Monitoring system for laboratory mice transportation: A novel concept for the measurement of physiological and environmental parameters
Abstract
Producción Científica
Full text
electronics Article Monitoring System for Laboratory Mice Transportation: A Novel Concept for the Measurement of Physiological and Environmental Parameters Carlos González-Sánchez 1, Juan-Carlos Fraile 2, Javier Pérez-Turiel 2,* , Ellen Damm 3, Jochen G. Schneider 3, Daniel Schmitt 1and Frank R. Ihmig 1 1Fraunhofer-Institut für Biomedizinische Technik (IBMT), 66280 Sulzbach/Saar, Germany; [email protected] (C.G.-S.); [email protected].de (D.S.); [email protected].de (F.R.I.) 2ITAP (Instituto de las Tecnologías Avanzadas de la Producción)—Universidad de Valladolid, Paseo del Cauce 59, 47011 Valladolid, Spain; [email protected] 3Luxembourg Centre for Systems Biomedicine, University of Luxembourg Esch-sur-Alzette L-4362, Luxembourg and Internal Medicine II, Saarland University Medical Center, 66421 Homburg, Germany; [email protected] (E.D.); [email protected] (J.G.S.) *Correspondence: [email protected]; Tel.: +34-983-423-355; Fax: +34-983-423-358 Received: 14 November 2018; Accepted: 20 December 2018; Published: 1 January 2019 Abstract: Laboratory mice are used in biomedical research as “models” for studying human disease. These mice may be subject to significant levels of stress during transportation that can cause alterations that could negatively affect the results of the performed investigation. Here, we present the design and realization of a prototypical transportation container for laboratory mice, which may contribute to improved laboratory animal welfare. This prototype incorporates electric potential integrated circuit (EPIC) sensors, which have been shown to allow the recording of physiological parameters (heart rate and breathing rate) and other sensors for recording environmental parameters during mouse transportation. This allows for the estimation of the stress levels suffered by mice. First experimental results for capturing physiological and environmental parameters are shown and discussed. Keywords: Laboratory animal transportation; EPIC sensors; mice stress; physiological signal processing; noninvasive measurements; hardware implementation 1. Introduction Animals are used in biomedical research as “models” for studying human biology and disease, and as test subjects for the development and testing of drugs and vaccines to improve human health, without putting the lives and safety of humans at risk. A moral code that minimizes the impact of research on animals is needed. The European Directive 2010/63/EU indicates that animal-welfare considerations should be given the highest priority in the context of animal keeping, breeding, and use [ 1 ]. The European Commission has created the “Platform on Animal Welfare” [ 2 ] to promote an enhanced dialogue on animal welfare issues that are relevant at the European Union level among competent authorities, businesses, civil society, and scientists. International animal welfare regulations [ 3 , 4 ] oblige the scientific community to improve the welfare of laboratory animals used in biomedical research. Animal ethics committees supervise the research to avoid animal suffering and the unethical behavior of researchers. In 1959, Russell and Burch [ 5 ] described the so-called “concept of the 3 Rs”: Replacement, Reduction, and Refinement. Replacement refers to the use of alternative methods as substitutes for Electronics 2019,8, 34; doi:10.3390/electronics8010034 www.mdpi.com/journal/electronics
Electronics 2019,8, 34 2 of 16 in vivo techniques. Reduction refers to any strategy that will result in minimizing the number of animals needed. Refinement refers to the modification of experimental procedures to minimize the pain and distress of animals used in research. Millions of animals are used each year in laboratories around the world. In the European Union, nearly 11.5 million laboratory animals were used in 2011, about 75% of which were rodents [6]. In research, it is important to minimize all external influences (human interaction and environmental factors) that could deteriorate animal welfare. Studies describe the impact of environmental factors such as light, noise, cage cleaning, and transport on the welfare and stress of laboratory rodents to show that the integrity and well-being of the animals being transported are necessary for their welfare and the quality of the research data [7,8]. Laboratory animals may be subject to significant levels of stress during transport that can cause alterations that pass unnoticed to the researcher but certainly could negatively affect the results of the performed investigation [ 9 ]. Such alterations are, for example: confinement in a transport box; hours of travel; unknown environments; and changes in temperature, light, and humidity. These factors produce high stress levels in animals [ 10 ], which cause changes in their physiological parameters [ 11 – 14 ], such as heart rate (HR), breathing rate (BR), blood pressure, body temperature, and hormone levels. This stress is not limited to the transported animal, because it can also affect their offspring [14,15]. After enduring high stress levels, mice require a long time for their HR and BR values to return to normal levels [ 12 , 13 , 16 ]. Several authors [ 17 , 18 ] have described the need for 24–48 h of rest for the immune system and corticosterone levels to stabilize after transport. Others [ 10 , 19 ] found that mice were not completely acclimated after three to four days by monitoring stress indicators based on animal behavior and corticosterone level. For this reason, it is useful to be able to measure both the physiological signals of the mouse and the environmental parameters of the transport container. This would allow evaluating the stress level that the mouse “suffered” during its transport. In general, heart rate variability (HRV) analysis provides a more accurate measurement of stress than simply observing cardiac rhythm, because the cardiac rhythm can vary greatly from one animal to another or can depend on physical activity. HRV is defined as beat-to-beat changes in HR or variations of the RR intervals (time between two R peaks in an electrocardiogram (ECG)) in consecutive cardiac cycles. Measuring interbeat intervals (IBI) and HRV are commonly used methods for monitoring the level of stress in small laboratory animals during transport [20,21]. Acquisition of HR in small laboratory animals is usually performed using invasive techniques with implantable radio link electrodes placed inside the animal [ 22 , 23 ]. With these invasive techniques, a complete ECG of the animal is obtained, but the placement of the electrodes causes stress on the animal and increases its risk of death. An alternative system for recording ECGs in conscious mice without anesthesia or implants is described in [ 24 ]. The system includes paw-sized conductive electrodes embedded in a platform configured to record ECGs when three single electrodes contact three paws. The measurement of physiological signals in mice using conventional techniques, such as immobilization or anesthesia, causes stress in the animal [ 25 ], since they do not allow them to move freely during transport and, therefore, should not be used. It is of great importance to use noninvasive techniques for monitoring physiological signals because this allows the acquisition of information about the mice’s condition with the aim to improve it according to the three Rs principle. This may improve the performance that an animal provides for science, further decreasing the number of animals required for research. We propose the use of noninvasive capacitive sensors to monitor the physiological signals of laboratory mice in order to avoid additional stress. Generally, capacitive sensors are used to measure HR in humans via either direct skin contact or through one and two layers of clothing with no dielectric gel and no grounding electrode [ 26 – 28 ] and also to measure BR in humans using a conductive
Electronics 2019,8, 34 3 of 16 textile-based wearable sensor [ 29 , 30 ]. However, we have not found reports of the use of capacitive sensors for monitoring physiological signals in mice during transport. This paper presents a new concept for a monitoring system for laboratory mouse transportation that incorporates a matrix of electric potential integrated circuit (EPIC) sensors for recording mouse physiological parameters (HR and BR) and other sensors for recording environmental parameters. We have already evaluated EPIC sensors for noninvasive breathing and heart monitoring in nonrestrained, nonsedated laboratory mice [ 31 ]. Here, the design and control strategy for a 4 × 4 matrix of EPIC sensors is presented. The integration of hardware (matrix of EPIC sensors, webcam, visible light and infrared sensors, accelerometer, humidity sensor, buzzer, and circuits for the voltage adaptation of inputs and signals) and software (sensor selection, signal conditioning, and acquisition) in a prototypical transport container is also described. The first experimental results for capturing physiological and environmental parameters are shown and discussed. 2. Materials and Methods 2.1. Legislation about Laboratory Animals Transport The international legislation on the transport of laboratory animals aims to achieve shipping compliance with the specific laws from each country through which the animals travel. The following documents may be used as reference: •EU Directive on the protection of animals during transport; •European Convention ETS 193 for the protection of animals during international transport; •International Air Transportation Association (IATA) Live Animal Regulations; •Guidelines for the Humane Transportation of Animals Research. Based on information contained in these documents, a short summary of the rules that may affect this research is presented below. Regulations on the design of container: •Adequate ventilation to avoid or minimize the entry of bacteria and viruses. • Provide appropriate gripping systems that do not compromise the animal (keeping the box horizontal, for example) or the personnel handling the container. • The material of which the container is made must be hard, rigid, and resistant, especially to moisture. For this, specially coated cardboard, fiberglass, aluminum, or more commonly, plastics may be employed. The interior surfaces should be durable and smooth, so that the animal cannot damage or gnaw them. Regulations on the needs of space: • The space specified for each animal must be broad enough to allow the animal a normal posture and to move freely, avoiding claustrophobic feelings that can cause stress, but not so large as to allow the animal to hit the walls hard in any case or crush with other animals. • It should also allow sufficient space between the upper part of the body and the top of the container, so that air can circulate freely. • The publication “Guidance on the transport of laboratory animals” [ 32 ] includes information on the minimum dimensions for different species of rodent, depending on their weight. Regulations on the environmental conditions: • Temperature: Suitable temperature for the transport of rodents may vary between 4 ◦ C and 34 ◦ C but should ideally be kept between 20 ◦C and 26 ◦C at all times. • Humidity: The acceptable range is between 40% and 70%. Very low humidity can cause respiratory problems and higher humidity favors the development of bacteria and microorganisms.
Electronics 2019,8, 34 4 of 16 • Amount of light: For common animals, the amount of light should be kept as small as possible, always less than 60 lux. Ideally, a 12 h ON and 12 h OFF pattern should be used, including dimmed simulation of sunrise and sunset. •Sound: Another major source of stress is the sound level, which must be kept as low as possible. Of course, this is often difficult during transport. Mice have the ability to make sounds between 23 Hz and 85 kHz, while the range for rats is between 250 Hz and 70 kHz. If they suffer stress, rats typically emit sounds at a frequency close to 22 kHz [33]. 2.2. Design of Monitoring System and Transport Box In the scientific literature, there are few designs of transport boxes that incorporate sensors to infer the stress level of the animal. A handling device for safely moving wild rats with physical partitions but without sensors is presented in [ 34 ]. A simple container, without sensors, consisting of a propylene tube, a high-efficiency particulate aerosol (HEPA) filter, and a rubber glove for transporting small animals to magnetic resonance imaging is described in [35]. The monitoring system for transportation of laboratory mice that we have designed incorporates a matrix of EPIC sensors and environmental sensors. EPIC sensors have the advantage of providing a strong and stable signal that can be captured even through dielectrics, such as plastic storage boxes and materials that form the nest and habitat of the animal (e.g., sawdust, paper, and bark). There are other physiological or behavioral signals capable of indicating the level of stress of mice: the type and frequency of sounds the mouse makes, the number of times it gets up on its hind legs, or the distance traveled. So, including a camera and/or microphone in the mouse transport box might be interesting for better control. Figures 1and 2gather all the information about the monitoring system concept and the transport box design. Electronics2019,8,xFORPEERREVIEW4of16 Amountoflight:Forcommonanimals,theamountoflightshouldbekeptassmallaspossible, alwayslessthan60lux.Ideally,a12hONand12hOFFpatternshouldbeused,including dimmedsimulationofsunriseandsunset. Sound:Anothermajorsourceofstressisthesoundlevel,whichmustbekeptaslowaspossible. Ofcourse,thisisoftendifficultduringtransport.Micehavetheabilitytomakesoundsbetween 23Hzand85kHz,whiletherangeforratsisbetween250Hzand70kHz.Iftheysufferstress,rats typicallyemitsoundsatafrequencycloseto22kHz[33]. 2.2.DesignofMonitoringSystemandTransportBox Inthescientificliterature,therearefewdesignsoftransportboxesthatincorporatesensorsto inferthestressleveloftheanimal.Ahandlingdeviceforsafelymovingwildratswithphysical partitionsbutwithoutsensorsispresentedin[34].Asimplecontainer,withoutsensors,consistingof apropylenetube,ahigh‐efficiencyparticulateaerosol(HEPA)filter,andarubberglovefor transportingsmallanimalstomagneticresonanceimagingisdescribedin[35]. Themonitoringsystemfortransportationoflaboratorymicethatwehavedesigned incorporatesamatrixofEPICsensorsandenvironmentalsensors.EPICsensorshavetheadvantage ofprovidingastrongandstablesignalthatcanbecapturedeventhroughdielectrics,suchasplastic storageboxesandmaterialsthatformthenestandhabitatoftheanimal(e.g.,sawdust,paper,and bark). Thereareotherphysiologicalorbehavioralsignalscapableofindicatingthelevelofstressof mice:thetypeandfrequencyofsoundsthemousemakes,thenumberoftimesitgetsuponitshind legs,orthedistancetraveled.So,includingacameraand/ormicrophoneinthemousetransportbox mightbeinterestingforbettercontrol.Figures1and2gatheralltheinformationaboutthe monitoringsystemconceptandthetransportboxdesign. Figure1.Modularconceptofthemonitoringsystem. Themonitoringsystemmusthavethefollowingcharacteristics: Useofarechargeablebatteryduetotheportabilityofthesystem. Capture,storage,andalarmmanagementforthefollowingenvironmentalparameters: Temperature,humidity,movement,lightintensityinsidethebox,andatmosphericpressure.A cameramustbeincludedtorecordtheinteriorofthetransportboxforvisualinformationabout thestateofthemice.Thiscouldalsobeusedtocalculatethepositionandmovementofthemice andassesstheirlevelofactivity. DetectionofthestresslevelofthemiceusingthematrixofEPICsensorsincludingpower managementaswellassignalcapture,filtering,andanalysis. Acousticaloropticalalertforvaluesoutsidetheacceptablerange. Figure 1. Modular concept of the monitoring system. The monitoring system must have the following characteristics: •Use of a rechargeable battery due to the portability of the system. • Capture, storage, and alarm management for the following environmental parameters: Temperature, humidity, movement, light intensity inside the box, and atmospheric pressure. A camera must be included to record the interior of the transport box for visual information about the state of the mice. This could also be used to calculate the position and movement of the mice and assess their level of activity. • Detection of the stress level of the mice using the matrix of EPIC sensors including power management as well as signal capture, filtering, and analysis.
Electronics 2019,8, 34 5 of 16 •Acoustical or optical alert for values outside the acceptable range. • Wireless access to information: The monitoring system must be able to exchange all data, preferably using WiFi, cellular networks, or RFID. This exchange could take place during transportation or once the transport box has already arrived at the destination, downloading the data to a tablet or similar device. Electronics2019,8,xFORPEERREVIEW5of16 Wirelessaccesstoinformation:Themonitoringsystemmustbeabletoexchangealldata, preferablyusingWiFi,cellularnetworks,orRFID.Thisexchangecouldtakeplaceduring transportationoroncethetransportboxhasalreadyarrivedatthedestination,downloadingthe datatoatabletorsimilardevice. Figure2.Prototypedesignofthetransportbox. 2.3.MatrixofEPICSensorsforCapturingPhysiologicalSignalsinMice Ourexperimentsforcapturingphysiologicalsignalsinlaboratorymiceweredevelopedwith C57BL6/J/LDLR−/− strainmice.AnimalexperimentswereapprovedbytheSaarlandUniversity MedicalCenteranimalexperimentationoffice(animalprotocol30/2012). Themeasurementplatformwascomposedof16EPICsensorsarrangedina4×4matrix.Figure3a showstheprintedcircuitboard(PCB)designedandbuilttocontrolthematrixofEPICsensors showninFigure3b.EachEPICsensorcoversanareaof10×10mm. (a)(b) Figure3.(a)Self‐designedprintedcircuitboard(PCB)usedtocontrolthe4×4matrixofelectric potentialintegratedcircuit(EPIC)sensors.(b)MatrixofEPICsensorsusedforphysiologicalsignals capturinginrodents. WeusedPS25251EPICcapacitivesensors(PlesseySemiconductors,Plymouth,UK)withthe followingfeatures:inputresistance(20GΩ),dry‐contactcapacitivecoupling,inputcapacitanceas lowas15pF,lower−3dBpointtypically200mHz,andupper−3dBpointtypically10kHz.It operatedwithabipolarpowersupplyfrom±2.4to±5.5V.Weusedtheinstrumentationamplifier INA128(TexasInstruments,Dallas,USA)toremovethecommonmodenoise.Thecaptured physiologicalsignalswerefedintoafirstorderhigh‐passfilterandamplifiedwithanOPA2131 (TexasInstruments,Dallas,USA),capturedbyadigitaloscilloscope(MDO4104B‐6,Tektronix, Beaverton,Oregon,USA),andstoredonaPC. EachEPICsensoroperatedproperlywhenthepositivepinreceivedavoltage(Vdd)lessthan 5.5V,thenegativepinavoltage(Vss)greaterthan–5.5V,andtheGNDpinshouldbeconnectedto0 Figure 2. Prototype design of the transport box. 2.3. Matrix of EPIC Sensors for Capturing Physiological Signals in Mice Our experiments for capturing physiological signals in laboratory mice were developed with C57BL6/J/LDLR − / − strain mice. Animal experiments were approved by the Saarland University Medical Center animal experimentation office (animal protocol 30/2012). The measurement platform was composed of 16 EPIC sensors arranged in a 4 × 4 matrix. Figure 3a shows the printed circuit board (PCB) designed and built to control the matrix of EPIC sensors shown in Figure 3b. Each EPIC sensor covers an area of 10 ×10 mm. Electronics2019,8,xFORPEERREVIEW5of16 Wirelessaccesstoinformation:Themonitoringsystemmustbeabletoexchangealldata, preferablyusingWiFi,cellularnetworks,orRFID.Thisexchangecouldtakeplaceduring transportationoroncethetransportboxhasalreadyarrivedatthedestination,downloadingthe datatoatabletorsimilardevice. Figure2.Prototypedesignofthetransportbox. 2.3.MatrixofEPICSensorsforCapturingPhysiologicalSignalsinMice Ourexperimentsforcapturingphysiologicalsignalsinlaboratorymiceweredevelopedwith C57BL6/J/LDLR−/− strainmice.AnimalexperimentswereapprovedbytheSaarlandUniversity MedicalCenteranimalexperimentationoffice(animalprotocol30/2012). Themeasurementplatformwascomposedof16EPICsensorsarrangedina4×4matrix.Figure3a showstheprintedcircuitboard(PCB)designedandbuilttocontrolthematrixofEPICsensors showninFigure3b.EachEPICsensorcoversanareaof10×10mm. (a)(b) Figure3.(a)Self‐designedprintedcircuitboard(PCB)usedtocontrolthe4×4matrixofelectric potentialintegratedcircuit(EPIC)sensors.(b)MatrixofEPICsensorsusedforphysiologicalsignals capturinginrodents. WeusedPS25251EPICcapacitivesensors(PlesseySemiconductors,Plymouth,UK)withthe followingfeatures:inputresistance(20GΩ),dry‐contactcapacitivecoupling,inputcapacitanceas lowas15pF,lower−3dBpointtypically200mHz,andupper−3dBpointtypically10kHz.It operatedwithabipolarpowersupplyfrom±2.4to±5.5V.Weusedtheinstrumentationamplifier INA128(TexasInstruments,Dallas,USA)toremovethecommonmodenoise.Thecaptured physiologicalsignalswerefedintoafirstorderhigh‐passfilterandamplifiedwithanOPA2131 (TexasInstruments,Dallas,USA),capturedbyadigitaloscilloscope(MDO4104B‐6,Tektronix, Beaverton,Oregon,USA),andstoredonaPC. EachEPICsensoroperatedproperlywhenthepositivepinreceivedavoltage(Vdd)lessthan 5.5V,thenegativepinavoltage(Vss)greaterthan–5.5V,andtheGNDpinshouldbeconnectedto0 Figure 3. ( a ) Self-designed printed circuit board (PCB) used to control the 4 × 4 matrix of electric potential integrated circuit (EPIC) sensors. ( b ) Matrix of EPIC sensors used for physiological signals capturing in rodents. We used PS25251 EPIC capacitive sensors (Plessey Semiconductors, Plymouth, UK) with the following features: input resistance (20 G Ω ), dry-contact capacitive coupling, input capacitance as low as 15 pF, lower − 3 dB point typically 200 mHz, and upper − 3 dB point typically 10 kHz. It operated with a bipolar power supply from ± 2.4 to ± 5.5 V. We used the instrumentation amplifier INA128 (Texas Instruments, Dallas, USA) to remove the common mode noise. The captured physiological signals were fed into a first order high-pass filter and amplified with an OPA2131 (Texas Instruments, Dallas, USA), captured by a digital oscilloscope (MDO4104B-6, Tektronix, Beaverton, Oregon, USA), and stored on a PC.
Electronics 2019,8, 34 6 of 16 Each EPIC sensor operated properly when the positive pin received a voltage (Vdd) less than 5.5 V, the negative pin a voltage (Vss) greater than − 5.5 V, and the GND pin should be connected to 0 V. There are different approaches to control the powering of the EPIC sensor matrix, depending on the way the three power pins of EPIC sensor are connected. The solution we implemented into the self-designed PCB was based on two main components: the 8-bit shift register 74HC595 integrated circuit and the ADG1606 analog multiplexer (Analog Devices, Norwood, USA). The integrated circuit 74HC595 is a very common shift register, which is capable of controlling up to eight outputs using only three signals: a clock signal, a switch, and the serial input signal. Another advantage is its ability to be daisy chained, that is, to use an output pin to transmit information to the input of a second integrated circuit. This could be connected to a third, and so on, requiring only three inputs for all of the shift registers. Because we needed to control 16 sensors, we used two daisy-chained integrated circuits and three signals for its control. In order to select the sensors to power, we considered all different possible combinations. There are 16 EPIC sensors with two states (ON/OFF), so the total number of combinations is 2 16 = 65,536. This can be represented by 2 bytes of information in binary code. Decimal values corresponding to each position in the 4 ×4 matrix are shown in Table 1. Table 1. Decimal values associated with each matrix position used for the sensors’ power supply. (1,1) (1,2) (1,3) (1,4) 1 2 4 8 (2,1) (2,2) (2,3) (2,4) 16 32 64 128 (3,1) (3,2) (3,3) (3,4) 256 512 1024 2048 (4,1) (4,2) (4,3) (4,4) 4096 8192 16384 32768 Any number between 0 and 65535 can be expressed as the sum of some of these terms. Its binary value is the 2-byte representation of the 4 × 4 matrix, in which each bit represents a sensor, where 0 = OFF, 1 = ON. For example, the binary representation of the code “33284” is “1000001000000100” (bit less significant to the right), which activates the sensors (1,3), (3,2), and (4,4). Note that the positions in the binary number are reversed relative to those shown in the table above, since the positions (1,1) (code 1) and (4.4) (code 32768), respectively, correspond to the least and most significant positions. The next part of our PCB circuit selected two of the outputs of the 16 EPIC sensors using two ADG1606 analog multiplexers (see Figure 4). By making use of four control signals (A0, A1, A2, and A3), we could select 1 of the 16 inputs (S1–S16), which was transmitted to the output D. Therefore, we needed eight signals to select a sensor. Electronics2019,8,xFORPEERREVIEW6of16 V.TherearedifferentapproachestocontrolthepoweringoftheEPICsensormatrix,dependingon thewaythethreepowerpinsofEPICsensorareconnected.Thesolutionweimplementedintothe self‐designedPCBwasbasedontwomaincomponents:the8‐bitshiftregister74HC595integrated circuitandtheADG1606analogmultiplexer(AnalogDevices,Norwood,USA). Theintegratedcircuit74HC595isaverycommonshiftregister,whichiscapableofcontrolling uptoeightoutputsusingonlythreesignals:aclocksignal,aswitch,andtheserialinputsignal. Anotheradvantageisitsabilitytobedaisychained,thatis,touseanoutputpintotransmit informationtotheinputofasecondintegratedcircuit.Thiscouldbeconnectedtoathird,andsoon, requiringonlythreeinputsforalloftheshiftregisters.Becauseweneededtocontrol16sensors,we usedtwodaisy‐chainedintegratedcircuitsandthreesignalsforitscontrol. Inordertoselectthesensorstopower,weconsideredalldifferentpossiblecombinations.There are16EPICsensorswithtwostates(ON/OFF),sothetotalnumberofcombinationsis2 16 =65536. Thiscanberepresentedby2bytesofinformationinbinarycode.Decimalvaluescorrespondingto eachpositioninthe4×4matrixareshowninTable1. Table1.Decimalvaluesassociatedwitheachmatrixpositionusedforthesensors’powersupply. (1,1)(1,2)(1,3)(1,4) 1248 (2,1)(2,2)(2,3)(2,4) 163264128 (3,1)(3,2)(3,3)(3,4) 25651210242048 (4,1)(4,2)(4,3)(4,4) 409681921638432768 Anynumberbetween0and65535canbeexpressedasthesumofsomeoftheseterms.Its binaryvalueisthe2‐byterepresentationofthe4×4matrix,inwhicheachbitrepresentsasensor, where0=OFF,1=ON. Forexample,thebinaryrepresentationofthecode“33284”is“1000001000000100”(bitless significanttotheright),whichactivatesthesensors(1,3),(3,2),and(4,4).Notethatthepositionsin thebinarynumberarereversedrelativetothoseshowninthetableabove,sincethepositions(1,1) (code1)and(4.4)(code32768),respectively,correspondtotheleastandmostsignificantpositions. ThenextpartofourPCBcircuitselectedtwooftheoutputsofthe16EPICsensorsusingtwo ADG1606analogmultiplexers(seeFigure4).Bymakinguseoffourcontrolsignals(A0,A1,A2,and A3),wecouldselect1ofthe16inputs(S1–S16),whichwastransmittedtotheoutputD.Therefore, weneededeightsignalstoselectasensor. Figure4.FunctionalblockdiagramoftheADG1606multiplexer. Tocontrolbothmultiplexers,wecouldmakeuseofeightoutputsofthecontrolboarddirectly, butthiswouldnotbeaveryefficientsolution.Itwasmoreefficienttouseanothershiftregister 74HC595,whichallowedustocontroltheeightoutputsusingonlythreesignals. WeprovidedanumbertoeachEPICsensorbasedonitspositioninthematrix(seeTable2).Its binaryrepresentationcorrespondedtothevaluethatweprovidedtothemultiplexercontrolsignals toselectthecorrespondingsensor.Forexample,toselectthesensoroutput(4,2),whichisassigned Figure 4. Functional block diagram of the ADG1606 multiplexer. To control both multiplexers, we could make use of eight outputs of the control board directly, but this would not be a very efficient solution. It was more efficient to use another shift register 74HC595, which allowed us to control the eight outputs using only three signals.
Electronics 2019,8, 34 7 of 16 We provided a number to each EPIC sensor based on its position in the matrix (see Table 2). Its binary representation corresponded to the value that we provided to the multiplexer control signals to select the corresponding sensor. For example, to select the sensor output (4,2), which is assigned the number 13, the control signals must be 1, 1, 0, 1, respectively. It must be considered that when the selections are inverted, the output signal is also inverted. For example, combinations 11111010 (sensor #15 and #10) and 10101111 (sensor #10 and #15) generate the same signal but inverted. The proposed solution allowed flexible and stable operation using only six inputs for the power control and output selection. Table 2. Values used for each matrix position that select the output to be processed. (1,1) (1,2) (1,3) (1,4) 0–0000 1–0001 2–0010 3–0011 (2,1) (2,2) (2,3) (2,4) 4–0100 5–0101 6–0110 7–0111 (3,1) (3,2) (3,3) (3,4) 8–1000 9–1001 10–1010 11–1011 (4,1) (4,2) (4,3) (4,4) 12–1100 13–1101 14–1110 15–1111 In the last part of the PCB circuit, previously selected signals were carried to the instrumentation amplifier that computed the difference and then amplified the signal. 2.4. Selection of Commercial Components and Modules The central processing unit (CPU) was responsible for carrying out the capture and analysis of the data obtained from the sensors, processing the information, and communicating with the user. For this purpose, the UDOO QUAD board (SECO USA Inc., Burlington, USA) was selected, which is a single board computer that can be equipped with Linux or Android operating systems. This development board had two microprocessors connected by serial communication. The main microprocessor corresponded to the NXP i.MX6 ARM ® Cortex ® -A9 quad core processor at 1 GHz performance. The second microprocessor was the Atmel SAM3X8E ARM Cortex-M3. This feature made the UDOO board compatible with all libraries developed for Arduino and all those accessories and development tools. The board also included a 1 Gb DDR3 RAM, a WiFi module, four USB ports (one of them OTG), an HDMI (touch-enabled), RJ45 Ethernet, analog input and output audio camera (CSI), hard disk (SATA) connectors, and an SD card connector that acted as a startup disk. A circuit was developed to generate the appropriate voltage to power the CPU board that switched the EPIC sensors ON/OFF and acquired the signal. This circuit worked with two voltages: positive (Vdd, +5 V) and negative (Vss, − 5 V), along with a reference (GND). We used a 9-V battery connected to a voltage regulator (Maxim MAX764), which regulated the output over a wide range of load currents and reduced power, capable of delivering up to 1.5 W, simultaneously. We decided to use commercially available sensors for the prototyping process of the transport box. Figure 5shows the CPU board and the selected environmental sensors. The following sensors have been used: Webcam—Logilink UA0072: For capturing images and video inside the transport box, we included a webcam featuring: 0.3 Megapixels, CMOS sensor (640 × 480), F14mm lenses, 54 ◦ angle, automatic exposure and brightness adjustments, USB 2.0 connection, refresh rate of 30 frames per second, and 24-bit color depth. In order to use infrared light, which does not interfere with mice since they are not sensitive to it, we removed the IR camera filter and replaced the integrated white power LEDs with 10 infrared ones. Visible light and infrared sensor—TAOS TSL2561: This sensor is able to convert a light intensity to a digital output of up to 16 bits of precision. It counts with two photodiodes, one for visible radiation
Electronics 2019,8, 34 8 of 16 and other for infrared radiation. It also has automatic ripple filtering at 50 or 60 Hz and supports I2C communication. Digital accelerometer—Analog Devices ADXL345: This small, very low power consumption accelerometer (0.1 µ A in standby mode and 40 µ A during measurement) is able to measure acceleration in three axes with at least 10 and up to 13 bits of precision. The measurement range is user adjustable and varies from ± 2 to ± 16 g. It allows to use serial peripheral interface (SPI) and I2C communication protocol. Barometric pressure sensor—Bosch BMP180: Based on piezoresistive technology, this sensor offers high accuracy, linearity, and long-term stability. It includes a temperature sensor to correct the pressure value, having a resolution of up to 0.01 hPa for pressure and up to 0.1 ◦ C for temperature. This sensor is capable of providing up to 128 measurements per second. Humidity sensor—TE Connectivity HTU21: This low-cost sensor is able to measure temperature in the − 40 to 125 ◦ C range and the relative humidity on a scale from 0% to 100%. Its accuracy is 0.04% for relative humidity (12 bits) and 0.01 ◦C (14 bits) for temperature. Buzzer: To implement sound alarms, a small piezoelectric buzzer was included. Its resonant frequency is 3 kHz, but using pulse width modulation (PWM), the pitch can be changed. Electronics2019,8,xFORPEERREVIEW8of16 accelerationinthreeaxeswithatleast10andupto13bitsofprecision.Themeasurementrangeis useradjustableandvariesfrom±2to±16g.Itallowstouseserialperipheralinterface(SPI)andI2C communicationprotocol. Barometricpressuresensor—BoschBMP180:Basedonpiezoresistivetechnology,thissensoroffers highaccuracy,linearity,andlong‐termstability.Itincludesatemperaturesensortocorrectthe pressurevalue,havingaresolutionofupto0.01hPaforpressureandupto0.1°Cfortemperature. Thissensoriscapableofprovidingupto128measurementspersecond. Humiditysensor—TEConnectivityHTU21:Thislow‐costsensorisabletomeasuretemperaturein the−40to125°Crangeandtherelativehumidityonascalefrom0%to100%.Itsaccuracyis0.04% forrelativehumidity(12bits)and0.01°C(14bits)fortemperature. Buzzer:Toimplementsoundalarms,asmallpiezoelectricbuzzerwasincluded.Itsresonant frequencyis3kHz,butusingpulsewidthmodulation(PWM),thepitchcanbechanged. (a) (b) (c) (d) (e) (f) Figure5.Centralprocessingunit(CPU)boardandselectedenvironmentalsensorsforthe prototypicaltransportbox:(a)UDOOQUADboard,(b)webcamandvisiblelight/infraredsensor,(c) digitalaccelerometer,(d)barometricpressuresensor,(e)humiditysensor,and(f)buzzer. 3.Results Astandardmailtransportboxwithcover,purchasedfromDeutschePost,wasselectedasthe containerforthesystemintegration.Thetransportboxweighed1.6kg,itsouterdimensionswere47 ×26.7×28cm,anditsvolumewas25L.Itwasmadeofimpact‐resistantandantibreakagematerial. Figure6showsthefunctionaltransportboxwiththeintegratedmonitoringsystem.Itincludedtwo platforms: Alowerplatformwasfixedtothebottomofthetransportboxandservedassupportforthecage whereanimalswerehousedduringtransport.Italsocontainedthe4×4matrixofEPICsensors. Anupperplatformwasfixedatthebottomoftheboxcover.ItincludedtheCPUboardandthe environmentalsensors. Figure 5. Central processing unit (CPU) board and selected environmental sensors for the prototypical transport box: ( a ) UDOO QUAD board, ( b ) webcam and visible light/infrared sensor, ( c ) digital accelerometer, (d) barometric pressure sensor, (e) humidity sensor, and (f) buzzer. 3. Results A standard mail transport box with cover, purchased from Deutsche Post, was selected as the container for the system integration. The transport box weighed 1.6 kg, its outer dimensions were 47 × 26.7 × 28 cm, and its volume was 25 L. It was made of impact-resistant and antibreakage material. Figure 6shows the functional transport box with the integrated monitoring system. It included two platforms:
Electronics 2019,8, 34 9 of 16 • A lower platform was fixed to the bottom of the transport box and served as support for the cage where animals were housed during transport. It also contained the 4 × 4 matrix of EPIC sensors. • An upper platform was fixed at the bottom of the box cover. It included the CPU board and the environmental sensors. Electronics2019,8,xFORPEERREVIEW9of16 Figure6.Functionaltransportboxwithintegratedmonitoringsystem.(Top)Backsideofthebox coverwithallhardwarecomponentsinstalled—CPUboardandenvironmentalsensors.(Bottom) MatrixofEPICsensorstocapturemousephysiologicalsignals. 3.1.MeasurementswithMatrixofEPICSensors ThissectionshowsexemplaryresultstovalidatetheuseoftheEPICsensormatrix.Weused MATLABsoftware(versionR2014b)toimplementalgorithmstoimprovethequalityofthecaptured physiologicalsignalsinordertocalculateHRandBR.Weimplementedanotchfiltertoremove50 Hznoise(anditsmultiples)andalow‐passfilter.Thiswasfollowedbyapolynomialsplinefitand subtractionfromtheoriginalsignal.Ifthesignalwasstillnoisy,weusedafifth‐orderpolynomial Savitzky–Golayfilterorgeneralizedmovingaveragetosmoothit.Timeandfrequencydomainplots weregeneratedaftereachoneofthesefilters. Figure7ashowsanexemplaryresultforacapturedrawphysiologicalsignalwhilethemouse hadonelimbincontactwithanEPICsensor.Figure7bshowstheresultafternotchfilteringtheraw signal. (a) Figure 6. Functional transport box with integrated monitoring system. (Top) Backside of the box cover with all hardware components installed—CPU board and environmental sensors. (Bottom) Matrix of EPIC sensors to capture mouse physiological signals. 3.1. Measurements with Matrix of EPIC Sensors This section shows exemplary results to validate the use of the EPIC sensor matrix. We used MATLAB software (version R2014b) to implement algorithms to improve the quality of the captured physiological signals in order to calculate HR and BR. We implemented a notch filter to remove 50 Hz noise (and its multiples) and a low-pass filter. This was followed by a polynomial spline fit and subtraction from the original signal. If the signal was still noisy, we used a fifth-order polynomial Savitzky–Golay filter or generalized moving average to smooth it. Time and frequency domain plots were generated after each one of these filters. Figure 7a shows an exemplary result for a captured raw physiological signal while the mouse had one limb in contact with an EPIC sensor. Figure 7b shows the result after notch filtering the raw signal. Electronics2019,8,xFORPEERREVIEW9of16 Figure6.Functionaltransportboxwithintegratedmonitoringsystem.(Top)Backsideofthebox coverwithallhardwarecomponentsinstalled—CPUboardandenvironmentalsensors.(Bottom) MatrixofEPICsensorstocapturemousephysiologicalsignals. 3.1.MeasurementswithMatrixofEPICSensors ThissectionshowsexemplaryresultstovalidatetheuseoftheEPICsensormatrix.Weused MATLABsoftware(versionR2014b)toimplementalgorithmstoimprovethequalityofthecaptured physiologicalsignalsinordertocalculateHRandBR.Weimplementedanotchfiltertoremove50 Hznoise(anditsmultiples)andalow‐passfilter.Thiswasfollowedbyapolynomialsplinefitand subtractionfromtheoriginalsignal.Ifthesignalwasstillnoisy,weusedafifth‐orderpolynomial Savitzky–Golayfilterorgeneralizedmovingaveragetosmoothit.Timeandfrequencydomainplots weregeneratedaftereachoneofthesefilters. Figure7ashowsanexemplaryresultforacapturedrawphysiologicalsignalwhilethemouse hadonelimbincontactwithanEPICsensor.Figure7bshowstheresultafternotchfilteringtheraw signal. (a) Figure 7. Cont.
Electronics 2019,8, 34 16 of 16 32. Laboratory Animal Science Association (LASA). Guidance on the transport of laboratory animals. Report of the Transport Working Group established by LASA. Lab. Anim. 2005,39, 1–39. [CrossRef] [PubMed] 33. Fawcett, A.; Rose, M. Guidelines for the housing of mice in scientific institutions. Animal Welfare Unit, NSW Department of Primary Industries, West Pennant Hills. Anim. Res. Rev. Panel 2012,1, 1–143. 34. Stryjek, R. A transportation device for rats. Lab Anim. 2010,39, 279. [CrossRef] [PubMed] 35. Zhan, H.; Tada, T.; Fujikura, E.; Matsumoto, K.; Tanaka, Y.; Hongo, K. A container for transporting small laboratory animals for magnetic resonance imaging. J. Neurosci. Methods 2005 ,144, 143–146. [CrossRef] [PubMed] © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).