Bacteriology Evaluation of the rapid RIDAQUICK Campylobacter® test in a general hospital Cristina Gómez-Camarasa a , José Gutiérrez-Fernández a,b, ⁎, Javier Moises Rodríguez-Granger a , Antonio Sampedro-Martínez a , Antonio Sorlózano-Puerto b , José María Navarro-Marí a a Laboratorio de Microbiología, Hospital Universitario Virgen de las Nieves, Spain b Departamento de Microbiología, Universidad de Granada, 18014 Granada, Spain abstractarticle info Article history: Received 27 August 2013 Received in revised form 28 October 2013 Accepted 1 November 2013 Available online 15 November 2013 Keywords: Campylobacter Immunochromatography Rapid test PCR Culture Diagnosis Stool Antigen test The study objective was to evaluate the effectiveness of the new rapid immunochromatographic test RIDAQUICK Campylobacter® (r-biopharm AG, Darmstadt, Germany) for the qualitative detection of Campylobacter antigens in pathologic feces from primary and specialist care patients. Three hundred feces samples were studied from patients with diarrhea, 50.6% from adults and 49.4% from children, which were received by our microbiology laboratory for coproculture. Campylobacter culture results, with or without PCR data, served as reference values for the comparative evaluation of RIDAQUICK Campylobacter® findings. Campylobacter was detected in 12.3% of samples. The diagnostic accuracy values of the RidaQuick Campylobacter® versus culture were: sensitivity of 87%, specificity of 97%, and positive and negative predictive values of 77% and 98%, respectively. RIDAQUICK Campylobacter® is a rapid test for the diagnosis of enteritis due to Campylobacter and could be an option for the clinical diagnosis of one of the main causes of bacterial enteritis in resource-limited settings. © 2014 Elsevier Inc. All rights reserved. 1. Introduction The genusCampylobacter belongs to the family Campylobacteriaceae (Hindiyeh et al., 2000) and contains 18 species, including Campylobacter jejuni, the cause of around 90% of acute Campylobacter enteritis cases in humans (Hindiyeh et al., 2000; Kawatsu et al., 2008; Granato et al., 2010).This zoonotic disease is prevalent worldwideand transmitted via water and food (Friedman et al., 2000; Granato et al., 2010). C. jejuni and Campylobacter coli colonize the digestive tract of numerous animals, especially birds, and most infections are produced by the consumption of their undercooked meat (Granato et al., 2010). The incubation period of the enteritis ranges from 1 to 7 days, and its signs and symptoms are non-specific. These infections are generally self-limiting (Skirrow and Titus, 2000) and only rarely give rise to extra-intestinal disease (e.g., bacteremia, reactive arthritis, or Guillain-Barré syndrome) (Hannu et al., 2002; Granato et al., 2010; Floch et al., 2012). Hence, antibiotic treatment is not usually indicated for diagnosed patients (Giltner et al., 2013). Nevertheless, detection of this etiologic agent is important for the differential diagnosis with other potential causes of diarrhea (Tissari and Rautelin, 2007) and because the infection may persist. Furthermore, specific antibiotic treatment may be indicated in immunodepressed patients and in children and the elderly, showing an appreciable rate of resistance to fluoroquinolones (Hindiyeh et al., 2000; Tolcin et al., 2000; Granato et al., 2008; Arsenault et al., 2011). For these reasons, rapid etiologic detection is vital in the aforementioned populations. The diagnosis of enteritis due to Campylobacter was traditionally performed using culture, visualization by Gram staining (fuchsin), and conventional biochemical identification methods such as APIsystems (bioMerieux Vitek, Inc, Hazelwood, MO, USA). Campylobacter is a curved Gram-negative rod that requires microaerophilic incubation for up to 48 h at 42 °C using selective media (Bessède et al., 2011a). More recent diagnostic techniques include the detection of antigens (enzyme-linked immunosorbent assay or immunochromatography) or DNA in feces and matrix-assisted laser desorption ionization–time of flight (MALDI-TOF) mass spectrometry of the colonies (Bessède et al., 2011b; Martiny et al., 2013). The antigen detection methods are of particular interest because they yield results within hours or minutes and are easy to apply, especially using the immunochromatographic approach (Floch et al., 2012). In addition, Campylobacter antigens can persist in the clinical sample in the absence of viable microorganisms, allowing primary infections to be detected by this method when the bacteria are not viable, which is a major potential clinical advantage. However, highly variable results have been obtained using commercialized kits (Table 1), and further knowledge is required on the functioning of these kits in clinical samples to establish their usefulness in the diagnosis of this infectious enteritis. The objective of this study was to evaluate the behavior of the new rapid immunochromatographic test RIDAQUICK Campylobacter® Diagnostic Microbiology and Infectious Disease 78 (2014) 101–104 ⁎Corresponding author. Tel.: 0034-958020072. E-mail address: [email protected] (J. Gutiérrez-Fernández). 0732-8893/$ –see front matter © 2014 Elsevier Inc. All rights reserved. http://dx.doi.org/10.1016/j.diagmicrobio.2013.11.009 Contents lists available at ScienceDirect Diagnostic Microbiology and Infectious Disease journal homepage: www.elsevier.com/locate/diagmicrobio
(r-biopharm AG, Darmstadt, Germany) in the detection of Campylobacter antigen in feces samples from patients with diarrhea. 2. Materials and methods Between March and May 2013, 300 consecutive samples of fresh feces samples were studied from patients diagnosed with diarrhea in primary (57.6%) and specialist (42.4%) care; 50.6% were from adults and 49.4% from children under 14 years of age (Table 2). The samples were received by the Microbiology Laboratory of University Hospital Virgen de las Nieves, Granada (Spain), for coproculture and were simultaneously studied with the RIDAQUICK Campylobacter® test to detect the presence of Campylobacter antigens. This hospital is a reference center in Andalusia (southern Spain) serving a population of around 440,000 individuals. Samples were gathered in sterile containers with no transport media and delivered to our laboratory under refrigeration (4 °C), with a maximum delay of 2 hours before their processing. They were processed for coproculture immediately on their reception by culture in CampyBAP® medium with 10% blood (Becton Dickinson, BD, Franklin Lakes, NJ, USA) using a 30-μg cefoxitin disk (BD BBL®) and incubated for 48 h at 42 °C in microaerophilic atmosphere (Campygen®, Oxoid, Basingstoke, UK). Suspicious colonies were identified by means of oxidase cytochrome tests (Difco, Detroit, MI, USA), Gram staining, and mass spectrometry using the Biotyper® system (Bruker Daltonics, Coventry, UK) (He et al., 2010). The samples were also seeded in the usual culture media for enteropathogens for 48 h (XLD® agar [BD] at 37 °C for recovery of Salmonella,Shigella, and Plesiomonas, and CIN® agar [BD] at 30 °C for recovery of Yersinia and Aeromonas)or 24 h (Selenito® broth [Difco] for recovery of Salmonella, followed by a subculture in Hektoen® agar [BD] at 37 °C). Colonies suspected of being enteropathogenic were identified by means of the Biotyper® system (Bruker Daltonics) and the usual biochemical tests (MicroScan; Siemens Healthcare, Rockville, MD, USA). Colonies identified as Salmonella were subjected to the agglutination test to determine the serogroup (Difco). The RIDAQUICK Campylobacter® was used according to the manufacturer’s instructions, mixing well 50 mg or 50 μL of the feces in a tube containing 0.5 mL of reagent A and 0.5 mL of reagent B. The mixture was incubated at room temperature for 5 min, and 150 μLof the supernatant was then placed in the sampling port of the device. After a 15-min interval at room temperature, the reading was made by 2 researchers using simple visual observation, and the result was accepted for evaluation when their readings coincided. The result was considered positive when violet red lines could be seen in the control (C) and test (T) bands, negative when the control line alone appeared, and invalid when no line could be observed. All samples tested were kept at −80 °C in aliquots. In samples with discrepant culture and immunochromatography results, the presence of Campylobacter DNA was also investigated using RIDAGENE Table 1 Results of tests for detection of Campylobacter antigen. Study Test Results No. of samples Tolcin et al., 2000 ProSpecT Campylobacter® (Alexon-Trend, Ramsey, MN, USA): EIA microplate assay. Se = 96%; Sp = 99% 164; GS: culture Endtz et al., 2000 Se = 80%; Sp = 100% 30; GS: culture Hindiyeh et al., 2000 Se = 89%; Sp = 99%; PPV = 80%; NPV = 99% 631; GS: culture Dediste et al., 2003 Se = 89.1%; Sp = 97.7%; PPV = 78.3%; NPV = 99% 1205; GS: culture Tissari and Rautelin, 2007 RidaScreen Campylobacter® (R-biopharm AG, Darmstadt, Germany): EIA microplate assay Se = 69%; Sp = 87% 1050; GS: culture Kawatsu et al., 2008 Campy in-house: immunochromatographic assay Se = 84.8%; Sp = 100% 222; GS: culture Bessède et al., 2011a In-house PCR Se = 80%; PPV = 85% 242; GS: Culture or PCR + EIA Multiplex PCR Seeplex Diarrhea ACE Detection® Se = 89%; PPV = 85% RidaScreen Campylobacter® (R-biopharm AG, Darmstadt, Germany): EIA microplate assay. Se = 91%; PPV = 85% Premier CAMPY® (Meridian Bioscience, Inc. Cincinnati, OH, USA): EIA microplate assay. Se = 95%; PPV = 80% ImmunoCard STAT! Campy® (Meridian Bioscience, Inc.): immunochromatographic assay. Se = 90%; PPV = 70% Dey et al., 2012 Se = 86%; Sp = 100%; GS: culture Se = 90%; Sp = 100%; GS: PCR 463 Floch et al., 2012 PPV = 80.6% 609; GS: culture Couturier et al., 2013 Se = 80%; Sp = 98% 500; GS: culture Granato et al., 2010 Se = 98.4%; Sp = 94.2%; PPV = 92.6%; NPV = 98.8%; GS: culture Se = 98.5%; Sp = 98.2%; PPV = 97.8%; NPV = 98.8%; GS: culture + PCR (discordant) 300 Giltner et al., 2013 Premier CAMPY® (Meridian Bioscience, Inc.: EIA microplate assay. Se = 100%; Sp = 98%; PPV = 91%; NPV = 100% 60; GS: culture Se = 75%; Sp = 96.5%; PPV = 42.9%; NPV = 99% 119; GS: culture EIA = enzyme immunoassay; Se = Sensitivity; Sp = Specificity; PPV = Positive predictive value; NPV = Negative predictive value; GS = Gold standard. 102 C. Gómez-Camarasa et al. / Diagnostic Microbiology and Infectious Disease 78 (2014) 101–104
Bacterial Stool Panel PCR® (r-Biopharm AG, Darmstadt, Germany), a qualitative multiplex real-time PCR assay (Mx3005p system; Strategene, La Jolla, CA, USA) that targets 16s-rDNA of Campylobacter spp. with a fluorogenic target-specific hydrolysis probe. Before the extraction (Maxwell DNA purification kit, Valencia, CA, USA), stool samples were diluted 1:3 with water, intensely vortexed, and centrifuged at 3000 rpm for 30 sec. The diagnostic accuracy of the RidaQuick Campylobacter® test was evaluated by calculating the sensitivity, specificity, and positive and negative predictive values, with exact binomial interval of ClopperPearson. Campylobacter was considered present when the microorganism was isolated or when its DNA was detected in samples with contradictory culture and immunochromatography findings. 3. Results Out of the 300 samples of feces studied (Table 2), the highest percentage with enteropathogens (12.3%) came from children aged less than 14 years referred by primary care centers. Campylobacter was detected in 37 (12.3%) of all samples, and the culture was positive (gold-standard method) for C. jejuni in 31 (83.78%) of these. A different enteropathogen was detected in 12 (8.11%) of the samples from children. The culture and immunochromatography results differed in 12 samples (4% of the total; Table 3), which were therefore studied by PCR; 6 of these samples were culture-negative and PCR-positive and 1 was culture positive (C. jejuni) and PCR-negative. Comparisons between the immunochromatography and the culture and/or PCR results are exhibited in Table 4. The 2 researchers independently gave the same result in all samples, and no sample was unreadable. There was no case in which the test was positive when the enteropathogen isolated was other than Campylobacter. The diagnostic accuracy values of the RidaQuick Campylobacter® versus culture were: sensitivity of 87% (95% confidence interval [CI]: 0.70–0.96), specificity of 97% (95% CI: 0.94–0.97), and positive and negative predictive values of 77% (95% CI: 0.60–0.90) and 98% (95% CI: 0.96–0.99), respectively. The diagnostic accuracy values versus culture plus PCR were: sensitivity of 89% (95% CI: 0.75– 0.97), specificity of 99% (95% CI: 0.97–1.00), and positive and negative predictive values of 94% (95% CI: 0.81–0.99) and 98% (95% CI: 0.96– 1.00), respectively. 4. Discussion Campylobacter is one of the main agents responsible for enteritis in our region, explaining our interest in the availability at primary and emergency care levels of a rapid, simple, and reliable test for its detection in feces samples. This could deliver major clinical benefits for the management of high-risk patients (Hannu et al., 2002; Granato et al., 2008; Floch et al., 2012) and for the early detection of outbreaks (Calciati et al., 2012) by detecting this agent in a rapid manner and with fewer false negatives due to inadequate sample transport. Most patients with intestinal disorders produced by Campylobacter are seen in primary care centers and do not usually require antibiotic treatment or hospitalization. However, besides the need for a rapid test to reduce the morbidity, protect high-risk patients, and detect outbreaks, its availability at primary care level could facilitate therapeutic decision making at hospital level in patients referred for specialist treatment due to their clinical symptoms, contributing to relieving the pressure on emergency departments. This type of test must show adequate diagnostic profitability and be easy to apply, and limitations in the results interpretation must be established. In the present study, application of the RIDAQUICK Campylobacter®, using only the material supplied by the manufacturer, may have helped to disclose the possible cause of the diarrhea at around 20 min after reception of the feces sample. The test was easy to perform and interpret, requires no specific training, and is not very costly (6 €), allowing its use to be recommended in primary and emergency care settings. Discordant results were obtained for 12 (4%) of the 300 samples (Table 3). Four of these (positive culture and negative immunochromatography) can be attributed to a lack of sensitivity of the test; the subsequent PCR test was positive in 3 of these 4 samples but negative in one, which may be because the amount of specific DNA was below the limit of detection of this multiplex PCR. A major limitation of our study is that PCR was only applied in cases of discrepancy between culture and rapid testing and not in cases of false-negative culture results due to suboptimal transport conditions or cases of false-negative rapid test results attributable to the potentially low sensitivity of these tests. Table 4 Results of antigen detection by immunochromatography with RIDAQUICK Campylobacter® versus culture with or without PCR. RIDAQUICK Campylobacter® Positive (no.) Negative (no.) Total (no.) Culture Positive (no.) 27 4 31 Negative (no.) 8 261 269 Culture with PCR Positive (no.) 33 4 37 Negative (no.) 2 261 263 Total 35 265 300 Table 2 Distribution of feces samples and presence of microorganisms by origin and age of the patients studied. Origin No. of feces samples Campylobacter positive Other enteropathogen Total Child Adult With enteropathogen Child Adult Child Adult Primary care 171 112 59 37 21 a 710(9 a Salmonella;1Aeromonas)0 Specialist care 129 36 93 11 1 8 2 (1 Salmonella;1Aeromonas)0 Total 300 148 152 48 22 15 12 0 a One case with simultaneous isolation of Campylobacter and Salmonella. Table 3 PCR results for samples with discrepant antigen detection and culture findings. Sample number Immunochromatography Culture PCR Interpretation 1+−− − 2+−++ 3−++ + 4+−++ 5+−− − 6+−++ 7+−++ 8−++ + 9+−++ 10 −+−+ 11 + −++ 12 −++ + Interpretation: culture plus PCR. 103C. Gómez-Camarasa et al. / Diagnostic Microbiology and Infectious Disease 78 (2014) 101–104
The limitation of PCRs, alongside their laborious nature and high cost, makes immunochromatography of clinical interest. Six of the discrepant samples showed negative culture results but positive PCR results, supporting the use of these rapid tests on the first contact of patients with the health system to reduce the false negatives due to bacteria viability losses. It has been reported that sample transport conditions require optimization in order to achieve more adequate results in bacterial culture (Wang et al., 1983). There were only 2 cases of false positives using immunochromatography (0.66%); however, their clinical significance is difficult to determine without knowledge of the antigen detected in the kit. Only 12 studies published in the English language have evaluated kits for the detection of Campylobacter antigen in feces. The sensitivity of the 8 enzyme immunoassays and 6 immunochromatographic kits ranged between 69% and 100% and the specificity between 87% and 100% (Table 1). The present study obtained poor sensitivity but adequate specificity. Only one of the aforementioned studies (Kawatsu et al., 2008) described the antibody used in the kit, a monoclonal antibody against a C. jejuni cell surface protein of 15 kDa. In addition, this rapid test provides no information on the resistance pattern of the causative Campylobacter isolate, and resistance testing is important in the microbiological diagnosis of campylobacteriosis at a time of increasing macrolide resistance in Campylobacter spp. (Coker et al., 2002; Wieczorek and Osek, 2013). In conclusion, detection of the Campylobacter antigen in pathological feces using RIDAQUICK Campylobacter® is an easily applied and rapid procedure that yields acceptable results for the clinical diagnosis of patients. References Arsenault J, Ravel A, Michel M, Berke O, Gosselin P. Do patients with recurrent reported episodes of campylobacteriosis differ from those with a single disease event? BMC Public Health 2011;11:32. Bessède E, Delcamp A, Sifré E, Buissonnière A, Mégraud F. New methods for detection of campylobacters in stool samples in comparison to culture. J Clin Microbiol 2011a; 49:941–4. Bessède E, Solecki O, Sifré E, Labadi L, Mégraud F. Identification of Campylobacter species and related organisms by matrix assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry. Clin Microbiol Infect 2011b;17:1735–9. Calciati E, Lafuente S, De Simó M, Balfagonc P, Bartolomé R, Caylà J. Campylobacter outbreak in a Barcelona school. Enferm Infecc Microbiol Clin 2012;30:243–5. Coker AO, Isokpehi RD, Thomas BN, Amisu KO, Obi CL. Human campylobacteriosis in developing countries. Emerg Infect Dis 2002;8:237–44. Couturier BA, Couturier MR, Kalp KJ, Fisher MA. Detection of non-jejuni and –coli Campylobacter species from stool specimens with an immunochromatographic antigen detection assay. J Clin Microbiol 2013;51:1935–7. Dediste A, Vandenberg O, Vlaes L, Ebraert A, Douat N, Bahwere P, Butzler JP. Evaluation of the ProSpecT Microplate Assay for detection of Campylobacter: a routine laboratory perspective. Clin Microbiol Infect 2003;9:1085–90. Dey SK, Nishimura S, Okitsu S, Hayakawa S, Mizuguchi M, Ushijima H. Comparison of immunochromatography, PCR and culture methods for the detection of Campylobacter bacteria. J Microbiol Methods 2012;91:566–8. Endtz HP, Ang CW, van den Braak N, Luijendijk A, Jacobs BC, de Man P, van Duin JM, van Belkum A, Verbrugh HA. Evaluation of a new commercial immunoassay for rapid detection of Campylobacter jejuni in stool samples. Eur J Clin Microbiol Infect Dis 2000;19:794–7. Floch P, Goret J, Bessède E, Lehours P, Mégraud F. Evaluation of the positive predictive value of a rapid immunochromatographic test to detect Campylobacter in stools. Gut Pathog 2012;4:17. Friedman CR, Neimann J, Wegener HC, Tauxe RV. Epidemiology of Campylobacter jejuni infections in the United States and other industrialized nations. In: Nachamkin I, Blaser MJ, editors. Campylobacter. 2nd ed. Washington, DC: ASM Press; 2000. p. 121–38. Giltner CL, Saeki S, Bobenchik AM, Humphries RM. Rapid detection of Campylobacter antigen by enzyme immunoassay leads to increased positivity rates. J Clin Microbiol 2013;51:618–20. Granato PA, Chen L, Holiday I, Rawling RA, Novak-Weekley SM, Quinlan T, Musser KA. Comparison of premier CAMPY enzyme immunoassay (EIA), ProSpecT Campylobacter EIA, and ImmunoCard STAT! CAMPY tests with culture for laboratory diagnosis of Campylobacter enteric infections. J Clin Microbiol 2010;48:4022–7. Hannu T, Mattila L, Rautelin H, Pelkonen P, Lahdenne P, Siitonen A, Leirisalo-Repo M. Campylobacter-triggered reactive arthritis: a population-based study. Rheumatology (Oxford) 2002;41:312–8. He Y, Li H, Lu X, Stratton CHW, Tang YV. Mass Spectrometry Biotyper System identifies enteric bacterial pathogens directly from colonies grown on selective stool culture media. J Clin Microbiol 2010;48:3888–92. Hindiyeh M, Jense S, Hohmann S, Benett H, Edwards C, Aldeen W, Croft A, Daly J, Mottice S, Carroll KC. Rapid detection of Campylobacter jejuni in stool specimens by an enzyme immunoassay and surveillance for Campylobacter upsaliensis in the greater Salt Lake City area. J Clin Microbiol 2000;38:3076–9. Kawatsu K, Kumeda Y, Taguchi M, Yamazaki-Matsune W, Kanki M, Inoue K. Development and evaluation of immunochromatographic assay for simple and rapid detection of Campylobacter jejuni and Campylobacter coli in human stool specimens. J Clin Microbiol 2008;46:1226–31. Martiny D, Visscher A, Catry B, Chatellier S, Vandenberg O. Optimization of Campylobacter growth conditions for further identification by matrix assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS). J Microbiol Methods 2013;94:221–3. Skirrow MB, Titus AE. Clinical aspects of Campylobacter infection. In: Nachamkin I, Blaser MJ, editors. Campylobacter. 2nd ed. Washington, DC: ASM Press; 2000. p. 69–88. Tissari P, Rautelin H. Evaluation of an enzyme immunoassay-based stool antigen test to detect Campylobacter jejuni and Campylobacter coli. Diagn Microbiol Infect Dis 2007;58:171–5. Tolcin R, LaSalvia MM, Kirkley BA, Vetter EA, Cockerill 3rd FR, Procop GW. Evaluation of the Alexon-Trend ProSpecT Campylobacter microplate assay. J Clin Microbiol 2000; 38:3853–5. Wang WL, Reller LB, Smallwood B, Luechtefeld NW, Blaser MJ. Evaluation of transport media for Campylobacter jejuni in human fecal specimens. J Clin Microbiol 1983;18: 803–7. Wieczorek K, Osek J. Antimicrobial resistance mechanisms among Campylobacter. Biomed Res Int. 2013;2013:340605. http://dx.doi.org/10.1155/2013/340605. [Epub 2013 Jun 24]. 104 C. Gómez-Camarasa et al. / Diagnostic Microbiology and Infectious Disease 78 (2014) 101–104