical detection of free-chlorine in Water samples facilitated by in-situ pH control using interdigitated microelectrodes, Sens Actuators B Chem 325 (2020) 128774. https://doi.org/10.1016/J.SNB.2020.128774. This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
[21] N. Creedon, C. Robinson, E. Kennedy, A.O. Riordan, Agriculture 4.0: Development of Serological on-Farm Immunosensor for Animal Health Applications, in: 2019 IEEE SENSORS, 2019: pp. 1–4. https://doi.org/10.1109/SENSORS43011.2019.8956677. [22] S. Barry, A. Pescaglini, A. Wahl, K. Dawson, A. O’Riordan, Reducing Charging Currents at Nanowire Sensors: Simulation, Fabrication and Evaluation, 2019 IEEE SENSORS, 2019: pp. 1– 4. https://doi.org/10.1109/SENSORS43011.2019.8956531. This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed
This preprint research paper has not been peer reviewed. Electronic copy available at: https://ssrn.com/abstract=5334734 Preprint not peer reviewed