Citation: Sejda, F.; Frydrýšek, K.; Pleva, L.; Pompach, M.; Hlinka, J.; Sadílek, M.; Murˇcinková, Z.; Krpec, P.; Havlíˇcek, M.; Madeja, R.; et al. Numerical Analysis of the Calcaneal Nail C-NAIL. Appl. Sci. 2022,12, 5265. https://doi.org/ 10.3390/app12105265 Academic Editor: Claudio Belvedere Received: 25 March 2022 Accepted: 18 May 2022 Published: 23 May 2022 Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. Copyright: © 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https:// creativecommons.org/licenses/by/ 4.0/). applied sciences Article Numerical Analysis of the Calcaneal Nail C-NAIL František Sejda 1,2,*, Karel Frydrýšek 1,2,* , Leopold Pleva 1,3, Martin Pompach 4, Josef Hlinka 5, Marek Sadílek 6, Zuzana Murˇcinková7, Pavel Krpec 8, Miroslav Havlíˇcek 9, Roman Madeja 1,3, Jana Pometlová1,3, Oldˇrich Uˇceˇn 10 and Kamila Dostálová11 1Institute of Emergency Medicine, Faculty of Medicine, University of Ostrava, Syllabova 19, Vítkovice, 703 00 Ostrava, Czech Republic; [email protected] (L.P.); r[email protected] (R.M.); [email protected] (J.P.) 2 Department of Applied Mechanics, Faculty of Mechanical Engineering, VSB-Technical University of Ostrava, 17. listopadu 2172/15, 708 00 Ostrava, Czech Republic 3Trauma Center, University Hospital Ostrava, 17. listopadu 1790, Poruba, 708 52 Ostrava, Czech Republic 4 Nemocnice Pardubického Kraje, Kyjevská44, 53 203 Pardubice, Czech Republic; [email protected] 5Department of Material Engineering, VSB-Technical University of Ostrava, 17. listopadu 15/2172, Poruba, 708 00 Ostrava, Czech Republic; [email protected] 6 Department of Material Technology, Faculty of Mechanical Engineering, VSB-Technical University of Ostrava, 17. listopadu 2172/15, 708 00 Ostrava, Czech Republic; [email protected] 7Faculty of Manufacturing, Technical University of Košice, 080 01 Prešov, Slovakia; [email protected] 8V-NASS, a.s., Halasova 2938/1a, Vítkovice, 703 00 Ostrava, Czech Republic; [email protected] 9Medin, a.s, Vlachovicka 619, 592 31 NovéMˇesto na Moravˇe, Czech Republic; miroslav[email protected] 10 Department of Machine and Industrial Design, Faculty of Mechanical Engineering, VSB-Technical University of Ostrava, 17. listopadu 2172/15, 708 00 Ostrava, Czech Republic; [email protected] 11 Centre of Advanced Innovation Technologies, VSB-Technical University of Ostrava, 17. listopadu 2172/15, 708 00 Ostrava, Czech Republic; [email protected] *Correspondence: [email protected] (F.S.); [email protected] (K.F.); Tel.: +420-737914234 (F.S.) Abstract: The presented article investigates the biomechanics of the calcaneal nail C-NAIL TM by numerical calculations and, partially, experimentally. This nail is widely used in trauma and orthopaedics. A numerical model of implants directly interacting with the bone tissue model obtained from CT scans was calculated. The material properties of the bone tissue can be described by several models; in this work, a non-homogeneous material model with isotropic elements and prescribed elastic modulus was used to provide a more accurate model of the applied force distribution on the individual parts of the implants. The critical areas of the nail and its fixtures were investigated using finite element strength calculations to verify their strength and reliability, contributing to the safety and faster and easier treatment of patients. These analyses suggest that the strength of the calcaneal nail C-NAIL, as well as the stabilization of bone fragments resulting from its use, are sufficient for clinical practice. Keywords: traumatology; orthopaedics; calcaneus C-NAIL; osteosynthesis; biomechanics; finite element analysis 1. Introduction The calcaneus or heel bone (Figure 1) is the biggest bone in the human foot, giving it its shape and functionality. In the foot skeleton, the heel bone forms the rear and bottom parts that bear the body weight through the talus bone. Therefore, the upper part of the heel bone joins the talus bone and forms a part of the ankle joint (talocrural joint) (Figure 2); the front part is adjacent to the cuboid bone (Figure 2) and in the rear part, the Achilles tendon attaches to it [1,2]. Appl. Sci. 2022,12, 5265. https://doi.org/10.3390/app12105265 https://www.mdpi.com/journal/applsci
Appl. Sci. 2022,12, 5265 2 of 14 Appl. Sci. 2022, 12, x FOR PEER REVIEW 2 of 15 Figure 1. X-ray of the foot. Figure 2. Anatomy–joint surfaces of the heel bone. A dislocated heel bone fracture counts among the most complicated injuries of the lower limb and, usually, is associated with permanent damage. Such a fracture also represents a major trauma for the patient. Patients have great difficulty walking, and the damage can lead to the development of arthrosis and pain; in most cases, it is impossible to walk at all. Therefore, osteosynthesis is often used as the treatment of choice. In addition, the bone is directly under the skin and is not covered by muscles but rather only by a fat pad, which (in the case of damage to the superficial structures) leads to additional complications in the healing of both the skin cover and the bone itself [2,3]. The first references to the treatment of calcaneal fractures date back to the time of Hippocrates (460–385 BC). The first recorded resting-based treatment leading to the stabilization of the bone fragments comes from French Petit and DeSault in 1720. In one of the earliest editions of the American medical journal (1880), Bailey described the treatment by rest, bandage, and saline solution. A major breakthrough in the treatment came with the discovery of X-rays, when the understanding of these fractures fundamentally changed and the first efforts at anatomical repositioning appeared. In 1913, the French physician Lerich performed the first osteosynthesis using a splint with screws and bone grafts. Throughout the 20th century, the development of new methods and approaches for open repositioning and internal fixation continued but for most surgeons, conservative treatment (administration of solutions and analgesics, immobilizing casts) still represented the method of choice. The truly massive use of internal and external fixation begun in the 1980s. Together with the development and widespread use of CT (computed tomography) scanning, more accurate diagnosis and classification of these fractures (e.g., according to Sanders) was made possible [2–4]. Figure 1. X-ray of the foot. Appl. Sci. 2022, 12, x FOR PEER REVIEW 2 of 15 Figure 1. X-ray of the foot. Figure 2. Anatomy–joint surfaces of the heel bone. A dislocated heel bone fracture counts among the most complicated injuries of the lower limb and, usually, is associated with permanent damage. Such a fracture also represents a major trauma for the patient. Patients have great difficulty walking, and the damage can lead to the development of arthrosis and pain; in most cases, it is impossible to walk at all. Therefore, osteosynthesis is often used as the treatment of choice. In addition, the bone is directly under the skin and is not covered by muscles but rather only by a fat pad, which (in the case of damage to the superficial structures) leads to additional complications in the healing of both the skin cover and the bone itself [2,3]. The first references to the treatment of calcaneal fractures date back to the time of Hippocrates (460–385 BC). The first recorded resting-based treatment leading to the stabilization of the bone fragments comes from French Petit and DeSault in 1720. In one of the earliest editions of the American medical journal (1880), Bailey described the treatment by rest, bandage, and saline solution. A major breakthrough in the treatment came with the discovery of X-rays, when the understanding of these fractures fundamentally changed and the first efforts at anatomical repositioning appeared. In 1913, the French physician Lerich performed the first osteosynthesis using a splint with screws and bone grafts. Throughout the 20th century, the development of new methods and approaches for open repositioning and internal fixation continued but for most surgeons, conservative treatment (administration of solutions and analgesics, immobilizing casts) still represented the method of choice. The truly massive use of internal and external fixation begun in the 1980s. Together with the development and widespread use of CT (computed tomography) scanning, more accurate diagnosis and classification of these fractures (e.g., according to Sanders) was made possible [2–4]. Figure 2. Anatomy–joint surfaces of the heel bone. A dislocated heel bone fracture counts among the most complicated injuries of the lower limb and, usually, is associated with permanent damage. Such a fracture also represents a major trauma for the patient. Patients have great difficulty walking, and the damage can lead to the development of arthrosis and pain; in most cases, it is impossible to walk at all. Therefore, osteosynthesis is often used as the treatment of choice. In addition, the bone is directly under the skin and is not covered by muscles but rather only by a fat pad, which (in the case of damage to the superficial structures) leads to additional complications in the healing of both the skin cover and the bone itself [2,3]. The first references to the treatment of calcaneal fractures date back to the time of Hippocrates (460–385 BC). The first recorded resting-based treatment leading to the stabilization of the bone fragments comes from French Petit and DeSault in 1720. In one of the earliest editions of the American medical journal (1880), Bailey described the treatment by rest, bandage, and saline solution. A major breakthrough in the treatment came with the discovery of X-rays, when the understanding of these fractures fundamentally changed and the first efforts at anatomical repositioning appeared. In 1913, the French physician Lerich performed the first osteosynthesis using a splint with screws and bone grafts. Throughout the 20th century, the development of new methods and approaches for open repositioning and internal fixation continued but for most surgeons, conservative treatment (administration of solutions and analgesics, immobilizing casts) still represented the method of choice. The truly massive use of internal and external fixation begun in the 1980s. Together with the development and widespread use of CT (computed tomography) scanning, more accurate diagnosis and classification of these fractures (e.g., according to Sanders) was made possible [2–4]. Today, open repositioning combined with internal fixation using a calcaneal plate is the standard method of treatment; however, it is occasionally complicated by problems
Appl. Sci. 2022,12, 5265 3 of 14 with healing of the surgical wound and possible infection. For this reason, new approaches to internal fixation (e.g., mini-invasive approach using nails that do not need such an extensive skin cover opening during insertion into the fracture, thus minimizing the risk of infection) have been developed. The success of heel bone treatment also depends on the strength of the used plates and calcaneal nails; in case of their failure, it is necessary to perform a new osteosynthesis. Despite the best efforts of the surgeons, the success rate of the treatment is not 100% (e.g., infection, osteolysis, etc.); however, modifications and improvements of these implants have led to a reduction in the probability of their failure [2–4]. Widely accepted methods of internal fixation include the method of fixing the plate with screws directly to the bone (Figures 3and 4). This method is used for fractures of the body of the bone. However, fractures often occur in the neck. In such cases, other methods are used, such as nailing, i.e., driving a nail into the bone cavity. Leading manufacturers of external and internal fixators for the heel bone include, for example, the Turkish Normmed, see [ 5 ], which specializes in the treatment of small bones, the Swiss Medartis [ 6 ], GPC Medical Ltd., see [7], the Czech manufacturer Medin a.s. [8], etc. Appl. Sci. 2022, 12, x FOR PEER REVIEW 3 of 15 Today, open repositioning combined with internal fixation using a calcaneal plate is the standard method of treatment; however, it is occasionally complicated by problems with healing of the surgical wound and possible infection. For this reason, new approaches to internal fixation (e.g., mini-invasive approach using nails that do not need such an extensive skin cover opening during insertion into the fracture, thus minimizing the risk of infection) have been developed. The success of heel bone treatment also depends on the strength of the used plates and calcaneal nails; in case of their failure, it is necessary to perform a new osteosynthesis. Despite the best efforts of the surgeons, the success rate of the treatment is not 100% (e.g., infection, osteolysis, etc.); however, modifications and improvements of these implants have led to a reduction in the probability of their failure [2–4]. Widely accepted methods of internal fixation include the method of fixing the plate with screws directly to the bone (Figures 3 and 4). This method is used for fractures of the body of the bone. However, fractures often occur in the neck. In such cases, other methods are used, such as nailing, i.e., driving a nail into the bone cavity. Leading manufacturers of external and internal fixators for the heel bone include, for example, the Turkish Normmed, see [5], which specializes in the treatment of small bones, the Swiss Medartis [6], GPC Medical Ltd., see [7], the Czech manufacturer Medin a.s. [8], etc. Figure 3. Internal fixator–plate for treatment of heel bone fracture (alternative method of treatment). Figure 4. Lateral incision and plate application during heel bone osteosynthesis (alternative method of treatment). The latter company, Medin a.s. [8], developed a calcaneal nail for angular stabilization under the commercial name C-NAIL (Figure 5). This calcaneal intraosseous nail is Figure 3. Internal fixator–plate for treatment of heel bone fracture (alternative method of treatment). Appl. Sci. 2022, 12, x FOR PEER REVIEW 3 of 15 Today, open repositioning combined with internal fixation using a calcaneal plate is the standard method of treatment; however, it is occasionally complicated by problems with healing of the surgical wound and possible infection. For this reason, new approaches to internal fixation (e.g., mini-invasive approach using nails that do not need such an extensive skin cover opening during insertion into the fracture, thus minimizing the risk of infection) have been developed. The success of heel bone treatment also depends on the strength of the used plates and calcaneal nails; in case of their failure, it is necessary to perform a new osteosynthesis. Despite the best efforts of the surgeons, the success rate of the treatment is not 100% (e.g., infection, osteolysis, etc.); however, modifications and improvements of these implants have led to a reduction in the probability of their failure [2–4]. Widely accepted methods of internal fixation include the method of fixing the plate with screws directly to the bone (Figures 3 and 4). This method is used for fractures of the body of the bone. However, fractures often occur in the neck. In such cases, other methods are used, such as nailing, i.e., driving a nail into the bone cavity. Leading manufacturers of external and internal fixators for the heel bone include, for example, the Turkish Normmed, see [5], which specializes in the treatment of small bones, the Swiss Medartis [6], GPC Medical Ltd., see [7], the Czech manufacturer Medin a.s. [8], etc. Figure 3. Internal fixator–plate for treatment of heel bone fracture (alternative method of treatment). Figure 4. Lateral incision and plate application during heel bone osteosynthesis (alternative method of treatment). The latter company, Medin a.s. [8], developed a calcaneal nail for angular stabilization under the commercial name C-NAIL (Figure 5). This calcaneal intraosseous nail is Figure 4. Lateral incision and plate application during heel bone osteosynthesis (alternative method of treatment). The latter company, Medin a.s. [ 8 ], developed a calcaneal nail for angular stabilization under the commercial name C-NAIL (Figure 5). This calcaneal intraosseous nail is used for the mini-invasive fixation of intraarticular and extraarticular fractures of the heel bone, stabilizing the fragments of the heel bone by the nail in conjunction with seven locking screws; this leads to the formation of angularly stable fixation. To achieve maximum stability, the sustentacular fragment (i.e., part or whole broken sustentaculum tali) is fixed into the nail with two locking screws guided by a targeting device [ 8 ]. Mini-invasiveness is one of the major advantages of using C-NAIL. A small lateral approach of about 3 cm from
Appl. Sci. 2022,12, 5265 4 of 14 the apex of the outer ankle towards the base of the 5th metatarsal is sufficient for fragment repositioning; further, just a few mini-incisions are needed for introducing Kirschner guide wires, the actual nail, and the individual screws. This mini-invasiveness significantly reduces the risk of possible infection. The stability of the implant and, thus, the firm fixation of the fragments constitute additional advantages. Appl. Sci. 2022, 12, x FOR PEER REVIEW 4 of 15 used for the mini-invasive fixation of intraarticular and extraarticular fractures of the heel bone, stabilizing the fragments of the heel bone by the nail in conjunction with seven locking screws; this leads to the formation of angularly stable fixation. To achieve maximum stability, the sustentacular fragment (i.e., part or whole broken sustentaculum tali) is fixed into the nail with two locking screws guided by a targeting device [8]. Mini-invasiveness is one of the major advantages of using C-NAIL. A small lateral approach of about 3 cm from the apex of the outer ankle towards the base of the 5th metatarsal is sufficient for fragment repositioning; further, just a few mini-incisions are needed for introducing Kirschner guide wires, the actual nail, and the individual screws. This mini-invasiveness significantly reduces the risk of possible infection. The stability of the implant and, thus, the firm fixation of the fragments constitute additional advantages. The lower age limit for using this treatment is given by the closure of the growth cleft of the child’s heel bone, which closes at the age of 16–17 years. The upper age limit is not specified, it depends on the biological condition of the patient. From doctoral experiences: the youngest patient was 17 years old, and the oldest patient was 75 years old. Figure 5. Internal fixator C-NAIL for treatment of the heel bone fractures and the targeting device. Presently, two types of materials are compatible with human tissue (titanium alloy Ti6Al4V and surgical stainless steel AISI 316L) are most widely used for osteosynthetic plates and nails [9], and locking screws are also from the same materials. The diameter of the screws is 3.5 mm, but the length varies from 22 mm to 70 mm [8]. Here, the finite element (FE) analysis of a calcaneal nail made of titanium alloy will be performed. A homogeneous isotropic linear elastic material model will be used for the actual analysis. The material constants were determined by tensile testing [10]. The modulus of elasticity E was determined to be 105 666 MPa, and the Poisson’s number, which indicates the ratio between longitudinal and transverse elongation, for this titanium is alloy μ titanium = 0.342, see [11]. 2. Numerical Bone Modelling The acquisition of a series of consecutive images forms the basis for creating a correct geometrical and numerical bone model. CT (computed tomography) or MRI (magnetic resonance imaging) are the methods most commonly used for imaging internal organs and bones. For soft tissue imaging, MRI imagery is preferable because of the use of magnetic fields and high-frequency electromagnetic waves (i.e., zero radiation load for both the doctor and the patient) [12]. CT images of the lower limb were used to create good quality models of the heel bone (Figure 6). The Materialise Mimics software [13] was used to subsequently obtain a CAD model of the calcaneus. The proper calculation of the bone model needs to separate the Figure 5. Internal fixator C-NAIL for treatment of the heel bone fractures and the targeting device. The lower age limit for using this treatment is given by the closure of the growth cleft of the child’s heel bone, which closes at the age of 16–17 years. The upper age limit is not specified, it depends on the biological condition of the patient. From doctoral experiences: the youngest patient was 17 years old, and the oldest patient was 75 years old. Presently, two types of materials are compatible with human tissue (titanium alloy Ti6Al4V and surgical stainless steel AISI 316L) are most widely used for osteosynthetic plates and nails [ 9 ], and locking screws are also from the same materials. The diameter of the screws is 3.5 mm, but the length varies from 22 mm to 70 mm [ 8 ]. Here, the finite element (FE) analysis of a calcaneal nail made of titanium alloy will be performed. A homogeneous isotropic linear elastic material model will be used for the actual analysis. The material constants were determined by tensile testing [ 10 ]. The modulus of elasticity E was determined to be 105 666 MPa, and the Poisson’s number, which indicates the ratio between longitudinal and transverse elongation, for this titanium is alloy µtitanium = 0.342, see [11]. 2. Numerical Bone Modelling The acquisition of a series of consecutive images forms the basis for creating a correct geometrical and numerical bone model. CT (computed tomography) or MRI (magnetic resonance imaging) are the methods most commonly used for imaging internal organs and bones. For soft tissue imaging, MRI imagery is preferable because of the use of magnetic fields and high-frequency electromagnetic waves (i.e., zero radiation load for both the doctor and the patient) [12]. CT images of the lower limb were used to create good quality models of the heel bone (Figure 6). The Materialise Mimics software [ 13 ] was used to subsequently obtain a CAD model of the calcaneus. The proper calculation of the bone model needs to separate the dense bone tissue from other soft tissues in the individual CT sections. The Mimics software performs this separation based on the Hounsfield units (HU), which is the density of individual pixels.
Appl. Sci. 2022,12, 5265 5 of 14 Appl. Sci. 2022, 12, x FOR PEER REVIEW 5 of 15 dense bone tissue from other soft tissues in the individual CT sections. The Mimics software performs this separation based on the Hounsfield units (HU), which is the density of individual pixels. HU can be perceived as the level of grey on the X-ray. CT images can be subsequently used to create a 3D model of a bone, organ, or soft tissue. The difficulty of this process largely depends on the quality of the CT imagery—where the CT quality is low, the soft tissues must be distinguished from the bone manually. The CAD model of the heel bone presented in Figure 7 was prepared from CT images provided by the University Hospital Ostrava, Czech Republic. This model was subsequently subjected to FE analysis and served as a basis for the stability analyses of calcaneal nails. Figure 6. CT image of the foot–sagittal plane. Figure 7. The anatomical CAD model created from CT imagery in the Mimics software. 3. Material Properties of the Bone Human bone tissue is a non-homogeneous and anisotropic material. The mechanical properties of bone tissue, which are genetically determined, are therefore strongly dependent on the direction of loading, age, sex and health of the person. Under compressive loading, the ultimate strength is higher than under bending loading [14]. Therefore, the search is on for suitable models describing the complicated mechanical properties of bone tissue that are as close to reality as possible. According to the literature, the use of a nonhomogeneous isotropic material model described by the elastic modulus E and the Poisson number μ for each element of the finite element mesh seems to be a sufficient approximation to the real bone. Figure 6. CT image of the foot–sagittal plane. HU can be perceived as the level of grey on the X-ray. CT images can be subsequently used to create a 3D model of a bone, organ, or soft tissue. The difficulty of this process largely depends on the quality of the CT imagery—where the CT quality is low, the soft tissues must be distinguished from the bone manually. The CAD model of the heel bone presented in Figure 7was prepared from CT images provided by the University Hospital Ostrava, Czech Republic. This model was subsequently subjected to FE analysis and served as a basis for the stability analyses of calcaneal nails. Appl. Sci. 2022, 12, x FOR PEER REVIEW 5 of 15 dense bone tissue from other soft tissues in the individual CT sections. The Mimics software performs this separation based on the Hounsfield units (HU), which is the density of individual pixels. HU can be perceived as the level of grey on the X-ray. CT images can be subsequently used to create a 3D model of a bone, organ, or soft tissue. The difficulty of this process largely depends on the quality of the CT imagery—where the CT quality is low, the soft tissues must be distinguished from the bone manually. The CAD model of the heel bone presented in Figure 7 was prepared from CT images provided by the University Hospital Ostrava, Czech Republic. This model was subsequently subjected to FE analysis and served as a basis for the stability analyses of calcaneal nails. Figure 6. CT image of the foot–sagittal plane. Figure 7. The anatomical CAD model created from CT imagery in the Mimics software. 3. Material Properties of the Bone Human bone tissue is a non-homogeneous and anisotropic material. The mechanical properties of bone tissue, which are genetically determined, are therefore strongly dependent on the direction of loading, age, sex and health of the person. Under compressive loading, the ultimate strength is higher than under bending loading [14]. Therefore, the search is on for suitable models describing the complicated mechanical properties of bone tissue that are as close to reality as possible. According to the literature, the use of a nonhomogeneous isotropic material model described by the elastic modulus E and the Poisson number μ for each element of the finite element mesh seems to be a sufficient approximation to the real bone. Figure 7. The anatomical CAD model created from CT imagery in the Mimics software. 3. Material Properties of the Bone Human bone tissue is a non-homogeneous and anisotropic material. The mechanical properties of bone tissue, which are genetically determined, are therefore strongly dependent on the direction of loading, age, sex and health of the person. Under compressive loading, the ultimate strength is higher than under bending loading [ 14 ]. Therefore, the search is on for suitable models describing the complicated mechanical properties of bone tissue that are as close to reality as possible. According to the literature, the use of a non-homogeneous isotropic material model described by the elastic modulus E and the Poisson number µ for each element of the finite element mesh seems to be a sufficient approximation to the real bone. The magnitudes of the elastic modulus are calculated from the magnitude of HU in the tissue volume. The density of the zone of interest according to HU can be further determined from the CT images. Using the Materialise Mimics software and mathematical and physical relationships, material properties can be assigned to each element. There is a large body of literature and scientific articles dealing with the conversion of Hounsfield units to densities ρ and elastic moduli E, see [ 14 – 19 ] and others. The mechanical properties
Appl. Sci. 2022,12, 5265 6 of 14 of bone tissue vary depending on gender, age, health status (smoker, diabetic), tissue type, genetics, etc. A method discussed in the literature [ 18 ] was used for recalculation of the density and the elasticity modulus (Figure 8). Appl. Sci. 2022, 12, x FOR PEER REVIEW 6 of 15 The magnitudes of the elastic modulus are calculated from the magnitude of HU in the tissue volume. The density of the zone of interest according to HU can be further determined from the CT images. Using the Materialise Mimics software and mathematical and physical relationships, material properties can be assigned to each element. There is a large body of literature and scientific articles dealing with the conversion of Hounsfield units to densities ρ and elastic moduli E, see [14–19] and others. The mechanical properties of bone tissue vary depending on gender, age, health status (smoker, diabetic), tissue type, genetics, etc. A method discussed in the literature [18] was used for recalculation of the density and the elasticity modulus (Figure 8). . Figure 8. Association of the elasticity modulus and Hounsfield units. This approximation of reality is sufficient for the presented research as it primarily focuses on the analysis and optimization of the calcaneal nail; in this paper, therefore, the bone itself is not in the spotlight of interest here. The calcaneal density ρ(x, y, z)/kg∙m−3/can be expressed as: = x + H U · y (1) where x, y, z are coordinate axes with a suitably selected origin of the coordinate system. Similarly, the elasticity modulus (x, y, z) can be expressed as: E(,,) =X∙ (,,) = 9354 ∙ 10 ∙𝜌 (,,) , (2) The distribution of the elasticity modulus is presented in Figures 9 and 10. Figure 8. Association of the elasticity modulus and Hounsfield units. This approximation of reality is sufficient for the presented research as it primarily focuses on the analysis and optimization of the calcaneal nail; in this paper, therefore, the bone itself is not in the spotlight of interest here. As a function, The calcaneal density ρ=ρ(x,y,z)/kg·m−3/can be expressed as: ρ=x+HU ×y(1) where x,yand zare coordinate axes with a suitably selected origin of the coordinate system. Values of HU =HU (x,y,z)are acquired via CT snapshots. Similarly, the elasticity modulus E=E(x,y,z)can be expressed by constants A,Bas: E=A×ρB (x,y,z)=9354 ×10 −7×ρ3.15 (x,y,z)(2) The distribution of the elasticity modulus is presented in Figures 9and 10. Appl. Sci. 2022, 12, x FOR PEER REVIEW 6 of 15 The magnitudes of the elastic modulus are calculated from the magnitude of HU in the tissue volume. The density of the zone of interest according to HU can be further determined from the CT images. Using the Materialise Mimics software and mathematical and physical relationships, material properties can be assigned to each element. There is a large body of literature and scientific articles dealing with the conversion of Hounsfield units to densities ρ and elastic moduli E, see [14–19] and others. The mechanical properties of bone tissue vary depending on gender, age, health status (smoker, diabetic), tissue type, genetics, etc. A method discussed in the literature [18] was used for recalculation of the density and the elasticity modulus (Figure 8). . Figure 8. Association of the elasticity modulus and Hounsfield units. This approximation of reality is sufficient for the presented research as it primarily focuses on the analysis and optimization of the calcaneal nail; in this paper, therefore, the bone itself is not in the spotlight of interest here. The calcaneal density ρ(x, y, z)/kg∙m−3/can be expressed as: = x + H U · y (1) where x, y, z are coordinate axes with a suitably selected origin of the coordinate system. Similarly, the elasticity modulus (x, y, z) can be expressed as: E(,,) =X∙ (,,) = 9354 ∙ 10 ∙𝜌 (,,) , (2) The distribution of the elasticity modulus is presented in Figures 9 and 10. Figure 9. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–transversal projection.
Appl. Sci. 2022,12, 5265 7 of 14 Appl. Sci. 2022, 12, x FOR PEER REVIEW 7 of 15 Figure 9. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–transversal projection. Figure 10. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–sagittal projection. 3.1. Strength Analysis of a Healthy Calcaneus without a C-NAIL Before the actual FEM calculations of the osteosynthetic implants, a stress–strain analysis of the whole (healthy) heel bone was performed as below. The computational model was based on a test machine constructed for static and dynamic (fatigue) experiments described in [20], where the anterior part of the bone was supported by a substitute of the cuboid bone (Figure 11), and the load was applied by a hydraulic test machine through the substitute of the tibial joint to simulate the normal loading of the foot. For the purpose of the experiment, these cuboid and tibial replacements were made of veterinary bone cement. This method (i.e., the replacement of the surrounding skeleton with veterinary or dental cement) has been previously described [21–23] and therefore will not be described in greater detail here. Figure 11. Boundary conditions. The computational model (Figure 11), similar to an experiment, replaces the interaction of the heel bone with the other leg bones by finite elements with the mechanical properties of cured epoxy putty [24]. This proxy (epoxy putty instead of veterinary cement, the Figure 10. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–sagittal projection. 3.1. Strength Analysis of a Healthy Calcaneus without a C-NAIL Before the actual FEM calculations of the osteosynthetic implants, a stress–strain analysis of the whole (healthy) heel bone was performed as below. The computational model was based on a test machine constructed for static and dynamic (fatigue) experiments described in [ 20 ], where the anterior part of the bone was supported by a substitute of the cuboid bone (Figure 11), and the load was applied by a hydraulic test machine through the substitute of the tibial joint to simulate the normal loading of the foot. For the purpose of the experiment, these cuboid and tibial replacements were made of veterinary bone cement. This method (i.e., the replacement of the surrounding skeleton with veterinary or dental cement) has been previously described [ 21 – 23 ] and therefore will not be described in greater detail here. Appl. Sci. 2022, 12, x FOR PEER REVIEW 7 of 15 Figure 9. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–transversal projection. Figure 10. Distribution of the elasticity modulus E/MPa/for the heel bone (100 materials acquired from CT images via Mimics software)–sagittal projection. 3.1. Strength Analysis of a Healthy Calcaneus without a C-NAIL Before the actual FEM calculations of the osteosynthetic implants, a stress–strain analysis of the whole (healthy) heel bone was performed as below. The computational model was based on a test machine constructed for static and dynamic (fatigue) experiments described in [20], where the anterior part of the bone was supported by a substitute of the cuboid bone (Figure 11), and the load was applied by a hydraulic test machine through the substitute of the tibial joint to simulate the normal loading of the foot. For the purpose of the experiment, these cuboid and tibial replacements were made of veterinary bone cement. This method (i.e., the replacement of the surrounding skeleton with veterinary or dental cement) has been previously described [21–23] and therefore will not be described in greater detail here. Figure 11. Boundary conditions. The computational model (Figure 11), similar to an experiment, replaces the interaction of the heel bone with the other leg bones by finite elements with the mechanical properties of cured epoxy putty [24]. This proxy (epoxy putty instead of veterinary cement, the Figure 11. Boundary conditions. The computational model (Figure 11), similar to an experiment, replaces the interaction of the heel bone with the other leg bones by finite elements with the mechanical properties of cured epoxy putty [ 24 ]. This proxy (epoxy putty instead of veterinary cement, the mechanical properties of which were unavailable) was selected on the basis of reasonable expert estimates of the mechanical properties of the modulus of elasticity of the used veterinary bone cement (DEMOTEC95-Demotec, Nidderau, Germany), E= 4830 MPa, µdemotec = 0.3.
Appl. Sci. 2022,12, 5265 8 of 14 Individual parts of the hydraulic testing machine are modelled as elements with mechanical properties of common structural steel, E= 210,000 MPa, µ = 0.3. FEM analysis was performed for a model of the heel bone created from CT images, with a material model respecting the mechanical properties of a real bone described element by element by the gradually changing modulus of elasticity according to Equation (2). In this experiment, the upper (proximal) solid elements representing the talus/upper part of the ankle joint were loaded with an axial force, see Equations (3) and (4). G=m·g= 120·9.81 = 1177.2 N (3) F=Kdyn·G= 1.47·1177.2 = 1722.84 N (4) where G/N/is the static gravitational force, F/N/is the dynamic force, and Kdyn is the dynamic coefficient converting the static problem to a quasi-static problem; g= 9.81/m s−2 /is the gravitational acceleration. To calculate the healthy (solid) calcaneus using FE analysis, a patient with a mass of m= 120 kg was considered. Calculations were performed for the maximum dynamic value that can be transferred to the heel bone, namely the total patient’s dynamic force F. In addition, a dynamic coefficient Kdyn = 1.47 was introduced as a simple solution to converting the static problem into a dynamic one (a common engineering approach; the Kdyn value was established based on our previous measurements of the dynamic load of the foot). The loading force is therefore equal to the product of the dynamic coefficient Kdyn and the patient’s gravitational force G. The elements representing the cuboid bone and the substrate are tightly constrained at all degrees of freedom on the free end faces (highlighted in blue; Figure 11). 3.2. Strength Analysis of a Healthy Heel Bone without C-NAIL The distribution of reduced stress under the HMH (von Mises stress) hypothesis for the solid (healthy) calcaneus generated from CT images and described by a variable linear homogeneous isotropic material model (distribution of 100 material properties across the elements in the calcaneus) is shown in (Figure 12). The largest value of the reduced stress is 11.9 MPa, which is still relatively low stress, located at the point of the contact of the bone model with the veterinary cement (Figure 11). Appl. Sci. 2022, 12, x FOR PEER REVIEW 9 of 15 Figure 12. Distribution of the reduced stress on the healthy heel bone produced from CT images, according to the HMH/MPa/hypothesis. Figure 13 shows the distribution of total displacements for the healthy heel bone under stress created from CT images as described by a variable linear homogeneous isotropic material model (100 materials). The greatest total displacement of 0.44 mm was detected in the sustentaculum tali region. Figure 13. The total displacements on healthy calcaneum created from CT imagery/mm/. The areas with greater values of the HMH stress are those that are most susceptible to fracture (Figure 14; lines between points A-B, B-C). The resulting HMH stress Figure 12. Distribution of the reduced stress on the healthy heel bone produced from CT images, according to the HMH/MPa/hypothesis.
Appl. Sci. 2022,12, 5265 9 of 14 Figure 13 shows the distribution of total displacements for the healthy heel bone under stress created from CT images as described by a variable linear homogeneous isotropic material model (100 materials). The greatest total displacement of 0.44 mm was detected in the sustentaculum tali region. Appl. Sci. 2022, 12, x FOR PEER REVIEW 9 of 15 Figure 12. Distribution of the reduced stress on the healthy heel bone produced from CT images, according to the HMH/MPa/hypothesis. Figure 13 shows the distribution of total displacements for the healthy heel bone under stress created from CT images as described by a variable linear homogeneous isotropic material model (100 materials). The greatest total displacement of 0.44 mm was detected in the sustentaculum tali region. Figure 13. The total displacements on healthy calcaneum created from CT imagery/mm/. The areas with greater values of the HMH stress are those that are most susceptible to fracture (Figure 14; lines between points A-B, B-C). The resulting HMH stress Figure 13. The total displacements on healthy calcaneum created from CT imagery/mm/. The areas with greater values of the HMH stress are those that are most susceptible to fracture (Figure 14; lines between points A-B, B-C). The resulting HMH stress distributions were presented to medical experts who confirmed that these fracture lines correspond to the typical fracture lines on a real bone. Appl. Sci. 2022, 12, x FOR PEER REVIEW 10 of 15 distributions were presented to medical experts who confirmed that these fracture lines correspond to the typical fracture lines on a real bone. Figure 14. The most likely fracture lines of a healthy heel bone. 4. Strength Analysis of the Calcaneal Nail C-NAIL in Interaction with the Heel Bone 4.1. Numerical Model The strength analysis of the C-NAIL calcaneal nail was performed on models of the heel bone virtually “cut” into seven fragments corresponding to complicated comminuted fractures (Figures 15 and 16) according to the Sanders IIB classification [25], similar to the experiment in [23]. Further model parameters, including the material and loading, are presented in Section 3.1. Figure 15. Sanders classification of calcaneal fractures, Type IIB—adapted with permission from Ref. [24]. Figure 14. The most likely fracture lines of a healthy heel bone. 4. Strength Analysis of the Calcaneal Nail C-NAIL in Interaction with the Heel Bone 4.1. Numerical Model The strength analysis of the C-NAIL calcaneal nail was performed on models of the heel bone virtually “cut” into seven fragments corresponding to complicated comminuted fractures (Figures 15 and 16) according to the Sanders IIB classification [ 25 ], similar to the experiment in [23].