Possible strategy to use differential mobility spectrometry in real time applications

Anttalainen, Osmo,Puton, Jaroslaw,Kontunen, Anton,Karjalainen, Markus,Kumpulainen, Markus,Oksala, Niku,Safaei, Zahra,Roine, Antti

Full text

ORIGINAL RESEARCH Possible strategy to use differential mobility spectrometry in real time applications Osmo Anttalainen 1 &Jaroslaw Puton 2 &Anton Kontunen 3 &Markus Karjalainen 3 &Pekka Kumpulainen 4 & Niku Oksala 1,4 &Zahra Safaei 5 &Antti Roine 1 Received: 13 March 2019 /Revised: 13 March 2019 /Accepted: 31 May 2019 #The Author(s) 2019 Abstract Differential Mobility Spectrometry (DMS), also called as FAIMS is a variation of atmospheric pressure ion mobility measurement techniques and is capable of providing information about the electric field - mobility dependence of ions. In this method, a combined electric field is used. This field consists of asymmetric oscillating electric field of high intensity and low static field component. Analytical information in DMS is 2-dimensional dependence of ionic current on oscillating field amplitude and the value of static field intensity. The measurement of DMS signal for whole ranges of both variables is time consuming and also generates lot of data. It is a disadvantage of DMS method, which limits the use of this otherwise powerful technology in real time applications that require a response time of few seconds. This paper presents a way to limit measurement time by heuristic knowledge of the properties of the data space and another method based on the concept of Shannon Entropy to find operating parameters satisfying both separation and signal to noise ratio requirements. Keywords Differential mobility spectrometry .DMS .FAIMS .Shannon entropy Introduction The purpose of spectrometric methods is to produce knowledge about the object under study by separating the sample matrix into the informative prime components. The spectral components, often manifested as intensity peaks or their combinations are used to reveal some descriptive feature or to detect such. The usefulness of a method is determined by its analytical power and time and efforts invested to sample preparation, measurement time and signal processing to obtain the desired output. Despite continuous improvement of technology there is no single ultimate method or instrument that provides both extreme analytical power and ease of use in the area of applications of analytical chemistry. Compromises between *Osmo Anttalainen [email protected] Jaroslaw Puton [email protected] Anton Kontunen [email protected] Markus Karjalainen [email protected] Pekka Kumpulainen [email protected] Niku Oksala [email protected] Zahra Safaei [email protected] Antti Roine [email protected] 1 Centre for Vascular Surgery and Interventional Radiology, Tampere University Hospital, Tampere, Finland 2 Institute of Chemistry, Military University of Technology, 00-908 Warsaw, Poland 3 Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland 4 Department of Surgery, Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland 5 Department of Green Chemistry, School of Technology, LappeenrantaLahti University of Technology LUT, Mikkeli, Finland International Journal for Ion Mobility Spectrometry https://doi.org/10.1007/s12127-019-00251-1 analytical performance and overall process time, cost and efforts, are common. Field Asymmetric Ion Mobility Spectrometry (FAIMS) and Differential Mobility Spectrometry (DMS) are atmospheric pressure techniques to separate ions with a high-frequency amplitude-asymmetric strong electric field [1] based on nonlinear velocity vs electric field characteristics of ion species. Whereas Ion Mobility Spectrometry in its various embodiments [2] has emerged to analytical applications [3,4]such as explosive detection [5], breath [6,7], pharmaceutical [8], food analysis [9] and detection of hazardous chemicals [10], only a few commercial stand-alone DMS based applications exist on the market [11–13]. The possible explanations of low popularity of DMS instruments are technical, since the realisation of the required asymmetric electric field with proper characteristics is a complex engineering challenge [14,15]. Although the spectra analysis time can be within a range of one second in some cases [16], the measurement time is a function of the desired parametric resolution. The output of a DMS measurement is a 2D-signal, which is called as dispersion plot. Creating a high resolution DMS dispersion plot can be time consuming, potentially lasting several minutes. This paper introduces one possible strategy to reduce the measurement time in order to enable applications such as real time detection of biomarkers in surgical smoke. Measurement time in real life DMS The separation of ions in DMS is based on the nonlinear electric field transport characteristics of the ions or ion clusters [1]. In DMS ions are transported with laminar gas flow between two parallel electrodes, which create time dependent asymmetric, transverse electric field whose intensity changes in time in such a way that time integral of the electric field is zero. The ideal shape of voltage waveform generating the electric field is a rectangular wave with duty cycle of 0.1..0.3. Therefore, in the part of the waveform period, electric field intensity is distinctly higher than for the rest of the period. The^high field^phase is often called separation field and the voltage that generates this field is called separation voltage U SV . The alternating electric field causes ions to oscillate between the electrodes and their net movement in the channel depends on the mobility difference between high and low field. The field dependent mobility is described as K(E/N)=K(0)(1 + α(E/N)) [17], where k(0) refers to mobility in low electric field and α(E/N) describes the field dependent component of the mobility. If the net movement in between the electrodes is not zero, ions move towards the electrode, lose their charges and neutralise. With a small superimposed DC field, created by and referred as compensation voltage U CV , the net movement can be adjusted to zero and ions can be transported with a flow between the electrodes. The DMS can be imagined as a filter which allows only portion of the ions of specific α(E/N) to pass at selected U SV and U CV values. The separation capability and also the signal intensity and thus signal-to-noise ratio depends on the geometry of the filter and the separation field strength. To obtain understanding of an ion’s behaviour in high field or detect existence of some ion species, both U SV and U CV are commonly scanned with discrete steps. Ionic current registered for these values constitutes the dispersion plot such as shown in Fig. 1. Dispersion plots are commonly used to visualise the measurement result. To measure one ionic current signal for one pair of U SV and U CV values of the dispersion plot, following steps need to be taken: –U SV is set to adjust the strength of an asymmetrically oscillating electric field Fig. 1 Left: An example of a dispersion plot where horizontal axis represents U CV ,verticalaxisU SV and colouring is based on ionic intensity. Right: Differential mobility spectra extracted from U SV values denoted with A, B and C. Horizontal axis is U CV , vertical axis ionic intensity Int. J. Ion Mobil. Spec. –U CV is adjusted to create DC field –Ions are transported between the planar channel with electrodes by flow –Number of surviving ions are measured with static electric field with amplifiers connected to electrode pair after the DMS filter –Measured signal is processed and stored into the measurement memory Normal strategy is to scan U CV over potential compensation value range while keeping U SV constant. After a complete U CV scan, U SV is stepped between given U SV range and this process is repeated until all desired U SV values are scanned. The measurement time of this process can be estimated as tscan ¼nSV*t SV þnCV*maxt flow;tCV ðÞþtsample þtspt  ð1Þ where –t scan is the time required for the complete scan –n SV is the number of separation voltage U SV steps, typically 10–100 –t SV is the setup time for separation voltage U SV ,typically 10–500 ms –n CV is the number of compensation voltage U CV steps, typically 100–1000 –t flow is the time required to get new steady state with new U SV and U CV parameters, typically 1–3 ms depending on the geometry and amount of the flow –t CV is the setup time for the compensation voltage, typically 0.1–2ms –max(t flow ,t CV ) is used because the next U CV can be set as soon as analog digital conversion (ADC) sampling is completed –t sample is the time required for the analog to digital conversion, including the number of measurement repetitions for digital filtering, typically 0.1–5ms –t spt is the time for signal processing required before next measurement step can be taken, 1–5ms The practical values depend strongly on the need of U SV and U CV resolution, desired signal to noise ratio, and hardware configuration and limitations. Using typical worst case values the t scan can end up to more than 20 min, which may well be suitable for laboratory measurements but is unpractical for real time measurements requiring response time in the range of seconds. To tackle the measurement time, it’spossibletolimit the number of U CV steps but this option is limited itself by the spectral peaks in the obtained spectrum. The peak position is an indication of ion type and too coarse selection of U CV will affect correct peak position detection. One option would be to cut down t sample . The sample rate of the ADC can be in the range of 100,000 samples per second and the majority of tsample time is contributed by the filtering requirements and limiting the filtering would sacrifice the signal-to-noise ratio. The parameter t flow can be varied, but enough residence time is necessary to ensure stabilisation of ionic current at the output of separator after changing the U CV . Also, using higher flow will reduce t flow and is seen as increased signal intensity, but it will also have a negative impact on peak width and thus, separation capability. It is notable that U SV can be set as soon as sampling of previous U SV -U CV pair is completed and therefore readiness for next sample acquisition is possible as soon as both t sample and t flow are completed. High voltage setup time t SV is fixed by the design of the high voltage source and tend to increase with U SV step size, because of the time needed to charge or discharge capacitances in the high voltage source. The signal processing time t spt is not a limiting factor for the measurement and digital devices such as modern embedded controllers or signal processors. The digital filtering and data transfer to the memory can be completed in milliseconds or less and the measurement cycle time depends more on other tasks the processor needs to execute. The most effective way to limit the total measurement time t scan is to limit the number of U SV ,U CV -pairs in the scan and thus find a way to limit n SV or n CV . Example of measurement time estimation The measurement results from DMS can be displayed as a heat map, where x-axis represents the compensation voltage U CV , y-axis the separation voltage U SV and the intensity of the signal is colour coded. This type of data presentation is visually appealing, informative for the user and is an outcome of the natural way of parameter scanning in DMS. An example of this kind of graph called a dispersion plot is shown in Fig. 1. For illustration purposes, using Owlstone V-OVG gas calibrator and a self-made permeation tube, low ppb-level heptanone sample was generated and mixed with small amount of water vapour. Concentration levels of heptanone and humidity were experimentally adjusted to demonstrate typical structure of dispersion plot in which reactant ions (hydrated protons) and two kinds (protonated monomer and proton bound dimer) of sample ions are visible. The measurements were performed and data were collected with ENVI-Analyzer manufactured by Environics Oy, Finland. The DMS was operated with square wave pulses at frequency of 250 kHz and duty cycle of 10.3%. The U CV resolution of the data was set to 200 equally distributed values between −1.89 V to 11.8 V. For U SV , the resolution was set to 45 equally distributed values between 240 V to 722 V. The total number of acquisition points in the scan was 9000. Each point in dispersion plot is an average of samples measured over 20 ms sampling period. The exact number of samples is not known because instrument’s program cycle and load time may affect to analog-toInt. J. Ion Mobil. Spec. digital conversion interval. The device collected the full 9000 dispersion plot points for positive and negative ions separately. The total time required to obtain scans for both polarities was 394 s, giving 21.9 ms estimate for single U SV U CV point measurement time t USV,UCV . This time may be short or long compared to some other instrument and should be considered here only as an illustrative measure. Anatomy of the dispersion plot A common reason to scan ionic current over U SV and U CV ranges is to get information to characterise ion’s high field properties or use the data for detection purposes. Figure 1 shows colour mapped coding from the two dimensional scan of a DMS measurement. The ion species, heptanone dimer, monomer and positive reactant ions, RIP are seen as fingers or branches shown against blue background. The background presents the floor of the intensities, which consists mainly of low level noise. Common sense interpretation of the figure indicates that interesting data lies in such branches of the dispersion plot in which the branches of the spectrum are well separated. Not all data in the dispersion plot are useful. Figure 1shows U CV scans extracted from the dispersion plot with fixed U SV values. These scans are named differential mobility spectra. With low U SV (Fig. 1right panel, A) ions are not separated. Mid range U SV (Fig. 1right panel B) reveals three peaks, from left to right: heptanone dimer, heptanone monomer and RIP. Figure 1, right panel, C shows the highest U SV . The RIP is no more visible and only the dimer and monomer peaks are seen. The peak intensities in Fig. 1A are highest but separation is poor and peaks in C are the lowest while separation of the two peaks is the best. The measurement of intensities along the branches in the dispersion plot in Fig. 1are shown in Fig. 2. The intensity is reduced as a function of U SV . The loss is caused by the reduction of an effective gap, when increased electric field intensity increases oscillation length, thus increasing contacts of ions to electrodes causing charge loss. This is discussed in detail in [18]. Fig. 2 Change of the peak intensity vs. separation voltages for heptanone dimer, monomer and RIP Int. J. Ion Mobil. Spec. Dispersion plot sub sections Based on the data shown in Figs. 1and 2the dispersion plot is annotated to descriptive sections shown in Fig. 3. The sections are: –A section, where branches caused by ion peaks are visually separated and well above noise level –B section, where branches are very close to noise level or have disappeared (see 1C) –C section, where no peak exists because no ion will pass the DMS filter with these U SV ,U CV values –D section, where almost no dispersion of ions can occur due to too low separation field (see 1, A) By intuition, the most useful measurement area lies in section A of Fig. 3. Considering that the measurement could be limited to the section A only, the potential savings of measurement time can be estimated from the surface areas of A compared to the full area. By arbitrary selection based on Fig. 2of corner points for (−1.89,300), (−1.89,600), (9600) and (2.5300) described by pairs of U CV and U SV values, the proportion of area A to the full area is 0.42, therefore measurement time could be decreased in the same proportion. Although the hardware would support measuring only this kind of section, the measurement time would still be too long for real time applications. Further time reduction could be possible if only a single differential mobility spectra was used. Assuming that relevant spectra could be found from section A, limited by its boundaries, the reduction of data could be dramatic. It can be estimated from Fig. 1, section B in right panel, that perhaps only half of a single spectrum data is necessary to be able to properly separate the three peaks in this example. Because the selection of U SV impacts to n CV , the next question is how to find the most representative U SV with good separation and signal-to-noise ratio. Entropy Shannon entropy - the definition and basic properties The theory of information contains tools for validation of the information content of data. One of the most important and known parameters is the entropy of the signal known as Shannon entropy H(x) [19], which can be defined as HxðÞ¼−∑N i¼1pilogkpi ðÞ ð2Þ where p i is the probability of obtaining the value x i of variable x and N is the number of possible values of a variable x. In signal processing the probability is represented usually by signal intensity measured as a function of some independent variable (e.g. frequency, wavelength, number of measurement, etc.). The logarithm base, k is usually equal to 2. For sequence of signal measurements, where results are distributed between discrete values, the Shannon entropy is zero, when all results have the same value. Shannon entropy increases when uniformity of distribution increases and will get maximum, when signal is uniformly distributed over the possible values. One way to interpret Shannon entropy is to see it as a measure of the quality of signal distribution. Shannon entropy and its minimum have been applied recently for liquid chromatrography - mass spectrometry (LCMS) [20]toselect Fig. 3 Descriptive area division in dispersion plot Int. J. Ion Mobil. Spec. the best mass-spectrum. The basis of this application is explained by Chatterjee et al. [21]. In this work the probability was estimated as ratio of the signal value for each point to the sum of signal intensities for all points of the spectrum. In our work another kind of approach for calculation of probability was selected. It was assumed that the result of signal intensity measurement in DMS is a random variable distributed between N intervals of equal value in such a way that ΔS¼Smax−Smin Nð3Þ where S max to S min is the range of possible values of DMS signal and S is the width of signal interval. Assuming that logarithm base k in (2) is equal to 2, total number of samples is n s and number of counts for i th signal interval is n i . For such approach the probability needed for calculation –the entropy can be estimated as pi¼ni ns. Then H¼−∑N i¼1 ni ns log2 ni ns  ð4Þ In the case of fully concentrated noiseless distribution ni =n s for i=jand n i = 0 for i≠j. Then, the entropy calculated with Eq. (4) is equal to 0, and it is minimum possible value of this parameter. In case of uniform distribution, the probability of finding signal in each interval is pi¼1 Nand thus the formula for calculation the entropy becomes HU¼−∑N i¼1 1 Nlog2 1 N  ¼log2NðÞ ð5Þ It can be proved that the value calculated with the formula (5) is the maximum value of entropy. The DMS dispersion plot can be considered as a matrix, which contains experiment vectors with a length of n CV for each U SV value. Each vector element is the intensity of the signal. In our approach each vector is considered rather as a set of random variable values than a spectrum. For high U SV values, i.e., for experimental conditions when ions do not pass through the DMS filter (section B in Fig. 3) the signal is the baseline with characteristic noises. This represents the lowest possible entropy. Appearing of peaks in the differential mobility spectrum for lower values of U SV cause distribution to expand towards larger uniformity, which will be seen as increased entropy. The low U SV values result to minimal or no separation and only for few U CV values the signal intensities differ from noise level, making the noise a major contributor for the entropy. Most of the ion related peaks exist between low and high U SV and thus the maximum of entropy sets somewhere between the low and high U SV . Entropy estimation The intensity probabilities were obtained with function hist() from graphics package of R Studio (version 3.5.1). The results of entropy values were calculated with Eq. 4for each value of U SV rows in dispersion plots measured for the heptanone. Measurements were performed with duty cycles 5.25% and 10.25% and are shown in Fig. 4. The entropy maximum for lower duty cycle was found at the higher limit of U SV range and entropy maximum with higher duty cycle is in mid-range of used U SV range. The difference can be explained on the basis of dispersion plots: The lower duty cycle contributes lower loss of the signal but needs higher field strength to cause dispersion, because the pulse length of U SV is shorter and thus entropy value increases as long as loss of ions becomes meaningful. When duty cycle is increased the dispersion gets stronger, but the loss increases as well and the maximum point is met with lower U SV than for lower duty cycle. The decreasing tendency of maximum entropy U SV is shown in Fig. 5. Altogether 202 experiments were completed with similar chemical setup as described before, but duty cycle Fig. 4 Left: Example of entropy as indicator of separation selection tool with duty cycle of 5.25%. A1: Dispersion plot, B1: Entropy vs separation voltage USV, C1: Differential mobility spectrum measured at separation voltage USVequal to 582 V for which maximum entropy was achieved. Right: Example of Entropy as indicator of separation selection tool, with duty cycle of 10.25%. A2: Dispersion plot, B2: Entropy vs separation voltage USV, C2:Differential mobility spectrum measured at separation voltage USV equal to 431 V for which maximum entropy was achieved Int. J. Ion Mobil. Spec. was stepped between 5 and 10.25%, each 9–11 times. The automated test run took about 24 h and there was some variation in water vapour concentration as well as in ambient pressure during the test, which may explain the large dispersion of single experiments results. This, somewhat unexpected dispersion of entropy maximum locations, can actually be used to determine the width of operative U SV window, when some parametric noise such as variation in sample concentration is expected. Discussion The original problem presented in this paper is to reduce the time needed for the measurement. The ENVI-Analyzer instrument used to collect heptanone data is used here as an example. The device allows script based control of U SV and U CV . The range and number of steps within a range and set-up delays can be set independently. For given U SV , also frequency and duty cycle can be defined. The averaging time for single intensity measurement can be defined as a parameter. The practical scanning speed for the used hardware is limited by the transients related to U SV and U CV step size and the noise level of the intensity measurement. The fluctuations of the intensity signal are related to characteristic noises of transimpedance amplifier used for measurement of the ion current, and electrical noise sources within the instrument and variations of the gas flow. The instrumental nonidealities define the averaging time and other parameters, which were found experimentally. In the examples shown in the Fig. 4, the ionic peaks are located within about 6 V U CV span. With the used instrument settings, this corresponds to about 100 U CV values and equals to 2.2 s measurement time. The reduction from original 9000 sample points to the proposed 100 points must be understood as a demonstration of measurement time reduction potential for limited type of applications rather than a generic result. DMS instruments are commonly used to scan over U SV ,U CV space to gather dispersion data, which is time consuming regardless of the hardware design. Applications aiming to detect or identify some set of chemicals need large amount of experimental trials for library creation. When instrument is used with the library of spectra or peak positions, the detection or identification speed depends on the speed of sample measurement - the scan time - and the algorithm how the sample data is evaluated against the experimental data. The entropy can be calculated from experimentally collected data sets during the library creation process and used to find candidate sets of U SV values which work well in indicative detection based on the peak positions. The fine resolution dispersion data can provide a lot of extra information compared to simple peak position detection and therefore, advanced measurement parameter set and detection algorithms could follow the indicative step. This two step approach would lead to improvement in the first stage detection time, especially if also the U CV span can be limited as illustrated within Fig. 3. Conclusion Our test data was limited to single chemical example, which does not predict the location of entropy maximum in a case of different analyte or in the case of complex multichemical matrix. When creating a database for detection applications Fig. 5 USV at maximum entropy with different duty cycles. The solid triangles shows the mean USV at a given duty cycle. The line connecting triangles is only to clarify the visualisation Int. J. Ion Mobil. Spec. containing large number of different conditions, it is likely that U SV for entropy maximum varies per sample. However, it seems possible that the knowledge of DMS dispersion plot characteristics and Shannon Entropy can be used to find section of DMS dispersion data and thus U SV and U CV parameters which provide good separation on ions together with good signal-to-noise ratio. Narrow dispersion plot, optimally single U SV results to short measurement time. Rapid measurement benefit applications which utilise this data to trigger more exhaustive data measurement and analysis algorithms. Using entropy as an indicator can be envisioned for parameter optimisation for hyphenated systems such as GC-DMS. The results shown here are early proof-of-concept and more experimenting is needed to find out the true potential of this concept. These experiments, as well as some theoretical considerations, should also enrich our knowledge of the properties of entropy as a measure of the quality of the spectra in DMS. We treat our work as the first step in this direction. Acknowledgements This study was financially supported (or partly supported) by the Competitive State Research Financing of the Expert Responsibility area of Tampere University Hospital (9 s045, 151B03, 9 T044, 9 U042, 150618, 9 U042, 9 V044 and 9X040); by Competitive funding to strengthen university research profiles funded by Academy of Finland, decision number 292477; and by Tampereen Tuberkuloosisäätiö (Tampere Tuberculosis Foundation). Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http:// creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. References 1. Buryakov IA, Krylov EV, Nazarov EG, Rasulev UK (1993) A new method of separation of multiatomic ions by mobility at atmospheric pressure using a high-frequency amplitude-asymmetric strong electric field. Int J Mass Spectrom Ion Process 128(3):143–148 2. Cumeras R, Figueras E, Davis CE, Baumbach JI, Grcia I (2015) Review on ion mobility spectrometry. Part 1: current instrumentation. Analyst 140(5):1376–1390 3. Cumeras R, Figueras E, Davis CE, Baumbach JI, Grcia I (2015) Review on ion mobility spectrometry. Part 2: hyphenated methods and effects of experimental parameters. Analyst 140(5):1391–1410 4. Armenta S, Alcala M, Blanco M (2011) A review of recent, unconventional applications of ion mobility spectrometry (ims). Anal Chim Acta 703(2):114–123 5. Ewing RG, Atkinson DA, Eiceman GA, Ewing GJ (2001) A critical review of ion mobility spectrometry for the detection of explosives and explosive related compounds. Talanta 54(3):515–529 6. Perl T, Carstens E, Hirn A, Quintel M, Vautz W, Nolte J, Jnger M (2009) Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry. Br J Anaesth 103(6):822– 827 7. Vautz W, Nolte J, Fobbe R, Baumbach JI (2009) Breath analysisperformance and potential of ion mobility spectrometry. J Breath Res 3(3):036004 8. ODonnell RM, Sun X, Harrington P d B (2008) Pharmaceutical applications of ion mobility spectrometry. TrAC Trends Anal Chem 27(1):44–53 9. Karpas Z (2013) Applications of ion mobility spectrometry (ims) in the field of foodomics. Food Res Int 54(1):1146–1151 10. Mkinen MA, Anttalainen OA, Sillanp MET (2010) Ion mobility spectrometry and its applications in detection of chemical warfare agents. Anal Chem 82(23):9594–9600 11. Chemical and biological detection & identification chemring - sensors & electronic system. http://www.chemringsensors.com/ capabilities/defense/chemical-and-biological-detectionandidentification. (Accessed on 01/16/2019) 12. Dms - breath analyzer —draper. https://www.draper.com/exploresolutions/dms-breath-analyzer#dmsbreath 13. Lonestar voc analyzer. https://www.owlstonemedical.com/ products/lonestar/. (Accessed on 01/16/2019) 14. Krylov EV, Coy SL, Vandermey J, Schneider BB, Covey TR, Nazarov EG (2010) Selection and generation of waveforms for differential mobility spectrometry. Rev Sci Instrum 81(2):024101 15. Papanastasiou D, Wollnik H, Rico G, Tadjimukhamedov F, Mueller W, Eiceman GA (2008) Differential mobility separation of ions using a rectangular asymmetric waveform. J Phys Chem A 112(16):3638–3645 16. 19212 lonestar slick sheet2send. https://www.owlstoneinc.com/ media/uploads/files/Lonestar_2Pager.pdf. (Accessed on 01/16/ 2019) 17. Krylov E, Nazarov EG, Miller RA, Tadjikov B, Eiceman GA (2002) Field dependence of mobilities for gas-phase-protonated monomers and proton-bound dimers of ketones by planar field asymmetric waveform ion mobility spectrometer (pfaims). J Phys Chem A 106(22):5437–5444 18. Krylov EV, Nazarov EG, Miller RA (2007) Differential mobility spectrometer: model of operation. Int J Mass Spectrom 266(1):76– 85 19. Shannon CE (1948) A mathematical theory of communication. Bell Syst Tech J 27(4):623–656 20. Chatterjee S, Major GH, Paull B, Rodriguez ES, Kaykhaii M, Linford MR (2018) Using pattern recognition entropy to select mass chromatograms to prepare total ion current chromatograms from raw liquid chromatographymass spectrometry data. J Chromatogr A 1558:21–28 21. Chatterjee S, Singh B, Diwan A, Lee ZR, Engelhard MH, Terry J, Tolley HD, Gallagher NB, Linford MR (2018) A perspective on two chemometrics tools: Pca and mcr, and introduction of a new one: pattern recognition entropy (pre), as applied to xps and tofSims depth profiles of organic and inorganic materials. Appl Surf Sci 433:994–1017 Publisher’snoteSpringer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. Int. J. Ion Mobil. Spec.