Millanetal. Eye and Vision (2023) 10:32 https://doi.org/10.1186/s40662-023-00350-5 RESEARCH Open Access © The Author(s) 2023. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/. The Creative Commons Public Domain Dedication waiver (http:// creat iveco mmons. org/ publi cdoma in/ zero/1. 0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. Eye and Vision Spatio-chromatic vision withmultifocal diffractive intraocular lens Maria S. Millan1* , Laura Clavé1,2, Aurora Torrents1, Jesús Armengol1 and Fidel Vega1 Abstract Background This study aims to detect alterations in the spatio-chromatic pseudophakic vision produced by multifocal diffractive intraocular lenses (IOLs) and provides a physical interpretation. Methods In vitro characterization of the imaging performance of two diffractive IOLs: AT LISA Tri (Zeiss) and FineVision (PhysIOL) in on-bench model eye illuminated with red (R, 625 nm), green (G, 530 nm) and blue (B, 455 nm) lights. We used the metrics: energy efficiency (EE), area under the modulation transfer function, longitudinal chromatic aberration (LCA), and halo intensity. Through-focus (TF) analysis and calculation of the expected defocus curve under white (W) daylight were included. In vivo visual acuity (VA) of 50 pseudophakics (60 eyes) was assessed under W, R, G, B lights at far and near. Two clinical experiments evaluated LCA and R, G, B TF-EE effects on pseudophakic vision and their relative importance. Results Clinical mean VA values under W light agreed with the predicted values at far and near for both IOLs. LCA measurements and R, G, B TF-EE curves were consistent with their lens design based on the 0th and 1st diffraction orders operative for far and near vision, respectively. LCA effects were compensated at near but noticed at far (− 0.75 D under B light). We detected strong asymmetry in visual resolution depending on the object distance and the illuminating wavelength—red predominance at far, blue predominance at near—in consistency with the TF-EE measurements. Conclusions Diffractive multifocal IOL designs produce asymmetries in the spatio-chromatic vision of pseudophakics beyond the alterations strictly due to LCA. VA asymmetry for far/near object distance under R and B illumination is clinically detectable in subjects implanted with IOLs with 0th and 1st diffraction orders for far and near vision, respectively. Such VA asymmetry cannot be explained solely from the influence of defocus, as would be derived from a chromatic difference of power, but mainly from the wavelength dependence of the EE. Keywords Presbyopia-correcting intraocular lens, Spatio-chromatic vision, Multifocal intraocular lens, Diffractive lens, Visual acuity, Energy efficiency, Modulation transfer function, Longitudinal chromatic aberration Background Modern cataract surgery with intraocular lens (IOL) implantation can restore human vision far beyond the degradation produced by the loss of transparency of the natural lens. Significant improvements have been introduced in IOL designs to compensate for common refractive errors (such as myopia, hyperopia, astigmatism), age-related insufficiencies (such as loss of accommodation or presbyopia), and some image degradations (such as high-order and chromatic aberrations). Remarkable *Correspondence: Maria S. Millan [email protected] 1 Applied Optics and Image Processing Research Group, Universitat Politècnica de Catalunya-BarcelonaTech, C/ Violinista Vellsolà, 37, Terrassa, 08222 Barcelona, Spain 2 Mataró Hospital, Consorci Sanitari del Maresme, Barcelona, Spain
Page 2 of 15 Millanetal. Eye and Vision (2023) 10:32 scientific and technological advances in eye modelling, ray-tracing calculation, invivo biometry, adaptive optics and laboratory testing on optical bench, along with intensive clinical research, have led to define a number of optical metrics that correlate with postoperative outcomes and can even be used to predict the visual quality of the average patient after surgery [1–3]. This may help surgeons to make a more informed decision about the IOL to choose in a scenario of over one hundred designs and products. Moreover, patients can even experience prospective vision before undergoing surgery by means of visual simulators with active elements able to display the optical function of a given IOL [4]. Although optical bench testing and computer eye model simulations are very useful for understanding IOL performance in a range of observation distance, there is a gap between physics and perception that requires investigation for further insight. Thus, for example, while quite a few studies, data, and calculations are carried out under monochromatic light (typically in the green spectral region corresponding to the maximum photopic efficiency), human vision is mostly realized under white light. In this field, considerable attention has been paid to the longitudinal chromatic aberration (LCA), caused by the dispersive nature of materials, which produces a variation of the refractive optical power with wavelength. Even though human vision is highly tolerant to LCA in the presence of natural monochromatic aberrations (about 2.1 D of chromatic difference of refraction in the visible spectral range from 400 to 700nm) [5], LCA has gained new interest because it can be manipulated through IOL design, in particular, with the introduction of hybrid refractive-diffractive IOLs (hereafter, for the sake of brevity, referred to as diffractive IOLs). This type of IOLs is based on engraving a diffractive profile on at least one of the surfaces of a refractive lens, which is used as a carrier platform. The hybrid component is specifically designed to provide coaxial multifocality, and therefore, distinct vision for several foci. The focus of lower optical power allows for distance vision, while the focus of higher optical power allows for near vision. Since a multifocal IOL forms simultaneous images of the same object, subsequent neural adaptation is necessary to allow the subject to focus on the image of interest despite being overlaid by at least one out-of-focus image [6]. In any case, it involves contrast reduction and the potential appearance of perceptual dysphotopsia. Diffractive bifocals provide enhanced vision for far and near distances whereas trifocals are intended to further improve vision at intermediate distance. Depending on the diffraction orders involved in its multifocality, the diffractive component can mitigate the LCA produced by the dispersive nature of the IOL material and the ocular media [7, 8]. The joint compensation of LCA and corneal spherical aberration (SA) is advantageous for contrast sensitivity enhancement [9] and has been used to design a diffractive bifocal IOL of low addition for extended range of vision [10]. LCA in pseudophakic eyes has been intensively evaluated [11–14] and measured [15–18], but has barely gained enough clinical relevance, possibly because of the natural attenuation of LCA in normal human vision [5] and the increasing use of IOL materials with relatively low chromatic dispersion (e.g., the pseudophakic chromatic difference of refraction with an IOL Abbe number of 47 is similar to that of normal human eyes [11] and IOLs with even larger Abbe number are currently available). Another perspective of the issue leads us to consider the fraction of the incident energy deviated to each focus. The energy efficiency (EE) metrics accounts for the distribution of energy between the different foci and is typically featured for the design wavelength (546 ± 10nm as recommended by the International Standard Organization ISO 11979-2:2014) [19]. In diffractive components, both the optical power and the EE depend strongly on the wavelength. Since the multiple foci are coaxial, their positions, relative peak intensities, contrast, and blur turn out to be physically dependent on the wavelength. As a result, the optical image quality—evaluated through modulation transfer function (MTF)-based metrics— becomes wavelength dependent as well. Labuz et al. [20] detected the effects of a 580nm high-pass red filter on the visual acuity (VA) and the contrast sensitivity of patients implanted with a low add diffractive IOL. In comparison with white light viewing, the red filter did not improve far vision but had an adverse effect at the near and intermediate distances. Their MTF measurements showed that the diffractive IOL was “intermediately dominant in the blue light but far dominant in the red light”. The result reported by Labuz etal. [20] was, in fact, initial evidence to further motivate our study. The effects of the optical power and EE wavelength dependency showed by multifocal diffractive IOLs on the visual quality of pseudophakic subjects are still an open issue. In this cross-sectional study with laboratory investigation, we searched for the possible spatio-chromatic changes with the observation distance that pseudophakic vision may experience because of a physical fact: the wavelength dependency of the multiple images formed by a diffractive multifocal IOL. Should there be noticeable changes, their evaluation and potential consequences would contribute new knowledge to vision science. In addition, the understanding, interpretation, and description of their effects in terms of the optical features of the diffractive lens would be of practical interest for IOL designers and clinical practitioners.
Page 3 of 15 Millanetal. Eye and Vision (2023) 10:32 We show the impact of the spectral dependency of diffractive IOLs on the spatial and chromatic vision of pseudophakic eyes and give a physical rationale based on the optical design of the lens. We assess the effects of the wavelength dependency of EE and compare them with the effects of LCA for different observation distances using both invitro and invivo methods. To this end, we characterize IOL performance using well established laboratory techniques based on a model eye on optical bench (optical experiment) [7]. A VA expectancy in a variety of observation conditions will be computed from the optical quality results and compared with the actual VA outcomes of pseudophakic patients in similar conditions. This study aims to assess the joint effects of LCA and EE on pseudophakic vision (clinical experiment 1) and their relative importance. To achieve the latter, LCA will be compensated in a second clinical examination (clinical experiment 2) and the effects of the EE alone will be evaluated and discussed. Methods Optical setup The test bench with the model eye used to measure the optical performance of the IOLs in vitro is shown in Additional file1: Fig. S1. The setup, described in detail elsewhere [3, 21, 22], consists of three parts: the illumination system, the model eye, and the image acquisition system. A light emitting diode (LED) illuminated a test object placed at the front focal plane of a collimator (200mm focal length) to locate the object optically at infinity from the model eye. We used a set of three red (R), green (G), and blue (B) LED sources, with various spectral band emissions (Additional file1: Fig. S2 and TableS1) and two object tests (Additional file1: Fig. S1): a pinhole (200μm) for EE and halo measurements, and a four-slit pattern for MTF measurements. The slits were 10μm wide. The model eye, formed by an artificial cornea lens and a wet cell where the IOL was immersed, met ISO 11979-2:2014 (model eye type 2) recommendations [19]. The cornea lens was an achromatic doublet (Lambda-X, Belgium) intended for the evaluation of aspheric IOLs; it induced SA = + 0.16μm (in terms of the Zernike c[4,0] coefficient) for a 5.15mm pupil at the IOL plane. An iris diaphragm placed in front of the artificial cornea controlled the lens aperture. The pupil diameters mentioned in this work are referred to the IOL plane (hereafter named IOL pupil). The image acquisition system was composed of a 10×, infinity-corrected, plan-achromatic, microscope objective assembled to an 8-bit CCD camera, mounted on a high precision, threeaxis translation holder for through-focus (TF) analysis. The image acquisition unit (microscope and camera) was nearly diffraction-limited across the visible spectrum with a cut-off frequency of 555 cycles/mm. To reduce the impact of electronic noise, each image was the result of temporal averaging eight frames at a time. Metrics The basic metrics used for the optical characterization of the IOLs were the EE, the area under the modulation transfer function (MTFa) and the halo size and intensity. We considered two IOL pupils of size 3.0mm and 4.5mm. EE and MTFa were measured under separate R, G and B illumination, within a TF span of image vergence ranging from − 4 D to + 3 D, in 0.10 D steps. The origin of image vergence and defocus (0.0 D) was set at the distance image (highest MTF value at 50 cycles/mm) for the G light (530nm, close to the standard design wavelength of 546nm) [19]. The spatial frequency 50 cycles/ mm corresponds, in an eye of 17mm focal length, to 15 cycles/degree in the object space. Negative dioptric value corresponds to near vision vergence according to the clinical convention. The EE was computed through the light-in-the bucket [23] measurement of the pinhole image formed by the model eye (Additional file 1: Fig. S3a). Basically, the image core energy (Ecore) to the total energy (Etotal = Ecore + Ebackground) ratio approaches the light-inthe bucket value in the experimental practice [7]. LCA equalled the refractive power difference calculated from the distance between the R and B EE peaks in the image space. Positive sign was assigned to LCA when the power magnitude for the B light was higher than that for the R light, and the negative sign was assigned the converse. The MTF for a given image vergence within the TF range (Additional file1: Fig. S3b) was computed from the image of the four-slit test formed by the model eye with the IOL immersed, as reported elsewhere [2]. The MTFa was calculated by integrating the MTF curve in the spatial frequency range from 0 to 50 cycles/mm. The TFMTFa data have been used to calculate the expected VA and depth-of-focus as measured in clinics through postoperative defocus curves [1, 2, 24]. For the sake of a closer prediction to the postoperative VA (logMAR) defocus curves, we considered the chromatic characteristics of the W LED used in the clinical experiments of this study. From the chromatic coordinates of the R, G, B LEDs in the CIE 1931 system {R(0.7017, 0.2981), G(0.1224, 0.7478), B(0.1506, 0.0262)} (Additional file1: Fig. S2 and TableS1), we calculated the R:G:B power ratio that would generate the W light of chromatic coordinates (0.3128, 0.3292) (6500K daylight). We used only the chromatic data of the LEDs, not their relative intensities, because the R, G, B image channels were normalized separately prior to the MTF calculation i.e., adjusted to cover the grey level dynamic range of
Page 4 of 15 Millanetal. Eye and Vision (2023) 10:32 the camera sensor with no saturation. Following the procedure described by Hooi [25] and Huang etal. [26] we determined the R:G:B power ratio 4:10:1. This ratio provided the weight coefficients to calculate the linear combinations of the polychromatic metrics EEpoly and MTFapoly in the TF range and, hence, to compute the expected VA and postoperative depth-of-focus, which are figures of clinical interest [1–3, 24]. To this end, we used the mathematical expression VA =Aexp B∗MTFa poly +C , with A = 1.828, B =− 0.23, andC = 0.014 , which was found to reach the correlation coefficient R2=0.94 in a former work [3]. Halo was characterized from the image of the pinhole test, in the far and near foci of each IOL, under the R, G, B lights. Logarithmic scale of intensity was used for the sake of halo visualization (Additional file1: Fig. S4). Intraocular lenses (IOLs) We used two off-the-shelf 20 D trifocal IOLs: AT LISA tri 839 MP (Carl Zeiss Meditec AG, Jena, Germany) and FineVision Micro F (PhysIOL S.A., Liège, Belgium) for the invitro on-bench optical testing. Both IOLs are made of hydrophilic acrylic materials with 1.46 refractive index and 58 Abbe number. They are refraction-based in far vision, meaning they use the 0thdiffraction order, which has no associated power. They have a combination of two diffractive profiles, of different add power, that use their 1stdiffraction orders for intermediate and near vision. Both designs are pupil dependent since their diffractive profiles have different heights across the aperture: • FineVision: apodized combined diffractive profiles 1 (+ 1.75 D) and 2 (+ 3.50 D) on the entire optic surface (6mm diameter). Active diffractive orders: Far (0th order of the two profiles), intermediate (1st order of 1st profile), near (1st order of 2nd profile and, with little contribution, 2nd order of 1st profile). Posterior aspheric surface. The lens induces SA (c[4,0] = − 0.11µm for 6mm entrance pupil). • AT LISA tri: two zones in the aperture. Central zone (4.34mm): combined diffractive profiles 1 (+ 1.66 D) and 2 (+ 3.33 D) for trifocal imaging. Periphery (until 6mm): bifocal, far and near imaging. Active diffractive orders: Far (0th order), intermediate (1st order of 1st profile), near (1st order of 2nd profile). The lens induces SA (c[4,0] = − 0.18µm for 6mm entrance pupil). These trifocal IOL designs have been intensively studied in related research. The interested reader will find further information elsewhere [22, 27–29]. Clinical data Fifty patients (60 eyes), aged 49 to 74years, were classified into three groups (Table 1): FineVision group, AT LISA tri groups 1 and 2. Clinical experiment 1 was followed by the FineVision group and the AT LISA tri group 1, whereas clinical experiment 2 by the AT LISA tri group 2 (Fig.1). These clinical experiments constituted a cross-sectional study which followed the tenets of the declaration of Helsinki. Subjects were fully informed about the study and provided written consent. The ethics committees of the Hospital de Mataró (Consorci Sanitari del Maresme, Barcelona, Spain) and other collaborative centers approved the clinical study (CEIm 20/19 LIO2019). All examinations were carried out by a single experienced optometrist (LC) using the same material and procedures. Eligible patients presented bilateral cataracts and no comorbidities. They underwent symmetrical bilateral cataract surgery, meaning they were implanted with the same type of lens, using similar technique—phacoemulsification followed by IOL Table 1 Sample data of clinical experiments FineVision subjects were recruited from Presbit (Sabadell) (12) and Creu Groga (Calella) (8). AT LISA tri group 1 subjects were recruited from Presbit (Sabadell) (10) and Creu Groga (Calella) (3), Eurolaser (Mataró) (1), and Hospital de Mataró (Mataró) (6). AT LISA tri group 2 subjects were recruited from Hospital de Mataró (Mataró) (10). All centers are in the province of Barcelona, Spain IOL = intraocular lens; SD = standard deviation Implanted IOL group Subjects (Eyes) Age (years) mean ± SD (min, max) IOL power (D) mean ± SD (min, max) Pupil (mm) mean ± SD (min, max) Clinical experiment 1 AT LISA tri group 1 20 (20) 65.05 ± 5.28 (54, 74) 21.17 ± 2.92 (14, 26) 3.43 ± 0.34 (2.91, 4.12) FineVision 20 (20) 63.70 ± 5.23 (54, 72) 20.72 ± 3.06 (19.0, 23.5) 3.74 ± 0.60 (3.00, 5.04) Clinical experiment 2 AT LISA tri group 2 10 (20) 64.70 ± 6.72 (49, 73) 21.13 ± 3.20 (14, 27) 3.51 ± 0.39 (2.9, 4.2)
Page 5 of 15 Millanetal. Eye and Vision (2023) 10:32 implantation into the capsular bag in both eyes. Eye retinoscopy and subjective refractions were performed for all patients. Specific inclusion criteria were preoperative refraction error (spherical equivalent) less than ± 5.0 D, postoperative best distance corrected VA better than 0.1 logMAR, availability and willingness to comply with the examination procedures. Since the measurements were not conventional in both experiments, dedication and collaboration were requested from the recruited subjects. Key exclusion criteria were complications during or post-surgery, abnormalities in colour vision, prior ocular pathology, or ocular surgery, including refractive procedures. The examination was done between one and six months after surgery. The chart was placed at 3.5m for far VA assessment, so the object vergence of − 0.25 D was included in the manifest refraction (by inserting an ophthalmic lens of + 0.25 D in the trial frame) to adjust far vision measurements to infinity. Near VA was tested using the same optotype, placed also at 3.5m, and adding a negative lens to simulate the near distance. The power of the negative lens was determined according to the conditions established in each experiment. All measurements were taken monocularly with the natural pupil. IOLMaster (Carl Zeiss Meditec, Jena, Germany) was used for optical biometry measurements, postoperative pupil size was included. Fig. 1 Visual acuity assessment. a Sequential W, R, G, B illumination of the optotype chart; b Flowchart for clinical experiment 1; c Flowchart for clinical experiment 2. VA, visual acuity; W, white; R, red; G, green; B, blue
Page 6 of 15 Millanetal. Eye and Vision (2023) 10:32 The refractive correction obtained under W light was used throughout the assessment. The background luminance of the optotypes was 25.3 ± 0.1cd/m2, constantly controlled with a Mavolux 5032C photometer. The room was kept in mesopic conditions to avoid any interference with the measurements. A set of high contrast optotype charts were designed for the purpose of our investigation in accordance with the recommendations of the Universal Ophthalmological Council of 1984 [30] and the guidelines of the ISO 8596:2018 [31]. They are further described elsewhere [32] and were printed with highresolution quality. The stimulus size of the successive lines followed decimal progression in 0.1 steps. Unlike the logarithmic progression, this design—ISO 8596:2018 compliant—permitted smaller increments in the stimulus size, allowing us to detect finer VA variations [32] in the vicinity of 0.0 logMAR. The order of the background colour presentation was randomized, and subjects were only prompted once for each VA measure. The last visual level where the subject correctly called 3 stimuli out of the 5 presented in the same line was taken as the criterion for determining the VA grade (ISO 8596:2018) [31]. Decimal VA values were converted to the logMAR equivalent values for data processing, statistical analysis, and presentation. Although the observational data were oriented to obtain evidence of physical facts and the enrolment of a few subjects would have likely sufficed, we decided to recruit more subjects to conform groups of conventional size in this type of studies. Statistical analysis was performed using SPSS software version 13.0 (SPSS Inc., Chicago, IL, USA). Descriptive statistics—mean ± standard deviation (SD)—characterized the sample. The Kolmogorov-Smirnov test did not confirm the normal distribution of data. The Wilcoxon test was applied to paired data to assess the VA differences. For independent data, the Mann-Whitney U test was used. A difference was considered statistically significant for a P value less than 0.05. Results Optical experiment Figure2 shows the TF-EE and TF-MTFa results obtained with the 3.0mm pupil and the R, G, B lights, whereas Additional file1: Fig. S5 shows the same for the 4.5mm pupil. The polychromatic TF-EEpoly and TF-MTFapoly plots (Fig.2c and Additional file1: Fig. S5c), were computed from the experimental R, G, B TF-EE and TFMTFa curves weighted by the coefficients that would generate (6500K) W light. Figure2 shows that both IOLs have two clear foci, for far and near vision. Between them, EE and MTFa metrics decrease smoothly, yet with a certain trend of recovery for intermediate distances. The curves corresponding to the simulated polychromatic W light are very close to those measured under G (530nm) illumination. LCA was measured for each IOL (Fig.2 and Table2). Since the on-bench eye model had an achromatic doublet for the artificial cornea, the LCA values can be considered as due to the IOL. LCA was very small and hardly measurable in the far focus of the two IOLs, but it exceeded 1.0 D (negative) in the near focus. Quite importantly for our study, the distribution of EE between the lens foci changes remarkably with wavelength (Fig.2, left column), as it can be expected from the optical path differences introduced by the diffractive step height in wavelengths other than the design [8, 33]. Thus, the R and B curves of the TF-EE differ clearly in opposite directions from the G curve: while the R light greatly benefits the far focus to the detriment of the near, the B light benefits the near focus at the expense of the far. The wavelength dependence of both the optical power and EE influences the contrast and size of the simultaneous images formed at the focal planes, with only one image being focused at any one time with the rest being out-offocus. This effect is illustrated in Fig.3 for the AT LISA tri. In the far focus (bottom row), the R image shows the best contrast (highest intensity in the image core and lowest intensity in the surrounding halo), followed by the G image and the B image, the latter showing the worst contrast. In the near focus (top row), however, the energy distribution is the opposite: the R image shows the worst contrast, closely followed by the G image, and the B image shows the best. Moreover, the size of the haloes, determined basically by the out-of-focus images [34], depends on the addition power existing between the far and near foci, which, in turn, depends on the wavelength. Thus, the largest halo corresponds to the highest add power (3.9 D, R light; Fig.3), whereas the smallest halo corresponds to the lowest add power (2.6 D, B light). We calculated the expected VA (logMAR) of the pseudophakic patients under W illumination from the TF-MTFapoly (calculated, in turn, from the R, G, and B TF-MTFa measurements, Fig.2) taken with a 3.0mm pupil [3]. Figure4 shows the expected defocus curves (blue line) for the AT LISA tri and FineVision IOLs. They predict a very good postoperative VA, close to 0.0 logMAR in far vision, which decreases smoothly in intermediate vision with some improvement in near. Defocus is represented at the spectacle plane [35] in Fig.4. Taken together, a sustained good visual quality (equal or better than 0.2 logMAR) can be expected for the average patient in a depth-of-focus range that goes from infinity to roughly 30cm from the subject (− 3.0 D defocus). This prediction is intended for comparison with the actual clinical VA outcomes as explained in the clinical experiments.
Page 7 of 15 Millanetal. Eye and Vision (2023) 10:32 Fig. 2 TF-EE and TF-MTFa measurements obtained in the laboratory experiment for the trifocal diffractive AT LISA tri (a) and FineVision (b) IOLs under R, G, B lights and 3.0 mm pupil; c Polychromatic TF-EE and TF-MTFa curves of both IOLs. TF-EE, through-focus energy efficiency; TF-MTFa, through-focus area under the modulation transfer function; R, red; G, green; B, blue; IOL, intraocular lens
Page 8 of 15 Millanetal. Eye and Vision (2023) 10:32 Clinical experiment 1 VA of pseudophakic subjects was tested at two fixed distances—far (0.0 D) and near (− 3.0 D), under successive W, R, G, and B illumination (Fig.1a, b). Figure5 shows the average VA outcomes obtained for the FineVision group and the AT LISA tri group 1. The VA results include the joint effects of the EE wavelength dependence of the diffractive IOL and the LCA of the pseudophakic eye. The mean VA outcomes are consistently similar for both trifocal IOLs under all four illuminations although slightly better for subjects with AT LISA tri. Despite the presence of LCA, the VA reached under W light was equal or better than any other colour light in both the far and near vision conditions. The (mean ± SD) values and their statistical significance are given in Tables3 and 4, respectively. The mean VA values with W illumination are represented with yellow dots in Fig.4. It is worth remarking the excellent agreement with the predicted values for the AT LISA tri group. The prediction for the FineVision group slightly overestimated the clinical results. In general, the mean VA values are better in far than in near vision for all the illumination conditions, except for the B light (Fig.5 and Additional file1: Fig. S6). Moreover, at near distance, the VA with B illumination is as good as the VA achieved with W light, with nonstatistically significant difference for both the AT LISA tri (P = 0.57) and FineVision (P > 0.99) groups (Table4). With R illumination, the visual quality worsens severely in near vision, more than with W and G lights (Fig.5, Table3, and Additional file1: Fig. S6). Note that the good VA achieved at far distance under R light (0.10 ± 0.06 logMAR for AT LISA tri, 0.13 ± 0.06 logMAR for FineVision) drops off dramatically to the worst VA at near (0.38 ± 0.10 logMAR for AT LISA tri, 0.44 ± 0.08 logMAR for FineVision), even worse than the poor VA outcomes obtained at far distance under B light (0.32 ± 0.10 logMAR for AT LISA tri, 0.37 ± 0.09 logMAR for FineVision). Clinical experiment 2 This experiment aims to bring to light the effects of two separate factors: one, the EE wavelength dependence of the IOL foci (determined by the operative diffractive orders) and, the other, the LCA of the pseudophakic eye. We want also to evaluate their relative influence on either far and near vision, as well as to emphasize the possible differences in comparison with the natural phakic human vision. The experiment consisted of four stages (Fig.1c) in the VA assessment: two concerned far vision with uncompensated (uc) LCA (stage 1) and compensated (c) LCA (stage 2), and the other two concerned near vision with uncompensated (uc) LCA (stage 3) and compensated (c) LCA (stage 4). For the LCA compensation of every Table 2 Longitudinal chromatic aberration for the far and near IOL foci, obtained from the through-focus energy efficiency values of Fig. 2 (3.0 mm pupil) The spectral range covers from the blue to the red light emitting diode lights (455 to 625nm) Intraocular lens Longitudinal chromatic aberration Far focus (D) Near focus (D) AT LISA tri 0.00 ± 0.10 − 1.20 ± 0.10 FineVision 0.10 ± 0.10 − 1.12 ± 0.10 Fig. 3 Red (R), green (G), blue (B) images of a pinhole test at near and far foci for AT LISA tri, 3.0 mm pupil. Energy efficiency (EE) values are provided. In the near focus (top row), the add power (D) is given for R, G, B lights. Front halo images and their halo profiles (right most panel) are presented in logarithmic scale of intensity for the sake of visibility
Page 9 of 15 Millanetal. Eye and Vision (2023) 10:32 subject in each illumination condition, we used additional trial ophthalmic lenses, with the (± 0.25 D) uncertainty assumed in ordinary clinical examinations. Note that stage 1 coincided with the first part of the experiment 1, but it was applied to a new group of subjects (namely, AT Lisa tri group 2). In stage 3 (near vision with uncompensated LCA), we determined the refractive addition for the best near distance corrected VA under W light and this refractive addition was kept unchanged with the other illuminations. LCA was individually compensated under R, G, B lights at far vision in stage 2 and at near vision in stage 4. For the sake of comparison, we present the results (mean ± SD) of the VA outcomes for the AT LISA tri groups 1 and 2 in Fig.6 and Table5. The VA values of the AT LISA tri group 1 were obtained with uncompensated (uc) LCA and constant addition of − 3.0 D at near vision. In far vision, the AT LISA group 2 (uc) showed, on average, slight hypermetropia under R light (0.26 ± 0.15 D) and moderate myopia under B light (− 0.75 ± 0.11 D). This chromatic difference of refraction was due to the LCA of the pseudophakic eye as a whole and was still similar to the natural LCA of a phakic eye [5]. After proper correction with ophthalmic lenses, they Fig. 4 Expected visual acuity (VA, logMAR) defocus curves (blue lines) of patients implanted with AT LISA tri (a) and FineVision (b) IOLs under (6500 K) white LED illumination. Actual clinical assessments (mean ± SD) of pseudophakic patients with best distance correction are represented by dots and error bars. In experiment 1, patients enrolled in the AT LISA tri group 1 and FineVision group were assessed in far and (− 3 D) near vision (yellow dots). In experiment 2, patients of AT LISA tri group 2 were further assessed in far and best near vision (red crosses). IOL, intraocular lens; LED, light emitting diode; uc, uncompensated longitudinal chromatic aberration